Direxion™ Fathom™-16 System M001195650Recalls & reported problems
Boston Scientific Corporation · Pre-Loaded Torqueable Microcatheter · Class 2
All models of Direxion Fathom-16 System (6)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729839644), received between Mar 26, 2020 and May 28, 2026 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Sep 16, 2026Class IIOpen, Classified
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Z-0252-2027 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Adverse Event Without Identified Device or Use Problem2
- Fracture2
- Break1
- Difficult to Remove1
- Failure to Advance1
- Material Deformation1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMay 28, 2026Difficult to Remove; Detachment of Device or Device Component
It was reported that the device fractured and detached, requiring intervention. \\A Direxion? Fathom?-16 System catheter was used for abdominal embolization to treat a hemorrhage. After injecting the embolization material into the target vessel, the catheter became stuck, could not be withdrawn from the body, and fractured during gentle traction. The outer nitinol segment broke at the proximal aspect, resulting in exposure of the inner component. The nitinol segment of the catheter remained inside the vessel, while the remaining portion of the catheter was removed. A Goose Neck snare was used and after several attempts, was able to retrieve the nitinol segment from the patient. The procedure was prolonged, but there were no complications and the patient fully recovered.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25318077 ↗MalfunctionFeb 9, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT MICROCATHETER FRACTURE OCCURRED. A DIREXION? FATHOM?-16 SYSTEM CATHETER-INFUSION WAS SELECTED FOR USE. DURING THE PROCEDURE, THE MICROCATHETER WAS INTRODUCED INTO THE PATIENT. HOWEVER, THE MICROCATHETER FRACTURED INTO SEVERAL SEGMENTS. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24292568 ↗MalfunctionFeb 9, 2026Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT MICROCATHETER FRACTURE OCCURRED. A DIREXION? FATHOM?-16 SYSTEM CATHETER-INFUSION WAS SELECTED FOR USE. DURING THE PROCEDURE, THE MICROCATHETER WAS INTRODUCED INTO THE PATIENT. HOWEVER, THE MICROCATHETER FRACTURED INTO SEVERAL SEGMENTS. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24292549 ↗MalfunctionJan 13, 2026Fracture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DEVICE FRACTURE OCCURRED. A DIREXION? FATHOM?-16 SYSTEM WAS SELECTED FOR A COILING PROCEDURE TO TREAT AN ENDOLEAK. DURING THE PROCEDURE, HOWEVER, THE DIREXION FRACTURED. A NEW DIREXION? FATHOM?-16 SYSTEM WAS SELECTED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS EXPECTED TO FULLY RECOVER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24053468 ↗MalfunctionApr 10, 2024Break; Break
IT WAS REPORTED THAT THE SHAFT BROKE. A DIREXION FATHOM-16 SYSTEM WAS SELECTED FOR HEPATIC ARTERIAL CHEMOEMBOLIZATION OF THE RIGHT LIVER METASTASIS. ULTRASOUND GUIDED PUNCTURE OF THE LEFT DISTAL RADIAL ARTERY WAS PERFORMED FOLLOWED BY PLACEMENT OF A NON-BOSTON SCIENTIFIC (BSC) INTRODUCER. CATHETERIZATION OF THE SUPERIOR MESENTERIC ARTERY AND THEN THE RIGHT HEPATIC ARTERY WAS THEN PERFORMED USING A NON-BSC CATHETER AND A NON-BSC .035 GUIDEWIRE. DURING DISTAL MICROCATHETERIZATION OF THE RIGHT NETWORK WITH THE DIREXION FATHOM-16 SYSTEM, THE CATHETER BROKE INSIDE DURING A WITHDRAWAL ATTEMPT. THE ENTIRE SYSTEM HAD TO BE REMOVED. THE PROCEDURE WAS REPEATED WITH A NEW DIREXION MICROCATHETER WITH CHEMOEMBOLIZATION TREATMENT INJECTION. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19082118 ↗MalfunctionOct 5, 2020Fracture; Material Deformation; Material Integrity Problem; Physical Resistance/Sticking; Fracture; Material Deformation; Material Integrity Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT THE GUIDEWIRE COATING PEELED. A DIREXION FATHOM-16 SYSTEM CATHETER INFUSION WAS SELECTED FOR USE IN A HEPATOCARCINOMA EMBOLIZATION. DURING THE PREPARATION, THE DIREXION MICROCATHETER WITH A PRELOADED GUIDEWIRE WAS PASSED INSIDE THE COIL. THE GUIDEWIRE WAS REMOVED FROM THE MICROCATHETER AND PLACED IT IN SALINE. THE COIL CONTAINING THE MICROCATHETER WAS FLUSHED. THE MICROCATHETER WAS REMOVED. RESISTANCE WAS NOTED UPON INSERTION DUE TO THE BENDING OF THE MICROGUIDE AND WAS UNABLE TO ENTER THE DIREXION. IT WAS OBSERVED THAT THE GUIDEWIRE WAS NOT SMOOTH AND ROUGHNESS IN THE GUIDEWIRE'S TIP, AS IF THE HYROPHILIC COATING WAS PEELED OFF AFTER PROPER HYDRATION. LASTLY, THE DIREXION WAS PLACED BACK ON THE COIL. THE COIL WAS ADVANCED. HOWEVER, THE SHAFT WAS BENT AND FRACTURED EXPOSING THE PEBAX LAYER. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10626615 ↗MalfunctionMar 26, 2020Failure to Advance; Detachment of Device or Device Component; Failure to Advance; Detachment of Device or Device Component
IT WAS REPORTED THAT THE TIP OF THE GUIDEWIRE FRACTURED. A 155CM DIREXION FATHOM-16 SYSTEM WAS SELECTED FOR USE. DURING PREPARATION, RESISTANCE WAS NOTED UPON REMOVING THE FATHOM GUIDEWIRE FROM THE CATHETER AND UPON INSERTING IT AGAIN. HOWEVER, THE TIP OF THE GUIDEWIRE WAS NOTED TO BE FRACTURED UPON REMOVAL. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 9882722 ↗
FDA clearance and classification
The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Direxion™ Fathom™-16 System M001195650
Has the Direxion™ Fathom™-16 System M001195650 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0252-2027 (Class II), began on Sep 16, 2026: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Direxion™ Fathom™-16 System M001195650?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Mar 26, 2020 and May 28, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (4), “Fracture” (2) and “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Direxion™ Fathom™-16 System M001195650?
None of the reports are classified as deaths, 1 as an injury and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Direxion™ Fathom™-16 System M001195650, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Direxion™ Fathom™-16 System, model M001195650). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths973 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths55 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page