Direxion™ M001195280Recalls & reported problems
Boston Scientific Corporation · Torqueable Microcatheter · Class 2
All models of Direxion (14)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729839392), received between Oct 25, 2021 and Dec 20, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture2
- Entrapment of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 20, 2022Fracture; Fracture
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 06-DEC-2022. IT WAS REPORTED THAT A FRACTURE OCCURRED. TWO DIREXION CATHETERS WERE SELECTED FOR USE. DURING THE PROCEDURE, THE CATHETERS FRACTURED A FEW CENTIMETERS FROM THE HUB ON A PORTION OF THE DEVICE EXTERNAL TO THE PATIENT. THE PROCEDURE WAS CANCELLED DUE TO UNFAVORABLE ANATOMY. THERE WERE NO PATIENT COMPLICATIONS. HOWEVER, DEVICE ANALYSIS REVEALED A FRACTURE 32.7CM FROM THE HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16016083 ↗MalfunctionApr 1, 2022Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. TWO NGMC/FATHOM/021/BERN/1RO/130 DIREXION FATHOM-16 SYSTEM WERE SELECTED FOR USE. DURING THE PROCEDURE, THE WIRE GOT STUCK IN THE CATHETER WAS NOTED. A NEW J-SHAPE DIREXION CATHETER WAS OPENED, HOWEVER, THE SAME ISSUE OCCURRED. THE PROCEDURE WAS THEN ABORTED DUE TO THIS EVENT. THERE WERE NO COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13972261 ↗MalfunctionOct 25, 2021Fracture; Fracture
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 04OCT2021. IT WAS REPORTED THAT DEVICE FRACTURED AT THE HUB. A DIREXION CATHETER-INFUSION WAS SELECTED FOR USE. THE CATHETER SURFACE WAS HYDRATED PRIOR TO REMOVAL FROM CARRIER HOOP. HOWEVER, UPON REMOVING THE CATHETER FROM THE HOOP, THE NITINOL TUBING WAS DISCONNECTED FROM THE HUB DUE TO EXCESSIVE FORCE BY THE PHYSICIAN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE SHAFT WAS SEPARATED 30.5CM FROM THE HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12689607 ↗
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page