Direxion™ HI-FLO™ Transend™-18 System M001195940Recalls & reported problems

Boston Scientific Corporation · Pre-Loaded Torqueable Microcatheter · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729839842), received between Sep 28, 2018 and May 14, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2026Fracture

    IT WAS REPORTED THAT SHAFT FRACTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE VESSEL IN THE LIVER. A NGMC/TRANSEND/027/BERN/1RO/130 DIREXION HI-FLO WAS SELECTED FOR USE. DURING THE PREPARATION, IT WAS NOTED THAT THE DEVICE WAS FRACTURED AND UNUSED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25174324 ↗
  • MalfunctionMar 28, 2024Fracture; Fracture

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A DIREXION HI-FLO TRANSEND-18 SYSTEM CATHETER-INFUSION WAS SELECTED FOR USE. DURING THE PROCEDURE, AN IDC-18 COIL WAS SUCCESSFULLY DEPLOYED. A SECOND NON-BSC COIL WAS SELECTED. HOWEVER, DURING ADVANCEMENT, RESISTANCE WAS ENCOUNTERED. IT WAS UNABLE TO PULL BACK OR DEPLOY THE COIL. UPON REMOVAL, THE CATHETER OUTER TUBE WAS FOUND BROKEN. THE INNER LINING WAS INTACT, THE CATHETER WAS WITHDRAWN IN ONE PIECE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18997054 ↗
  • MalfunctionSep 28, 2018Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE OUTER SHAFT OF THE CATHETER WAS BROKEN AND DETACHED. A DIREXION HI-FLO TRANSEND-18 SYSTEM WAS SELECTED FOR USE. DURING THE PROCEDURE, WHEN THE MICROCATHETER WAS INTRODUCED THROUGH A NON-BSC DEVICE WITH THE TRANSEND 18 GUIDEWIRE, THE OUTER SHAFT OF THE MICROCATHETER BROKE AND DETACHED. THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7918663 ↗

Other Catheter, continuous flush devices