Captivator EMR M00561610Recalls & reported problems

Boston Scientific Corporation · Endoscopic Mucosal Resection Device · Class 2

All models of Captivator EMR (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729842699), received between Feb 12, 2020 and Jun 22, 2026 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Failure to Fire1
  • Premature Activation1
  • Suction Problem1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Code Available1
  • No Consequences Or Impact To Patient1
  • Perforation1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 22, 2026Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A CAPTIVATOR EMR SNARE WAS USED FOR A GASTROSCOPY PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE THE SNARE BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25580052 ↗
  • MalfunctionOct 19, 2022Break; Failure to Fire; Break; Failure to Fire

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CAPTIVATOR EMR DEVICE WAS USED IN THE ESOPHAGUS DURING AN ENDOSCOPIC MUCOSAL RESECTION (EMR) PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, IT WAS NOTICED THAT THE SUTURE THREAD WAS BROKEN, AND THE BANDS COULD NOT BE RELEASED. THE PROCEDURE WAS COMPLETED WITH ANOTHER CAPTIVATOR EMR DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15632414 ↗
  • InjuryMar 1, 2022Use of Device Problem; Suction Problem; Use of Device Problem; Suction Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CAPTIVATOR ENDOSCOPIC MUCOSAL RESECTION DEVICE WAS USED DURING A ENDOSCOPIC MUCOSAL RESECTION (EMR) PROCEDURE PERFORMED ON (B)(6), 2022. WHEN THE DEVICE WAS OPENED THE BLUE CAP WAS DROPPED/LOST SO THE BLUE CAP WAS MISSING. THE TECHNICIAN USED THEIR FINGER TO COMPLETE SUCTION LOOP INSTEAD. THE DEVICE WAS SUCCESSFULLY USED IN THE ANTRUM. WHEN USED IN THE FUNDUS, THE BAND CAPTURED MUSCULARIS DUE TO TOO MUCH SUCTION AND THE BAND WAS DEPLOYED OVER THE WHOLE WALL AND WAS NOT REALIZED UNTIL AFTER IT WAS CUT, AND A PERFORATION OCCURRED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NON-BSC CLIP DEVICE. THE PATIENT OUTCOME WAS REPORTED TO BE FULLY RECOVERED. THERE WAS NO ADDITIONAL INTERVENTION REQUIRED AND THE PATIENT WAS NOT ADMITTED TO THE HOSPITAL BEYOND STANDARD OF CARE. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.

    Patient problems: Perforation

    MAUDE report 13639856 ↗
  • MalfunctionJun 16, 2021Premature Activation; Appropriate Device Problem Term/Code Not Available; Premature Activation; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CAPTIVATOR ENDOSCOPIC MUCOSAL RESECTION DEVICE WAS USED DURING AN ENDOSCOPIC MUCOSAL RESECTION (EMR) PROCEDURE PERFORMED ON (B)(6) 2021. DURING PREPARATION, IT WAS NOTED THAT TWO BANDS DEPLOYED WITH JUST ONE TURN. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. REPORTEDLY, THE PATIENT WAS ALREADY SEDATED WHEN THE PROCEDURE WAS CANCELLED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12008939 ↗
  • MalfunctionMay 4, 2020Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CAPTIVATOR ENDOSCOPIC MUCOSAL RESECTION DEVICE WAS USED DURING AN ENDOSCOPIC MUCOSAL RESECTION (EMR) PROCEDURE PERFORMED IN THE STOMACH ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING UNPACKING, THE SUTURE WAS FOUND CUT AND BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER CAPTIVATOR ENDOSCOPIC MUCOSAL RESECTION DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10020800 ↗
  • InjuryFeb 12, 2020Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CAPTIVATOR ENDOSCOPIC MUCOSAL RESECTION DEVICE WAS USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) WITH ENDOSCOPIC MUCOSAL RESECTION PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE SUTURE BROKE IN THE WORKING CHANNEL OF THE SCOPE. REPORTEDLY, THE BANDS FELL OFF THE DEVICE INSIDE OF THE PATIENT. TWO OF THE BANDS WERE RETRIEVED FROM THE BACK OF THE THROAT USING FORCEPS, AND THREE BANDS MIGRATED INTO THE STOMACH. THE SIXTH BAND COULD NOT BE LOCATED SO A BRONCHOSCOPE WAS USED TO CHECK THE LUNGS. THE BRONCHOSCOPY CONFIRMED THAT THE SIXTH BAND DID NOT MIGRATE TO THE PATIENT'S LUNGS. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WAS NO SERIOUS INJURY NOR ADVERSE PATIENT EFFECTS AS A RESULT OF THIS EVENT.

    Patient problems: No Code Available

    MAUDE report 9699203 ↗