Skip to content

Direxion™ Fathom™-16 System M001195610Recalls & reported problems

Boston Scientific Corporation · Pre-Loaded Torqueable Microcatheter · Class 2

All models of Direxion Fathom-16 System (6)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729839606), received between Nov 27, 2018 and Nov 12, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 16, 2026Class IIOpen, Classified

    The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

    Z-0249-2027 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break2
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Difficult to Remove1
  • Fracture1
  • Material Deformation1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 12, 2025Break; Difficult to Remove; Adverse Event Without Identified Device or Use Problem; Break; Difficult to Remove; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE DEVICE WAS BROKEN. A DIREXION? FATHOM?-16 SYSTEM CATHETER INFUSION WAS SELECTED FOR USE. DURING THE PROCEDURE, RESISTANCE WAS ENCOUNTERED WHILE WITHDRAWING THE CATHETER. UPON REMOVAL, IT WAS OBSERVED THAT THE CATHETER HAD FRACTURED. THE PROCEDURE WAS COMPLETED USING WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23533621 ↗
  • MalfunctionAug 29, 2024Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation

    IT WAS REPORTED THAT THE CORE WIRE FRACTURED. A DIREXION INFUSION CATHETER KIT SUPPLIED WITH A FATHOM GUIDEWIRE WAS SELECTED FOR USE. DURING USE, IT WAS NOTED THAT THE LAST FEW FATHOM GUIDEWIRES SUPPLIED IN THE DIREXION CATHETER KIT LOST THEIR SHAPE QUICKLY, EVEN DURING THE FIRST RECONFIGURATION. IN ONE FATHOM GUIDEWIRE, IT RESULTED A FRACTURE OF THE CORE WIRE. THE PROCEDURE WAS COMPLETED WITH A NEW DEVICE, SAME MODEL. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20101444 ↗
  • MalfunctionOct 31, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE MICROCATHETER WAS FRACTURED. THE TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND NON-CALCIFIED VESSEL. A DIREXION FATHOM-16 SYSTEM WAS SELECTED FOR USE IN AN OVARIAN VEIN EMBOLIZATION PROCEDURE TO TREAT PELVIC CONGESTION SYNDROME. DURING THE PROCEDURE, IT WAS NOTED THAT RESISTANCE WAS FELT DURING COIL DEPLOYMENT. THE WHOLE MICROCATHETER WAS REMOVED AND A FRACTURE WAS FOUND HALFWAY DOWN THE MICROCATHETER. THE FRACTURED PORTION WAS IN THE SHEATH. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18041584 ↗
  • MalfunctionNov 27, 2018Break; Material Integrity Problem; Break; Material Integrity Problem

    IT WAS REPORTED THE HYDROPHILIC COVER OF THE DEVICE WAS DETACHED. A DIREXION FATHOM-16 SYSTEM WAS SELECTED FOR USE. IT WAS FOUND OUT THAT THERE WAS FAILURE OF PROXIMAL RUPTURE OF THE CATHETER WITH DETACHMENT OF THE HYDROPHILIC COVER. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8105859 ↗

FDA clearance and classification

The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Direxion™ Fathom™-16 System M001195610

Has the Direxion™ Fathom™-16 System M001195610 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0249-2027 (Class II), began on Sep 16, 2026: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Direxion™ Fathom™-16 System M001195610?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Nov 27, 2018 and Nov 12, 2025. In those reports, the most common specific device problems are “Break” (2), “Detachment of Device or Device Component” (1) and “Difficult to Remove” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Direxion™ Fathom™-16 System M001195610?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Direxion™ Fathom™-16 System M001195610, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Direxion™ Fathom™-16 System, model M001195610). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, continuous flush devices

All Catheter, continuous flush devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page