Skip to content

Emerge™ Push H7493919012150Recalls & reported problems

Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729805991), received between May 30, 2023 and Nov 25, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1
  • Failure to Advance1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 25, 2024Failure to Advance; Detachment of Device or Device Component; Failure to Advance; Detachment of Device or Device Component

    IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. THE PATIENT PRESENTED WITH SEVERE CORONARY ARTERY DISEASE. A 1.50MM X 12MM EMERGE PUSH BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, THE DEVICE BECAME STUCK WHILE CROSSING THE LESION AND WHEN AN ATTEMPT WAS MADE TO REMOVE THE DEVICE, THE BALLOON TIP BROKE. THE REMAINDER OF THE CATHETER WAS SUCCESSFULLY REMOVED. THE PROCEDURE WAS SUSPENDED, AND THE PATIENT WAS TRANSFERRED TO A HIGHER LEVEL OF CARE AT A LARGER HOSPITAL.

    Patient problems: Foreign Body In Patient

    MAUDE report 20775339 ↗
  • MalfunctionMay 30, 2023Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. AN UNKNOWN BOSTON SCIENTIFIC (BSC) BALLOON CATHETER WAS USED; HOWEVER, THE SHAFT WAS BROKE. THE PROCEDURE WAS COMPLETED WITH A NON-BSC DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED. IT WAS FURTHER REPORTED THAT THE DEVICE WAS A 1.50MM X 12MM EMERGE PUSH BALLOON CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17026026 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page