Coyote™ H74939186250810Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Coyote (55)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 08714729797180), received between Mar 25, 2019 and Jun 4, 2025 — classified as 2 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture5
- Adverse Event Without Identified Device or Use Problem2
- Entrapment of Device2
- Difficult to Advance1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient2
- Device Embedded In Tissue or Plaque1
- No Consequences Or Impact To Patient1
- Vascular Dissection1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 4, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 2.5MM X 80MM X 150CM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON BURST EVEN WITHOUT REACHING THE BURST RATED PRESSURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22142129 ↗MalfunctionOct 22, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE STENOSED TARGET LESION WAS LOCATED IN THE TIBIAL ARTERY. A 2.5MM X 80MM X 150CM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT 14 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20509199 ↗InjuryJun 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ELEGANCE CLINICAL STUDY IT WAS REPORTED THAT VESSEL DISSECTION OCCURRED. IN (B)(6) 2023, THE SUBJECT UNDERWENT TREATMENT WITH RANGER DRUG COATED BALLOON AND ELUVIA DRUG-ELUTING STENT AS A PART OF THE ELEGANCE CLINICAL TRIAL. THE TARGET LESION #001 WAS IN THE RIGHT MID POPLITEAL ARTERY EXTENDING INTO RIGHT DISTAL POPLITEAL ARTERY WITH PROXIMAL REFERENCE VESSEL DIAMETER OF 4 MM AND DISTAL REFERENCE VESSEL DIAMETER OF 4 MM WITH LESION LENGTH OF 60 MM AND 75% STENOSIS AND WAS CLASSIFIED AS TASC II D LESION. TREATMENT OF TARGET LESION WAS PERFORMED BY DILATION USING STUDY DEVICE, 4 MM X 60 MM OF RANGER DRUG-COATED BALLOON. POST TREATMENT, THE FINAL RESIDUAL STENOSIS WAS NOTED TO BE 30%. THE TARGET LESION #002 WAS IN THE RIGHT DISTAL SUPERFICIAL FEMORAL ARTERY WITH PROXIMAL REFERENCE VESSEL DIAMETER OF 6 MM AND DISTAL REFERENCE VESSEL DIAMETER OF 6 MM WITH LESION LENGTH OF 15 MM AND 80% STENOSIS AND WAS CLASSIFIED AS TASC II D LESION. TREATMENT OF TARGET LESION WAS PERFORMED BY PLACEMENT OF STUDY DEVICE, 6 MM X 40 MM ELUVIA DRUG-ELUTING STENT. FOLLOWING POST-DILATION WAS PERFORMED WITH 6 MM X 40 MM MUSTANG PTA BALLOON AND 4 MM X 40 MM ADVANCE PTA BALLOON AND THE FINAL RESIDUAL STENOSIS W
Patient problems: Vascular Dissection
MAUDE report 17038799 ↗MalfunctionJul 25, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A 2.5MM X 120MM X 150CM STERLING SL AND A 2.5MM X 80MM X 150CM COYOTE BALLOON CATHETERS WERE ADVANCED FOR DILATION. BOTH BALLOON REACHED THE LESION SITE. HOWEVER, DURING INFLATION AT 12 ATMOSPHERES, BOTH BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. THE PATIENT WAS IN STABLE CONDITION FOLLOWING THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15098935 ↗MalfunctionNov 29, 2021Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT CATHETER ENTRAPMENT OCCURRED. THE TARGET LESION WAS LOCATED IN A VESSEL BELOW THE KNEE. A 300CM STRAIGHT TIP V-14 CONTROL WIRE WAS ADVANCED TO CROSS THE LESION. HOWEVER, IT WAS NOTED THAT THE WIRE TIP WAS TORN APART FROM THE WIRE BODY AND NEED TO USE A SNARE TO REMOVE IT OUT. IT WAS ALSO FOUND THAT THE WIRE BECAME STUCK IN THE 2.5MM X 80MM X 150CM COYOTE BALLOON CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12887104 ↗MalfunctionAug 18, 2021Difficult to Advance; Adverse Event Without Identified Device or Use Problem; Difficult to Advance; Adverse Event Without Identified Device or Use Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 27JUL2021. IT WAS REPORTED THAT THE CATHETER COULD NOT MOVE OVER THE GUIDEWIRE. A 2.5X80X150 COYOTE BALLOON WAS SELECTED FOR A PROCEDURE. DURING THE PROCEDURE, IT WAS REPORTED THAT THE COYOTE WAS NOT ABLE TO MOVE OVER THE .014 V-14 GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING AN ALTERNATE DEVICE. HOWEVER, DEVICE ANALYSIS REVEALED THAT AN UNKNOWN .014 WIRE WAS STUCK IN THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12336029 ↗InjuryDec 30, 2019Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT A DEVICE COMPONENT WAS LEFT INSIDE THE PATIENT. THE BALLOON PROTECTOR WAS NOT REMOVED PRIOR TO INTRODUCING A 2.5MM X 80MM X 150CM COYOTE BALLOON CATHETER INTO THE PATIENT. THE BALLOON PROTECTOR SLEEVE CAME OFF AND WENT DOWN THE PATIENT'S ARTERY AND WAS NOT ABLE TO BE REMOVED. THE PROCEDURE WAS IMMEDIATELY STOPPED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS DOING OKAY. IT WAS FURTHER REPORTED THAT PATIENT COMPLICATIONS OCCURRED. PROTECTIVE OUTER SHEATH FROM 2.5X80X150 COYOTE BALLOON WAS RETAINED IN PATIENTS BODY, AND THE PROCEDURE PERFORMED WAS: EXPLORATION DISTAL LEFT ANTERIOR TIBIALIS ARTERY WITH REMOVAL OF FOREIGN BODY AND VEIN PATCH ANGIOPLASTY. PREOPERATIVE DIAGNOSIS WAS NON- HEALING ULCER ON LEFT FOOT, AND THE PROCEDURE PERFORMED WAS: PERCUTANEOUS ULTRASOUND GUIDED DYNAMIC AORTOGRAM WITH BILATERAL LOWER EXTREMITY RUNOFF WITH INTERPRETATION LEFT POPLITEAL AND ANTERIOR TIBIALIS ARTERY DIRECTIONAL ATHERECTOMY WITH LEFT POPLITEAL ARTERY DRUG COATED BALLOON ANGIOPLASTY AND LEFT AT ANGIOPLASTY. PATIENT WAS DISCHARGED HOME ON 12DEC2019. NO PROGNOSIS NOTES FOUND YET.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9532593 ↗MalfunctionNov 4, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE HAD OCCURRED. A TIBIALPLASTY PROCEDURE WAS BEING PERFORMED. THE 2.5MM X 80MM X 150CM COYOTE BALLOON RUPTURED ON THE FIRST INFLATION AT NOMINAL PRESSURE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DIFFERENT DEVICE WITHOUT ISSUE OR PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9274201 ↗MalfunctionApr 15, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA BRACHIAL ARTERY. THE 60% STENOSED, 50 X 9.62MM, CONCENTRIC, DE NOVO TARGET LESION WAS LOCATED IN A MILDLY TORTUOUS AND NON-CALCIFIED BRACHIOCEPHALIC VEIN. A 2.5MM X 80MM X 150MM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, UPON INITIAL INFLATION AT 12 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8515609 ↗MalfunctionMar 25, 2019Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT CATHETER ENTRAPMENT OCCURRED. AFTER A 300 CM VICTORY GUIDE WIRE CROSSED THE LESION, A 2.5MM X 80MM X 150CM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, THE WIRE BECAME STUCK WITH THE BALLOON SHAFT AND IT WOULD NOT MOVE, WHEN BOTH DEVICES WERE COMPATIBLE. ACCORDING TO THE TECHNICIAN, THE GUIDE WIRE AND THE BALLOON CATHETER MAY HAVE NOT PROPERLY FLUSHED. THE PROCEDURE WAS COMPLETED WITH ANOTHER WIRE AND BALLOON CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8449189 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026