SYNERGY™ H7493926012220Recalls & reported problems
Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant
All models of SYNERGY (74)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Synergy everolimus-eluting coronary stent is used to prop open narrowed heart arteries. Most of the 72 reports describe problems during placement — the stent or its shaft breaking, bending, or failing to cross tight, calcified lesions, and sometimes coming apart inside the patient — along with a smaller number mentioning clot formation, chest pain, heart attack, and6 deaths and 25 injuries. No recalls name this device, and no reports were filed in either of the past two 12-month periods.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 72
- Deaths reported
- 6
- Injuries reported
- 25
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 72 adverse event reports that list this exact device (DI 08714729840022), received between Nov 11, 2016 and Sep 20, 2023 — classified as 6 deaths, 25 injuries, 41 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (22 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated22
- Break16
- Failure to Advance9
- Bent7
- Adverse Event Without Identified Device or Use Problem6
- Device Damaged Prior to Use6
- Difficult to Remove5
- Obstruction of Flow5
Patient problems
- No Consequences Or Impact To Patient39
- No Patient Involvement8
- Thrombosis7
- Death6
- No Known Impact Or Consequence To Patient5
- Chest Pain4
- Device Embedded In Tissue or Plaque4
- Myocardial Infarction3
Recent adverse event reports
The 25 most recent of 72 reports, in the reporter's own words as published by the FDA.
InjurySep 20, 2023Obstruction of Flow; Obstruction of Flow
EVOLVE_LV_CLV CLINICAL STUD. IT WAS REPORTED THAT MYOCARDIAL INFARCTION AND IN-STENT RESTENOSIS (ISR) OCCURRED. ON (B)(6) 2021, THE SUBJECT PRESENTED WITH STABLE ANGINA AND WAS REFERRED FOR CARDIAC CATHETERIZATION. THE TARGET LESION WAS LOCATED IN THE PROXIMAL LEFT ANTERIOR DESCENDING ARTERY (LAD) WITH 80% STENOSIS AND WAS 16 MM LONG WITH A REFERENCE VESSEL DIAMETER OF 5.3 MM AND TIMI FLOW OF 3. THE TARGET LESION WAS TREATED WITH PRE-DILATION AND PLACEMENT OF A 4.00 MM X 20 MM STUDY STENT. FOLLOWING POST-DILATION, RESIDUAL STENOSIS WAS 0% WITH TIMI FLOW OF 3. IN ADDITION, A NON-TARGET LESION LOCATED IN MID LAD EXTENDING TO DISTAL LAD WAS TREATED WITH PLACEMENT OF A 2.50 MM X 16 MM SYNERGY DRUG ELUTING STENT (DES). A SECOND NON-TARGET LESION LOCATED IN 1ST DIAGONAL ARTERY WAS TREATED WITH A 2.25 MM X 12 MM SYNERGY DES. THE SUBJECT WAS DISCHARGED ON ASPIRIN AND CLOPIDOGREL. ON (B)(6) 2023, THE SUBJECT PRESENTED IN THE EMERGENCY DEPARTMENT (ED) WITH PRESSURE-LIKE SUBSTERNAL CHEST PAIN RADIATING TO BILATERAL ARMS AND LASTING 40 MINUTES IN DURATION WITH ASSOCIATED SHORTNESS OF BREATH. THE SUBJECT TOOK NITROGLYCERINE WHICH DID NOT PROVIDE COMPLETE PAIN RELIEF AND PARAMEDICS HAD GIVEN THE
Patient problems: Dyspnea; Myocardial Infarction; Restenosis
MAUDE report 17780482 ↗MalfunctionAug 23, 2021Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 2.25 X 12 SYNERGY II DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING THE PROCEDURE, THE MIDDLE OF THE SHAFT BROKE. NO PATIENT COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED THAT THE FRACTURE OCCURRED OUTSIDE THE PATIENT'S BODY AND THE DEVICE WAS SIMPLY PULLED FROM THE GUIDE CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12353599 ↗MalfunctionJul 8, 2021Break; Failure to Advance; Break; Failure to Advance
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 2.25 X 12 SYNERGY II DRUG ELUTING CORONARY STENT SYSTEM WAS ADVANCED TO THE LESION. THE SYNERGY STENT FAILED TO CROSS THE LESION. IT WAS OBSERVED THAT THE SHAFT BROKE. THE SYNERGY STENT WAS REMOVED. AND REMOVED. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO FURTHER COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12140421 ↗MalfunctionJan 7, 2021Defective Device; Defective Device
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2020. IT WAS REPORTED THAT STENT DEFECT WAS ENCOUNTERED. A 2.25 X 12 SYNERGY II DRUG-ELUTING STENT WAS SELECTED FOR USE. HOWEVER, BEFORE USING THE DEVICE INTO THE PATIENT, IT WAS NOTED THAT STENT HAS A DEFECT ON IT AND COULD NOT BE USED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A STENT DAMAGE.
Patient problems: No Patient Involvement
MAUDE report 11131729 ↗MalfunctionNov 18, 2020Failure to Advance; Failure to Advance
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 30-OCT-2020. IT WAS REPORTED THAT CROSSING DIFFICULTIES WERE ENCOUNTERED. THE TARGET LESION WAS LOCATED IN THE HIGHLY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.25 X 12 SYNERGY DRUG-ELUTING STENT WAS ADVANCED BUT FAILED TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED STENT DAMAGE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10860552 ↗MalfunctionOct 26, 2020Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 2.25 X 12MM SYNERGY DRUG-ELUTING STENT WAS SELECTED FOR USE. HOWEVER, THE MID SHAFT BROKE AS IT WAS BEING ADVANCED THROUGH THE GUIDE. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10735917 ↗MalfunctionSep 28, 2020Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT SHAFT FRACTURE AND STENT DAMAGE OCCURRED. A 2.25 X 12MM SYNERGY DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT IN A SIDE BRANCH OF A CORONARY VESSEL. AFTER AN UNSUCCESSFUL ATTEMPT TO DELIVER THE STENT, THE CATHETER SHAFT FRACTURED 7 TO 8 INCHES FROM THE HUB DURING REMOVAL. THE FRACTURED PORTION OF THE DEVICE WAS PULLED OUT WITH THE UNDEPLOYED DAMAGED STENT STILL ATTACHED. THE PROCEDURE WAS COMPLETED AND THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10597916 ↗InjurySep 15, 2020Failure to Advance; Failure to Advance
IT WAS REPORTED THAT THE PATIENT WAS EXPOSED TO EXCESSIVE RADIATION. THE 85% STENOSED TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND MILDLY CALCIFIED VESSEL, WITH A 90% BEND. FOLLOWING PRE-DILATION WITH A 2.0 X 12 EMERGE BALLOON, A 2.25 X 12 SYNERGY DRUG-ELUTING STENT WAS ADVANCED BUT FAILED TO CROSS THE LESION. THIS CAUSED A SIGNIFICANT DELAY OF THE PROCEDURE WHICH RESULTED TO EXCESSIVE RADIATION EXPOSURE OF THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Radiation Exposure, Unintended
MAUDE report 10536093 ↗MalfunctionAug 26, 2020Material Deformation; Material Deformation
IT WAS REPORTED THAT STENT DAMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND NON-CALCIFIED OSTIAL FIRST DIAGONAL ARTERY. AFTER PLACING A STENT IN THE LAD (LEFT ANTERIOR DESCENDING), A 2.25 X 12 SYNERGY II DRUG-ELUTING STENT WAS ADVANCED BUT COULD NOT CROSS LAD STENT. THE DEVICE WAS REMOVED AND NOTED A DAMAGED TO THE STENT STRUTS. IT WAS NOTED THAT IT WAS ASSUMED THE STENT HIT THE LAD STENT. THE LESION WAS THEN PRE-DILATED AND THE PROCEDURE WAS COMPLETED BY PLACING A NON-BSC STENT. NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10455688 ↗InjuryAug 25, 2020Difficult to Remove; Device Dislodged or Dislocated; Difficult to Remove; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT DIED. THE 50% STENOSED TARGET LESION WAS LOCATED IN THE NON-CALCIFIED AND SEVERELY TORTUOUS LEFT INTERNAL MAMMARY ARTERY TO DISTAL LEFT ANTERIOR DESCENDING ARTERY. A 2.25X16MM SYNERGY II DRUG-ELUTING STENT WAS ADVANCED BUT FAILED TO CROSS THE LESION. SUBSEQUENTLY, A 2.25X12MM SYNERGY II DRUG-ELUTING STENT WAS USED BUT HAD DIFFICULTY ADVANCING. THE DEVICE WAS ATTEMPTED TO REMOVE BUT THE STENT WAS DISLODGED AT THE ANASTOMOSIS. A 2.0X12MM EMERGE BALLOON CATHETER WAS ADVANCED AND DEPLOY THE STENT WITH TWO INFLATIONS AT 18 ATMOSPHERES FOR 16 SECONDS EACH. AFTERWARDS, A SLOW FLOW DOWN IN THE VESSEL WAS OBSERVED. NITROGLYCERIN WAS GIVEN TO RESTORE NORMAL BLOOD FLOW; HOWEVER, THE VESSEL STILL HAD SLOW FLOW. THE PHYSCIAN TRIED GOING BACK IN WITH AN INTERVENTIONAL WIRE TO BALLOON IT UP BUT IT ENDED LEAVING IT AS IS. THE PATIENT WAS TRANSFERRED TO THE FLOOR AND LATER IN THE AFTERNOON, THE PATIENT CODED AND EXPIRED LATER THE SAME NIGHT.
Patient problems: Death; No Code Available; Stenosis
MAUDE report 10449978 ↗MalfunctionAug 4, 2020Difficult to Advance; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED DIAGONAL BRANCH. AFTER PRE-DILATION WAS PERFORMED WITH A 2X12MM EMERGE BALLOON CATHETER. A 2.25 X 12 AND A 2.50 X 16 SYNERGY DRUG-ELUTING STENTS WERE SELECTED FOR USE; BUT, DURING INSERTION, THE STENTS LOSE WIRE ACCESS. THE STENTS WERE COMPLETELY REMOVED FROM THE PATIENT'S BODY. THE PHYSICIAN DECIDED NOT TO PROCEED AND ABORTED THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10362997 ↗MalfunctionJul 8, 2020Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE CIRCUMFLEX ARTERY. A 2.25 X 12 SYNERGY II DRUG-ELUTING STENT WAS SELECTED FOR USE. HOWEVER, WHILE ADVANCING THE STENT DELIVERY SYSTEM IN THE GUIDE CATHETER, THE SHAFT BROKE IN HALF MID-WAY. BOTH PIECES OF THE DELIVERY SYSTEM WAS RETRIEVED INTACT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10246484 ↗MalfunctionJun 10, 2020Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 2.25 X 12 SYNERGY DRUG-ELUTING STENT WAS ADVANCED BUT FAILED TO CROSS THE LESION AND THE SHAFT WAS BROKEN. NO KNOWN PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10139641 ↗InjuryMar 5, 2020Fracture; Fracture
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 2.25 X 12 SYNERGY II DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, IT WAS FOUND THAT THE STENT FRACTURED INSIDE THE PATIENT. THE PATIENT WAS TREATED BY IMPLANTING ANOTHER STENT OVER THE FRACTURED STENT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENTS STATUS IS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9794761 ↗MalfunctionFeb 14, 2020Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE CIRCUMFLEX ARTERY. A 2.25 X 12 SYNERGY DRUG-ELUTING STENT ADVANCED FOR TREATMENT. THE PHYSICIAN TRIED TO TWIST AND TORQUE THE DEVICE PASSED THE LESION BUT, WHEN THE DEVICE WAS REMOVED FROM THE PATIENT'S BODY, IT BROKE MID SHAFT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9712918 ↗InjuryOct 17, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE CIRCUMFLEX ARTERY. A 2.25 X 12 SYNERGY DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, WHILE TRYING TO CROSS THROUGH STENT STRUTS, THE STENT CAME OFF THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9206013 ↗InjurySep 12, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGMENT OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.25 X 12 SYNERGY DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, PRIOR TO DEPLOYMENT, THE STENT CAME OFF THE BALLOON. THE DISLODGED STENT WAS TRAPPED BY ANOTHER STENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER SYNERGY STENT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS FINE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9009958 ↗MalfunctionAug 19, 2019Material Deformation; Material Deformation
IT WAS REPORTED THAT STENT DAMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY. A 2.25 X 12MM SYNERGY II DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, WHEN TRYING TO PLACE THE STENT IN THE ARTERY, IT WAS OBSERVED THAT THE PROXIMAL END OF THE STENT WAS DAMAGED. THE STENT STAYED INTACT WITH THE BALLOON AND WAS REMOVED SAFELY FROM THE PATIENT'S BODY. THE PHYSICIAN FELT THAT THE STENT DAMAGE OCCURRED WHEN TRYING TO ADVANCE IT THROUGH A PREVIOUSLY PLACED STENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER 2.25 X 12MM SYNERGY II DRUG-ELUTING STENT WITH THE ASSISTANCE OF USING A GUIDEZILLA EXTENSION GUIDE CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8905559 ↗MalfunctionAug 2, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. A 2.25 X 12 SYNERGY DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, UPON LOADING THE STENT ONTO THE WIRE, IT WAS NOTED THAT THE STENT WAS STRIPPED FROM THE BALLOON. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT SERIOUS INJURY OR ADVERSE EVENT WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8854859 ↗InjuryMay 29, 2019Difficult to Remove; Device Dislodged or Dislocated; Difficult to Remove; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED AND RETRIEVAL DIFFICULTY WAS ENCOUNTERED. THE TARGET LESION WAS LOCATED IN THE PROXIMAL CIRCUMFLEX ARTERY. A 2.25 X 12 SYNERGY DRUG-ELUTING STENT WAS ADVANCED TO TREAT THE LESION. HOWEVER, THE STENT WAS DISLODGED IN THE OSTIAL CIRCUMFLEX AND IT WAS UNABLE TO BE RETRIEVED. THE PATIENT WAS SENT FOR BYPASS SURGERY. NO FURTHER PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8650555 ↗MalfunctionApr 19, 2019Failure to Advance; Failure to Advance
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 28MAR2019. IT WAS REPORTED THAT CROSSING DIFFICULTIES WERE ENCOUNTERED. A 2.25MM X 12MM SYNERGY DRUG-ELUTING STENT WAS ADVANCED BUT FAILED TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED STENT DAMAGE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8531450 ↗MalfunctionMar 26, 2019Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCOUNTERED. THE TARGET LESION WAS LOCATED IN A CORONARY ARTERY. A 2.25 X 12MM SYNERGY STENT WAS DEPLOYED FOR TREATMENT. HOWEVER, UPON REMOVAL IT WAS OBSERVED THAT THE STENT BALLOON WAS STUCK AND IT REQUIRES TOO MUCH FORCE TO REMOVE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8454110 ↗MalfunctionMar 26, 2019Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCOUNTERED. THE TARGET LESION WAS LOCATED IN A CORONARY ARTERY. A 2.25 X 12MM SYNERGY STENT WAS DEPLOYED FOR TREATMENT. HOWEVER, UPON REMOVAL IT WAS OBSERVED THAT THE STENT BALLOON WAS STUCK AND IT REQUIRES TOO MUCH FORCE TO REMOVE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8454103 ↗MalfunctionDec 17, 2018Failure to Advance; Failure to Advance
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 28-NOV-2018. IT WAS REPORTED THAT CROSSING DIFFICULTIES WERE ENCOUNTERED. THE TARGET LESION WAS LOCATED IN A CORONARY ARTERY. A 2.25 X 12MM SYNERGY II DRUG-ELUTING STENT WAS ADVANCED BUT FAILED TO CROSS THE LESION. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A HYPOTUBE DETACHED/SEPARATED. A SYNERGY II US MR 2.25 X 12MM STENT DELIVERY SYSTEM (SDS) WAS RETURNED FOR ANALYSIS. A VISUAL AND MICROSCOPIC EXAMINATION OF THE CRIMPED STENT FOUND NO ISSUES WITH THE STENT. THERE WERE NO SIGNS OF DAMAGE, STRETCHING OR LIFTING OF THE STENT STRUTS. THE STENT SHOWED NO SIGNS OF MOVEMENT AND WAS SET BETWEEN THE PROXIMAL AND DISTAL MARKER BANDS. THE CRIMPED STENT OUTER DIAMETER (OD) WAS MEASURED AND IS WITHIN MAX CRIMPED STENT PROFILE MEASUREMENT. THE BALLOON CONES WERE REVIEWED AND NO ISSUES WERE NOTED. THE BALLOON WINGS WERE TIGHTLY WRAPPED AND EVENLY FOLDED AND WERE NOT SUBJECTED TO POSITIVE PRESSURE. A VISUAL AND MICROSCOPIC EXAMINATION OF THE BUMPER TIP SHOWED NO DAMAGE. A VISUAL AND TACTILE EX
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8170037 ↗DeathDec 7, 2018Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT STENT THROMBOSIS AND DEATH OCCURRED. A 2.25 X 12MM SYNERGY II DRUG-ELUTING STENT (DES) WAS IMPLANTED IN THE POSTERIOR "DESCEDING" ARTERY (PDA). TWO STENTS WERE PLACED IN THE LEFT CIRCUMFLEX ARTERY, A 2.25 X 16MM SYNERGY II DES AND A 2.25 X 12MM SYNERGY II DES. TWENTY MINUTES AFTER THE PATIENT WAS TAKEN OFF THE TABLE, THE PATIENT EXPERIENCED CHEST PAIN. ELECTROCARDIOGRAPHY REVEALED STENT THROMBOSIS ON THE THREE IMPLANTED STENTS. THE PHYSICIAN PLACED A 2.25 X 8MM STENT DISTAL TO PREVIOUSLY IMPLANTED STENT IN THE PDA. THE FOLLOWING DAY, A NON-BSC TEMPORARY VENTRICULAR SUPPORT DEVICE WAS IMPLANTED IN THE PROXIMAL RIGHT CORONARY ARTERY BECAUSE THE PATIENT WAS IN SHOCK. AFTER ONE DAY, THE PATIENT DIED. THE PHYSICIAN BELIEVES THAT THE PATIENT HAD A NEUROLOGICAL EVENT PRIOR TO REMOVING THE NON-BSC DEVICE.
Patient problems: Chest Pain; Death; Thrombosis
MAUDE report 8143337 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page