Epic™ Vascular H74939200091070Recalls & reported problems
Boston Scientific Corporation · Self-Expanding Stent System · Class 3
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729805892), received between Jun 28, 2018 and Dec 22, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Bent” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Bent1
- Difficult to Remove1
- Entrapment of Device1
- Failure to Advance1
- Material Deformation1
- Migration or Expulsion of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2023Entrapment of Device; Difficult to Remove; Entrapment of Device; Difficult to Remove
IT WAS REPORTED THAT DIFFICULTY IN REMOVING FROM GUIDEWIRE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN SEVERELY TORTUOUS AND SEVERELY CALCIFIED COMMON ILIAC ARTERY. AFTER CROSSING THE LESION WITH A NON-BSC GUIDEWIRE, A 9 X 100 X 75 EPIC STENT WAS ADVANCED AND SUCCESSFULLY DEPLOYED. AN ATTEMPT TO WITHDRAW THE DELIVERY SYSTEM WAS MADE, BUT IT WAS DIFFICULT TO REMOVE AND WAS STUCK TOGETHER WITH THE NON-BSC GUIDEWIRE. BOTH DEVICES WERE REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18383876 ↗MalfunctionNov 18, 2021Failure to Advance; Material Deformation; Failure to Advance; Material Deformation
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 09-NOV-2021. IT WAS REPORTED THAT CROSSING DIFFICULTIES AND KINK OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA CONTRALATERAL APPROACH. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND MODERATELY CALCIFIED EXTERNAL ILIAC ARTERY. FOLLOWING A 6F INTRODUCER SHEATH WAS PLACED, A .035 NON-BSC GUIDEWIRE WAS CROSSED THE LESION. PRE-DILATION WAS PERFORMED RESULTING TO A 60% RESIDUAL STENOSIS. A 9X100X75 EPIC VASCULAR STENT WAS ADVANCED FOR TREATMENT. HOWEVER, THE DEVICE UNABLE TO CROSS THE LESION AND KINK WAS NOTED. THE PROCEDURE WAS COMPLETED WITHOUT REPLACING THE DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED STENT DEPLOYED OUTSIDE OF THE SHEATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12836349 ↗InjuryJun 28, 2018Bent; Migration or Expulsion of Device; Bent; Migration or Expulsion of Device
IT WAS REPORTED THAT STENT DAMAGE AND MIGRATION OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED ILIAC ARTERY. AFTER PRE-DILATION, A 9X100X75MM EPIC¿ VASCULAR STENT WAS IMPLANTED FOR TREATMENT. HOWEVER, WHEN PHYSICIAN TRIED TO ADVANCE THE SHEATH FURTHER WHILE USING DILATOR, IT WAS NOTED THAT THE EDGE OF THE STENT WAS CAUGHT IN THE GAP AND GOT FLATTENED. UPON PUSHING FURTHER, THE STENT SHIFTED. THE PROCEDURE WAS COMPLETED BY IMPLANTING ANOTHER STENT INSIDE THE EPIC STENT. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED. THE FLOW UNTIL THE DISTAL WAS ALSO GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7647823 ↗
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Omnilink Elite 1012631-59Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page