Direxion™ Fathom™-16 System M001195670Recalls & reported problems
Boston Scientific Corporation · Pre-Loaded Torqueable Microcatheter · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729839668), received between Apr 7, 2023 and Apr 1, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 1, 2026Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DEVICE FRACTURED. A DIREXION FATHOM-16 SYSTEM WAS SELECTED FOR USE. HOWEVER, DURING THE PROCEDURE, IT WAS NOTED THAT THE DEVICE BECAME FRACTURED. THE DEVICE WAS REMOVED INTACT. THE PROCEDURE WAS COMPLETED WITH ANOTHER DIREXION FATHOM MICROCATHETER. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION WAS POSITIVE POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24756367 ↗MalfunctionOct 11, 2024Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A DIREXION FATHOM-16 SYSTEM WAS USE FOR GENICULAR ARTERY EMBOLIZATION(GAE). DURING THE PROCEDURE, THE DIREXION CATHETER WAS NOT FUNCTIONING AS EXPECTED. UPON REMOVAL, THE CATHETER WAS FOUND TO BE BROKEN ABOUT 5 INCHES FROM THE HUB. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20429057 ↗MalfunctionOct 1, 2024Break; Material Integrity Problem; Break; Material Integrity Problem
IT WAS REPORTED THAT INTERNAL CATHETER BREAK OCCURRED. THE NON- STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND NON-CALCIFIED LEFT KNEE VESSEL. A 021/BERN/2RO/130 DIREXION CATHETER INFUSION WAS SELECTED FOR USE. DURING THE EMBOLIZATION PROCEDURE, THE CATHETER FELT OBSTRUCTED AND WAS NO LONGER DELIVERING ANY BEADS. UPON REMOVAL OF THE MICROCATHETER, IT WAS OBSERVED THAT THE CATHETER WAS BROKEN ALONG THE SHAFT AND TWISTED, 6-7 INCHES DISTAL TO THE HUB. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20349015 ↗MalfunctionApr 7, 2023Fracture; Fracture
IT WAS REPORTED THAT THE CATHETER WAS FRACTURED. THE TARGET LESION WAS LOCATED IN THE SPLENIC ARTERY. A DIREXION FATHOM-16 SYSTEM WAS SELECTED FOR USE. DURING THE EMBOLIZATION PROCESS WITH A COIL, THE CATHETER WAS NOTED TO BE FRACTURED AFTER THE PROCEDURE ON SUPRA DISTAL. THE DEVICE WAS SIMPLY REMOVED WITHOUT ANY INTERVENTION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND PATIENT WAS STABLE POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16702004 ↗
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.