Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729803591), received between Dec 21, 2016 and Dec 21, 2016 — classified as 1 malfunction.

The most frequently reported specific device problem is “Unable to Obtain Readings” (1 report).

3 other recalls cover devices cleared under the same FDA submission (P920047); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Unable to Obtain Readings1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 21, 2016Unable to Obtain Readings; Unable to Obtain Readings

    REPORTABLE BASED ON RETURNED DEVICE ANALYSIS COMPLETED ON DECEMBER 5, 2016. IT WAS REPORTED THAT NO ELECTROGRAMS DISPLAYED. THE TARGET LOCATION WAS THE RIGHT ATRIUM. DURING THE PROCEDURE, NO ELECTROGRAMS WERE DISPLAYING FROM THE BLAZER PRIME® HTD CATHETER AND THE CABLE HAD NOISE. THE NOISE WAS RESOLVED WITH EXCHANGING THE CABLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS FINE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED FLUID UNDER RING SEAL 1 AND A BLUE FIBER ENCASED IN THE SEAL OF R3.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6196040 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P920047. One submission often covers many models, so check the product and lot details in each record.

  • Z-0202-2019 · Jun 20, 2018 · Open, Classified
    Boston Scientific Maestro 4000 Cardiac Ablation System, Firmware Version 5.14, Model Numbers: M00440000, M0044000D0, M0044000H0,M0044000R0, M0044000Z0,
  • Z-0973-2011 · Jul 14, 2009 · Terminated
    RF Ablation System Foot Switch, used with the EPT-1000 RF Generator, Foot Switch Model Number UPN M0048400, and in kits with UPNs M0044810 and M0044811. Intended use: The EPT-1000
  • Z-1617-2008 · Feb 14, 2008 · Terminated
    Boston Scientific SteeroCath-T Ablation Catheter. This catheter is indicated for general indications, treatment of supraventricular tachycardia (svt) and ventricular tachycardia (

Other Cardiac ablation percutaneous catheter devices

All Cardiac ablation percutaneous catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page