Epic™ Vascular H74939200061020Recalls & reported problems

Boston Scientific Corporation · Self-Expanding Stent System · Class 3

All models of Epic Vascular (50)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729805205), received between May 30, 2018 and Apr 26, 2024 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Defective Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Device2
  • Activation Failure1
  • Bent1
  • Break1
  • Difficult to Advance1
  • Positioning Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 26, 2024Difficult to Advance; Positioning Problem; Difficult to Advance; Positioning Problem

    IT WAS REPORTED THAT STENT MIGRATED AND DID NOT FULLY COVER THE STENOSIS. THE PATIENT WAS PRESENTED WITH LEFT FEMORAL ARTERY OCCLUSION. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6X100X120 EPIC VASCULAR WAS ADVANCED FOR TREATMENT. HOWEVER, THE STENT WAS DIFFICULT TO DELIVER TO THE TARGET POSITION. WHEN DEPLOYED, IT WAS NOTED THAT THE STENT MIGRATED AND DID NOT FULLY COVER THE STENOSIS. THERE WAS NO ADDITIONAL INTERVENTION PERFORMED. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19190875 ↗
  • MalfunctionMar 27, 2024Defective Device; Defective Device

    IT WAS REPORTED THAT STENT PROFILE PROBLEM OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY (SFA). A 6X100X120 EPIC VASCULAR STENT WAS DEPLOYED AND POST-DILATED WITH A NON-BOSTON SCIENTIFIC BALLOON; HOWEVER, A VISIBLE SIZE DIFFERENCE BETWEEN THE STENT AND A PREVIOUSLY DEPLOYED 6X150 STENT IN THE DISTAL SFA WAS NOTED. THE PHYSICIAN BELIEVES THAT THE DEPLOYED STENT HAD AN INCORRECT SIZE. THE STENT REMAINED IMPLANTED; HOWEVER, THE STENT DID NOT FULLY COVER TARGET LESION. THE PROCEDURE WAS COMPLETED. THERE WERE NO COMPLICATIONS REPORTED AND THE PATIENT WAS OKAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18985381 ↗
  • InjuryAug 16, 2021Defective Device; Defective Device

    IT WAS REPORTED THAT STENT FORESHORTENING OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY CALCIFIED ARTERY. A 6X100X120 EPIC VASCULAR STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING DEPLOYMENT, THE STENT WAS FORESHORTENED AND WAS STUCK IN THE CENTER AREA AND MISSED THE TARGET LESION. THE PHYSICIAN EXTENDED THE STENT AND ADDED AN ADDITIONAL STENT TO THE STACKED AREA AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12325599 ↗
  • InjuryOct 30, 2019Activation Failure; Activation Failure

    IT WAS REPORTED THAT STENT DID NOT FULLY EXPAND POST DEPLOYMENT. THE TARGET LESION WAS IN THE SEVERELY TORTUOUS AND MILDLY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6X100X120 EPIC VASCULAR STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING DEPLOYMENT THE PROXIMAL END OF THE STENT STARTED TO STACK AND WAS DEPLOYING WITHIN ITSELF. SUBSEQUENTLY, A BALLOON WAS INFLATED AND DRAGGED WITHIN THE STENT TO REMEDY THE PROXIMAL END OF THE STENT. THE STENT DELIVERY SYSTEM WAS REMOVED AND THERE WAS NO PATIENT COMPLICATIONS REPORTED. THE PATIENT WAS FINE POST PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9255409 ↗
  • InjuryMay 30, 2018Bent; Break; Bent; Break

    IT WAS REPORTED THAT STENT DAMAGE AND FRACTURE OCCURRED. VASCULAR ACCESS WAS OBTAINED UTILIZING RETROGRADE APPROACH FROM THE FOOT. THE CHRONIC TOTALLY OCCLUDED (CTO) TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND CALCIFIED SUPERFICIAL FEMORAL ARTERY (SFA). A 6X100X120 EPIC¿ VASCULAR STENT WAS ADVANCED THROUGH THE HUNTERS CANAL BACK PROXIMAL TO LOWER MID SFA THROUGH THE SEGMENT WHERE THE CTO WAS CROSSED EARLIER BY THE RETROGRADE APPROACH. THE CTO SEGMENT WAS HYPER-DILATED FOR THE NATURE OF THE CTO. THE PHYSICIAN SLOWLY FLOWERED OPEN THE EPIC DISTALLY AND ONCE ANCHORED, CONTINUED TO THUMBWHEEL RELEASE THE STENT. AT WHICH POINT THE STENT STARTED TO QUICKLY FORESHORTEN IN THE ECTATIC AREA. THE STENT THEN APPEARED TO INVAGINATE ON ITSELF AND POSSIBLY FRACTURED IN THAT SEGMENT. THE LESION WAS A CALCIFIED CTO, HOWEVER, THE LOCATION WHERE THE STENT HAD AN ISSUE WAS A VERY DILATED AREA. A 5X200 CHARGER PTA BALLOON CATHETER WAS INITIALLY ADVANCED FROM THE TRADITIONAL APPROACH BUT FAILED TO CROSS AS THE STENT WAS NOT ALLOWING PASSAGE IN THAT DIRECTION. THEREFORE AN ADDITIONAL 5X100 STERLING BALLOON CATHETER WAS THEN ADVANCED FROM THE FOOT TO CROSS AND POST-DILATE THE STENT. THE PROCEDUR

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7555360 ↗