Direxion™ Transend™-14 System M001195810Recalls & reported problems
Boston Scientific Corporation · Pre-Loaded Torqueable Microcatheter · Class 2
All models of Direxion Transend-14 System (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729839743), received between Feb 23, 2021 and Oct 25, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Difficult to Advance1
- Difficult to Insert1
- Entrapment of Device1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 25, 2023Break; Difficult to Advance; Break; Difficult to Advance
IT WAS REPORTED THAT THE MICROCATHETER EXTERNAL SHEATH BROKE IN TWO PIECES. A DIREXION CATHETER WAS SELECTED FOR USE IN A RENAL EMBOLIZATION PROCEDURE TO TREAT A RENAL TUMOR. DURING THE PROCEDURE, WHEN ENTERING THE MICROCATHETER, THE EXTERNAL SHEATH WAS BROKEN IN TWO PIECES. THE PROBLEM WAS DISCOVERED BECAUSE IT WAS NOT POSSIBLE TO MOVE THE CATHETER FORWARD. THE MICROCATHETER WAS REMOVED OUT OF THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE METHOD. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18007166 ↗MalfunctionMay 23, 2022Difficult to Insert; Difficult to Insert
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 03MAY2022. IT WAS REPORTED THAT THE DEVICE COULD NOT TRACK ON DIAGNOSTIC CATHETER. VASCULAR ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY. THE TARGET LESION WAS LOCATED IN THE GASTROINTESTINAL AREA. A 130CM DIREXION TRANSEND-14 SYSTEM WAS SELECTED FOR USE. THE DEVICE WAS HYDRATED 15 SECONDS PRIOR TO REMOVAL OF THE HOOP. UPON INTRODUCTION, THE DEVICE COULD NOT TRACK ON DIAGNOSTIC CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO COMPLICATIONS WERE REPORTED AND PATIENT WAS STABLE POST PROCEDURE. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE DEVICE WAS FRACTURED 66.7CM FROM THE HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14475568 ↗MalfunctionSep 13, 2021Physical Resistance/Sticking; Physical Resistance/Sticking
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 25AUG2021. IT WAS REPORTED THAT DEVICE HAD RESISTANCE AND WAS NOT TRACKING. THE TARGET LESION WAS LOCATED IN THE FEMORAL ARTERY. A DIREXION TRANSEND -14 SYSTEM WAS SELECTED FOR USE. DURING THE PROCEDURE, UPON ADVANCING THE MICROCATHETER INTO THE LESION, THE DEVICE COULD NOT TRACK AND SHOWED RESISTANCE. THE DEVICE WAS COMPLETELY REMOVED ALONG WITH THE DIAGNOSTIC CATHETER. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE SHAFT FRACTURE 46CM FROM THE HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12464360 ↗MalfunctionFeb 23, 2021Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT RESISTANCE WAS NOTED WHEN WITHDRAWING THE CATHETER. THE 50% STENOSED TARGET LESION WAS LOCATED IN THE NOT SO TORTUOUS AND MILDLY CALCIFIED ULNAR ARTERY. A DIREXION TRANSEND-14 SYSTEM WAS SELECTED FOR USE. DELIVERY WAS MADE USING THE MICROCATHETER AND A CONTINUOUS SALINE FLUSH WAS MAINTAINED BETWEEN THE GUIDING CATHETER AND THE MICROCATHETER, AND BETWEEN THE MICROCATHETER AND THE GUIDEWIRE DURING THE PROCEDURE. DURING WITHDRAWAL, IT WAS NOTED THAT THERE WAS RESISTANCE IN WITHDRAWING THE DIREXION MICROCATHETER. THE DEVICE WAS REMOVED ALONG WITH THE 5F DIAGNOSTIC CATHETER AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. IT WAS NOTED THAT EXCESSIVE FORCE WAS NOT REQUIRED. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT WAS STABLE POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11366855 ↗
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.