Skip to content

Epic™ Vascular H74939200068020Recalls & reported problems

Boston Scientific Corporation · Self-Expanding Stent System · Class 3

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729805366), received between Feb 10, 2017 and Dec 2, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Bent” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Bent1
  • Contamination1
  • Device Damaged Prior to Use1
  • Incomplete or Missing Packaging1
  • Premature Activation1

Patient problems

  • Intimal Dissection1
  • No Clinical Signs, Symptoms or Conditions1
  • No Patient Involvement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 2, 2025Contamination; Incomplete or Missing Packaging; Contamination; Incomplete or Missing Packaging

    IT WAS REPORTED THAT THE STENT WAS NOT IN A SEALED STERILE PACKAGE.THE TARGET LESION WAS LOCATED IN THE RIGHT LOWER LEG. A 6X80X120 EPIC VASCULAR WAS SELECTED FOR RIGHT LOWER LEG ANGIOGRAPHY. HOWEVER, IT WAS NOTED THAT THE STENT WAS IN AN UNOPENED BOX BUT NOT IN A SEALED STERILE PACKAGE. THE DEVICE WAS IMMEDIATELY SET ASIDE AS SOON AS THE BOX WAS OPENED AND WAS NEVER INTRODUCED TO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23695620 ↗
  • MalfunctionJun 20, 2017Premature Activation; Device Damaged Prior to Use; Premature Activation; Device Damaged Prior to Use

    IT WAS REPORTED THAT INADVERTENT STENT DEPLOYMENT OCCURRED. A 6MMX80MMX120MM EPIC¿ STENT WAS SELECTED FOR USE. HOWEVER UPON OPENING THE PACKAGE, IT WAS NOTED THAT THE TIP OF THE STENT WAS PARTIALLY OUT OF THE DELIVERY SYSTEM. THE DEVICE WAS NOT USED AND THE PROCEDURE WAS COMPLETED WITH A 6MMX80MM INNOVA STENT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.

    Patient problems: No Patient Involvement

    MAUDE report 6654078 ↗
  • InjuryFeb 10, 2017Bent; Bent

    SAME CASE AS MDR ID# 2134265-2017-00764. IT WAS REPORTED THAT STENT DAMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE RIGHT SUPERFICIAL FEMORAL ARTERY. AN EPIC¿ VASCULAR STENT WAS ADVANCED AND IMPLANTED. THE DELIVERY SYSTEM AND GUIDEWIRE WERE INADVERTENTLY REMOVED TOGETHER FROM THE PATIENT. A NEW WIRE WAS ENGAGED AND POST-DILATION WAS PERFORMED. THE PHYSICIAN WAS NOT SATISFIED WITH THE RESULTS AND IMPLANTED A SECOND EPIC¿ VASCULAR DISTAL TO THE FIRST. UPON REMOVAL OF THE DELIVERY SYSTEM, THE NOSECONE BECAME STUCK ON THE PREVIOUSLY IMPLANTED STENT, CAUSING THE IMPLANTED STENT TO ELONGATE. THE DEVICE WAS REMOVED BY DISENGAGING THE WHEEL LOCK ON THE HANDLE AND RESHEATHING THE DEPLOYMENT SYSTEM. THE SECOND EPIC STENT WAS POST-DILATED. IT WAS NOTED THAT A VESSEL DISSECTION HAD OCCURRED. A NON-BSC STENT WAS DEPLOYED TO COVER BOTH PREVIOUSLY IMPLANTED STENTS. THERE WERE NO FURTHER PATIENT COMPLICATIONS REPORTED.

    Patient problems: Intimal Dissection

    MAUDE report 6321670 ↗

Other STENT, ILIAC devices

All STENT, ILIAC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page