Direxion™ M001195200Recalls & reported problems
Boston Scientific Corporation · Torqueable Microcatheter · Class 2
All models of Direxion (14)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729839316), received between Oct 19, 2018 and Mar 29, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
- Material Twisted/Bent1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 29, 2022Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking
IT WAS REPORTED THAT THE MICROCATHETER WAS FRACTURED. A DIREXION CATHETER-INFUSION WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT RESISTANCE WAS FELT AND THE DEVICE SNAPPED. THE MICROCATHETER WAS TAKEN OUT AND A FRACTURE WAS OBSERVED IN THE MIDDLE OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITH ALTERNATIVE METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13943748 ↗MalfunctionSep 24, 2021Material Twisted/Bent; Material Twisted/Bent
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 03SEP2021. IT WAS REPORTED THAT THE DEVICE WAS BENT. THE TARGET LESION WAS LOCATED IN THE ILIAC AREA. A DIREXION CATHETER-INFUSION WAS SELECTED FOR USE. DURING THE PREPARATION, THE CATHETER SURFACE WAS HYDRATED PRIOR TO REMOVAL FROM CARRIER HOOP. HOWEVER, IT WAS NOTED THAT THE DEVICE GOT BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE DEVICE WAS FRACTURED 25.5CM FROM THE HUB. IT WAS FURTHER REPORTED THAT THE FRACTURE WAS NOTED UPON FULLY REMOVING THE DEVICE FROM THE PATIENT'S BODY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12528450 ↗MalfunctionOct 19, 2018Break; Break
IT WAS REPORTED THAT THE CATHETER BROKE. A 021 STRAIGHT 1RO/105 DIREXION CATHETER WAS SELECTED FOR USE. DURING THE PROCEDURE, WHEN TRACKING THE DEVICE THROUGH THE ANATOMY, IT WAS NOTED THAT THE CORE OF THE NITINOL LAYER CRACKED AND SPLIT APART. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS POST-PROCEDURE WAS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7984766 ↗
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.