Epic™ Vascular H74939200081070Recalls & reported problems
Boston Scientific Corporation · Self-Expanding Stent System · Class 3
All models of Epic Vascular (50)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729805571), received between Nov 27, 2017 and Sep 28, 2020 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Detachment Of Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment Of Device Component1
- Material Deformation1
- Occlusion Within Device1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Patient Involvement1
- Reocclusion1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2020Material Deformation; Material Deformation
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 21-SEP-2020. IT WAS REPORTED THAT SHAFT KINK OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND SEVERELY CALCIFIED SHUNT. AN 8X100X75MM EPIC STENT WAS SELECTED FOR USE. HOWEVER, WHEN THE DEVICE WAS TAKEN OUT, THE SHAFT WAS FOUND BENT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED STENT PARTIAL DEPLOYMENT.
Patient problems: No Patient Involvement
MAUDE report 10593192 ↗InjuryMay 29, 2018Detachment Of Device Component; Detachment Of Device Component
SAME CASE AS MDR ID: 2134265-2018-05320. IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED LEFT COMMON ILIAC ARTERY AND THE THROMBOSED PREVIOUSLY PLACED WALLSTENT IN THE EXTERNAL ILIAC ARTERY (EIA). THE WALLSTENT OF EIA AND BLOOD VESSEL WERE TORTUOUS TO AN UNBELIEVABLE EXTENT. A NON-BSC GUIDE WIRE WAS ADVANCED VIA RIGHT FEMORAL APPROACH BUT FAILED TO CROSS. THE DEVICE WAS THEN ADVANCED VIA LEFT FEMORAL BUT IT WAS STILL UNABLE TO CROSS. SUBSEQUENTLY, WITH THE AID OF A NON-BSC SHEATH, THE WIRE WAS ADVANCED VIA LEFT BRACHIAL BUT STILL FAILED. APPROACH WAS PERFORMED AGAIN VIA LEFT FEMORAL AND THE GUIDE WIRE WAS ABLE TO CROSS. A 3MM-20MM STERLING BALLOON CATHETER WAS ADVANCED FOR PRE-DILATATION. FOLLOWING PLAIN OLD BALLOON ANGIOPLASTY WITH A 6MM-60MM STERLING BALLOON CATHETER, A 8X100X75 EPIC¿ VASCULAR STENT WAS PLACED. UPON REMOVING THE STENT DELIVERY SYSTEM (SDS) POST-DEPLOYMENT, RESISTANCE WAS ENCOUNTERED. WHEN THE DEVICE WAS PULLED, A SNAPPING SOUND WAS HEARD. THE SDS WAS REMOVED, HOWEVER; IT WAS NOTICED THAT THE TIP GOT DETACHED AND THE DETACHED FRAGMENT COULD NOT BE FOUND. THE PHYSICIAN ATTEMPT
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7548569 ↗InjuryNov 27, 2017Occlusion Within Device; Occlusion Within Device
SAME CASE AS MDR ID 2134265-2017-11238. IT WAS REPORTED THAT IN-STENT RESTENOSIS (ISR) OCCURRED. ON AN UNSPECIFIED DATE, AN EPIC STENT WAS IMPLANTED TO TREAT A TARGET LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA). IN (B)(6) 2017, THE PATIENT PRESENTED FOR TREATMENT OF THE 100% RESTENOSED 8X100X75 EPIC¿ VASCULAR STENT LOCATED IN A MILDLY TORTUOUS AND MODERATELY CALCIFIED SFA. A 7.0MMX40MMX135 CM (4F) STERLING¿ BALLOON CATHETER WAS ADVANCED FOR DILATATION. UPON INFLATION AT 12 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.
Patient problems: Reocclusion
MAUDE report 7059500 ↗
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Omnilink Elite 1012631-59Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page