Skip to content

Coyote™ H74939186400610Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Coyote (55)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729796633), received between Nov 8, 2023 and Nov 8, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Deflate” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Deflate1
  • Leak/Splash1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 8, 2023Leak/Splash; Material Rupture; Failure to Deflate; Leak/Splash; Material Rupture; Failure to Deflate

    IT WAS REPORTED THAT THE COYOTE BALLOON LEAKED AND WAS UNABLE TO BE DEFLATED. A COYOTE, 4.0MM X 60MM X 150CM, OTW WAS SELECTED FOR USE TO TREAT A CASE OF IN-STENT RESTENOSIS IN THE RIGHT LOWER EXTREMITY. THE TARGET LESION WAS SEVERELY CALCIFIED AND MODERATELY TORTUOUS WITH 100% STENOSIS. THE FIRST COYOTE BALLOON TRACKED WELL OVER THE UNKNOWN 6FR PEDAL SHEATH. A RANGER BALLOON WAS UNABLE TO TRACK OVER THE SHEATH, AND WAS REPLACED WITH A STERLING BALLOON, WHICH TRACKED REASONABLY WELL. THE SECOND COYOTE BALLOON WAS ADVANCED TO THE TARGET LESION AND INFLATED, AND IT WAS OBSERVED THAT THE BALLOON WAS LEAKING. THE COYOTE BALLOON WAS UNABLE TO BE DEFLATED, AND THE ENTIRE SYSTEM WAS PULLED FROM THE PEDAL ACCESS. THE PROCEDURE WAS CANCELLED, AS THE IN-STENT STENOSIS WAS DEEMED TO BE TOO EXTENSIVE. THERE WERE NO REPORTED PATIENT COMPLICATIONS. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. THE SECOND COYOTE BALLOON WAS INFLATED TO NOMINAL PRESSURE WHEN THE LEAK WAS OBSERVED. THE COYOTE BALLOON WAS REMOVED INTACT FROM THE PATIENT IN A SEMI-DEFLATED STATE OVER THE COURSE OF 10-15 SECONDS BY WALKING THE BALLOON OFF THE GUIDEWIRE. ONCE REMOVED FROM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18093639 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates