Skip to content

Epic™ Vascular H74939200084020Recalls & reported problems

Boston Scientific Corporation · Self-Expanding Stent System · Class 3

All models of Epic Vascular (50)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729805649), received between Feb 6, 2020 and Nov 18, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Premature Activation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Activation3
  • Activation, Positioning or Separation Problem1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 18, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. AN UNSPECIFIED STENT WAS SELECTED FOR USE DURING THE PROCEDURE. HOWEVER, IT WAS NOTED THAT THE DISTAL TIP OF THE DEVICE WAS DAMAGED. THE STENT WAS THEN REPLACED WITH AN 8X40X120 EPIC VASCULAR STENT; HOWEVER, THE DEVICE COULD NOT CROSS THE LESION. CONSEQUENTLY, THE PROCEDURE COULD NOT BE COMPLETED. NO PATIENT INJURY WAS REPORTED. IT WAS FURTHER REPORTED THAT TWO 8X40X120 EPIC VASCULAR STENTS WERE USED IN THE PROCEDURE, MOREOVER, THE PATIENT WAS LOCALLY SEDATED WHEN THE ISSUE OCCURRED, AND THE DEVICE WAS REMOVED USING THE NORMAL METHOD DURING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20708471 ↗
  • MalfunctionDec 2, 2022Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    IT WAS REPORTED THAT STENT PARTIAL DEPLOYMENT OCCURRED. THE STENOSED TARGET LESION WAS LOCATED IN THE ILIAC ARTERY. DURING UNPACKING OF AN 8X40X120 EPIC VASCULAR STENT AND WHEN IT WAS ABOUT TO ADVANCE THROUGH THE GUIDEWIRE, THE TIP OF THE STENT WAS FOUND TO BE PARTIALLY DEPLOYED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15906488 ↗
  • MalfunctionAug 30, 2022Premature Activation; Premature Activation

    IT WAS REPORTED THAT STENT INADVERTENT DEPLOYMENT OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY. THE 80% STENOSED TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND SEVERELY CALCIFIED RIGHT ILIAC ARTERY. AN 8X40X120 EPIC VASCULAR STENT WAS SELECTED FOR USE; HOWEVER, WHEN THE PHYSICIAN OPENED THE PACKAGE, HE FOUND THE STENT WAS OUTSIDE THE PROTECTIVE DELIVERY SHEATH. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15320873 ↗
  • MalfunctionMay 24, 2022Premature Activation; Premature Activation

    IT WAS REPORTED THAT STENT PREMATURE DEPLOYMENT OCCURRED. THE TARGET LESION WAS LOCATED IN ILIAC ARTERY. A 8X40X120 EPIC VASCULAR STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING THE PROCEDURE, WHEN THE STENT WAS FLUSHED, THE TIP OF THE STENT AUTOMATICALLY DEPLOYED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14492828 ↗
  • MalfunctionFeb 6, 2020Premature Activation; Premature Activation

    IT WAS REPORTED THAT STENT PARTIAL INADVERTENT DEPLOYMENT OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED ILIAC ARTERY. DURING UNPACKING OF A 8X40X120 EPIC STENT, IT WAS NOTED THAT THE TAIL END OF THE STENT HAD DEPLOYED 0.5CM. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED AND THE PATIENT STATUS WAS STABLE.

    Patient problems: No Patient Involvement

    MAUDE report 9676747 ↗

Other STENT, ILIAC devices

All STENT, ILIAC devices →