Rubicon™ 14 H74939212014150Recalls & reported problems
Boston Scientific Corporation · Support Catheter · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729832140), received between Oct 26, 2018 and Feb 10, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Difficult to Advance1
- Difficult to Remove1
- Entrapment of Device1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 10, 2026Fracture; Difficult to Advance
IT WAS REPORTED THAT DEVICE FRACTURE OCCURRED. A 14, 150CM (BOX/5) RUBICON WAS SELECTED FOR USE. DURING THE PROCEDURE, WHEN THE DEVICE ATTEMPTED TO LOAD THE GUIDEWIRE, IT WOULD NOT PASS AND IT WAS FOUND THAT THE DEVICE WAS FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24309630 ↗MalfunctionDec 2, 2025Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT DEVICE ENTRAPMENT OCCURRED. A RUBICON 14 SUPPORT CATHETER WAS SELECTED FOR USE IN A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE TO TREAT VASCULAR STENOSIS. DURING THE PROCEDURE, THE WIRE BECAME STUCK IN THE CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23693232 ↗MalfunctionFeb 16, 2022Difficult to Remove; Difficult to Remove
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 25JAN2022. IT WAS REPORTED THAT A MICROWIRE WAS DIFFICULT TO REMOVE FROM THE CATHETER. THE 100% OCCLUDED TARGET LESION WAS LOCATED IN THE NON TORTUOUS AND NON CALCIFIED TIBIALIS ANTERIOR DEXTRA. A 150CM RUBICON 14 WAS SELECTED FOR USE. DURING THE PROCEDURE, AT THE TIME OF ANGIOGRAPHY EVALUATION, THE MICROWIRE WAS DIFFICULT TO REMOVE FROM THE RUBICON 14 MICROCATHETER SO IT BROKE ON THE RUBICON 14 MICROCATHETER. THE DEVICE WAS REMOVED VIA JR 6FR GUIDE CATHETER AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT WAS STABLE POST PROCEDURE. HOWEVER, DEVICE ANALYSIS REVEALED THAT THERE WAS A SEPARATION LOCATED 166.5CM FROM THE HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13546430 ↗InjuryOct 26, 2018Break; Break
IT WAS REPORTED VIA USER FACILITY REPORT # MW5079859 THAT THE WIRE TIP BROKE AND REMAINED INSIDE PATIENT'S BODY. A RUBICON 14 WAS SELECTED FOR USE. DURING PROCEDURE, WHILE TRYING TO CROSS THE CHRONICALLY OCCLUDED TIBIAL VESSEL, IT WAS NOTICED THAT 3MM OF THE TIP OF WIRE BROKE OFF IN THE HEAVILY CALCIFIED PORTION OF THE OCCLUSION. FURTHERMORE, THE DETACHED PORTION DID NOT AFFECT ARTERIAL FLOW OR CHANGE THE PATIENT'S CONDITION AND WAS LEFT IN PLACE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8010659 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page