Epic™ Vascular H74939200098020Recalls & reported problems
Boston Scientific Corporation · Self-Expanding Stent System · Class 3
All models of Epic Vascular (50)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729805861), received between Feb 10, 2021 and Feb 6, 2024 — classified as 3 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Deformation1
- Material Separation1
Patient problems
- Foreign Body In Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 6, 2024Material Separation; Material Separation
IT WAS REPORTED VIA USER MEDWATCH MW5149467 THAT MATERIAL SEPARATION AND UNRETRIEVABLE DEVICE COMPONENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT EXTERNAL ILIAC ARTERY. A 9X80X120 EPIC VASCULAR STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING DEPLOYMENT, THE NOSE CONE DETACHED FROM THE STENT. THE DETACHED COMPONENT COULD NOT BE RETRIEVED AS IT WAS NOT RADIOPAQUE AND COULD NOT BE SEEN WHERE IT WAS LODGED. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 18653009 ↗InjuryNov 27, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. THE STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY CALCIFIED ARTERY. A 9 X 80 X 120 EPIC STENT WAS ADVANCED FOR TREATMENT. HOWEVER, UPON REMOVE, IT WAS NOTED THAT THE TIP WAS MISSING FROM THE NOSE CONE FROM THE DELIVERY SYSTEM. MULTIPLE ANGIOGRAMS WERE DONE ON THE VESSELS. THE TIP WAS NOT RADIOPAQUE AND THERE WERE NO EMBOLISMS CAUSED FROM THE TIP. THE DETACHED PORTION REMAINS INSIDE THE PATIENT AND THEY ARE ARRANGED FOR A FOLLOW UP TO CHECK THE ANKLE BRACHIAL INDEX AND AN ULTRASONOGRAM.
Patient problems: Foreign Body In Patient
MAUDE report 18208690 ↗InjuryFeb 10, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT STENT DAMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE RIGHT COMMON ILIAC ARTERY. A 9X80X120 EPIC VASCULAR STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING PROCEDURE, IT WAS NOTED THAT THE STENT SEEMED TO FORESHORTEN BY AT LEAST 20 MM. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED. IT WAS FURTHER REPORTED THAT THE DEVICE WAS NOT REMOVED AND THE PROCEDURE WAS COMPLETED BY USING ADDITIONAL STENT TO COVER THE STENOSED AREA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11306768 ↗
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Omnilink Elite 1012631-59Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page