Direxion™ Fathom™-16 System M001195650Recalls & reported problems
Boston Scientific Corporation · Pre-Loaded Torqueable Microcatheter · Class 2
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729839644), received between Mar 26, 2020 and Feb 9, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Adverse Event Without Identified Device or Use Problem2
- Fracture2
- Break1
- Failure to Advance1
- Material Deformation1
- Material Integrity Problem1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 9, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT MICROCATHETER FRACTURE OCCURRED. A DIREXION? FATHOM?-16 SYSTEM CATHETER-INFUSION WAS SELECTED FOR USE. DURING THE PROCEDURE, THE MICROCATHETER WAS INTRODUCED INTO THE PATIENT. HOWEVER, THE MICROCATHETER FRACTURED INTO SEVERAL SEGMENTS. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24292568 ↗MalfunctionFeb 9, 2026Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT MICROCATHETER FRACTURE OCCURRED. A DIREXION? FATHOM?-16 SYSTEM CATHETER-INFUSION WAS SELECTED FOR USE. DURING THE PROCEDURE, THE MICROCATHETER WAS INTRODUCED INTO THE PATIENT. HOWEVER, THE MICROCATHETER FRACTURED INTO SEVERAL SEGMENTS. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24292549 ↗MalfunctionJan 13, 2026Fracture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DEVICE FRACTURE OCCURRED. A DIREXION? FATHOM?-16 SYSTEM WAS SELECTED FOR A COILING PROCEDURE TO TREAT AN ENDOLEAK. DURING THE PROCEDURE, HOWEVER, THE DIREXION FRACTURED. A NEW DIREXION? FATHOM?-16 SYSTEM WAS SELECTED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS EXPECTED TO FULLY RECOVER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24053468 ↗MalfunctionApr 10, 2024Break; Break
IT WAS REPORTED THAT THE SHAFT BROKE. A DIREXION FATHOM-16 SYSTEM WAS SELECTED FOR HEPATIC ARTERIAL CHEMOEMBOLIZATION OF THE RIGHT LIVER METASTASIS. ULTRASOUND GUIDED PUNCTURE OF THE LEFT DISTAL RADIAL ARTERY WAS PERFORMED FOLLOWED BY PLACEMENT OF A NON-BOSTON SCIENTIFIC (BSC) INTRODUCER. CATHETERIZATION OF THE SUPERIOR MESENTERIC ARTERY AND THEN THE RIGHT HEPATIC ARTERY WAS THEN PERFORMED USING A NON-BSC CATHETER AND A NON-BSC .035 GUIDEWIRE. DURING DISTAL MICROCATHETERIZATION OF THE RIGHT NETWORK WITH THE DIREXION FATHOM-16 SYSTEM, THE CATHETER BROKE INSIDE DURING A WITHDRAWAL ATTEMPT. THE ENTIRE SYSTEM HAD TO BE REMOVED. THE PROCEDURE WAS REPEATED WITH A NEW DIREXION MICROCATHETER WITH CHEMOEMBOLIZATION TREATMENT INJECTION. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19082118 ↗MalfunctionOct 5, 2020Fracture; Material Deformation; Material Integrity Problem; Physical Resistance/Sticking; Fracture; Material Deformation; Material Integrity Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT THE GUIDEWIRE COATING PEELED. A DIREXION FATHOM-16 SYSTEM CATHETER INFUSION WAS SELECTED FOR USE IN A HEPATOCARCINOMA EMBOLIZATION. DURING THE PREPARATION, THE DIREXION MICROCATHETER WITH A PRELOADED GUIDEWIRE WAS PASSED INSIDE THE COIL. THE GUIDEWIRE WAS REMOVED FROM THE MICROCATHETER AND PLACED IT IN SALINE. THE COIL CONTAINING THE MICROCATHETER WAS FLUSHED. THE MICROCATHETER WAS REMOVED. RESISTANCE WAS NOTED UPON INSERTION DUE TO THE BENDING OF THE MICROGUIDE AND WAS UNABLE TO ENTER THE DIREXION. IT WAS OBSERVED THAT THE GUIDEWIRE WAS NOT SMOOTH AND ROUGHNESS IN THE GUIDEWIRE'S TIP, AS IF THE HYROPHILIC COATING WAS PEELED OFF AFTER PROPER HYDRATION. LASTLY, THE DIREXION WAS PLACED BACK ON THE COIL. THE COIL WAS ADVANCED. HOWEVER, THE SHAFT WAS BENT AND FRACTURED EXPOSING THE PEBAX LAYER. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10626615 ↗MalfunctionMar 26, 2020Failure to Advance; Detachment of Device or Device Component; Failure to Advance; Detachment of Device or Device Component
IT WAS REPORTED THAT THE TIP OF THE GUIDEWIRE FRACTURED. A 155CM DIREXION FATHOM-16 SYSTEM WAS SELECTED FOR USE. DURING PREPARATION, RESISTANCE WAS NOTED UPON REMOVING THE FATHOM GUIDEWIRE FROM THE CATHETER AND UPON INSERTING IT AGAIN. HOWEVER, THE TIP OF THE GUIDEWIRE WAS NOTED TO BE FRACTURED UPON REMOVAL. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 9882722 ↗
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.