Capio SLIM M0068318260Recalls & reported problems

Boston Scientific Corporation · Suture Capturing Device · Class 2

All models of Capio SLIM (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Boston Scientific Capio SLIM is a suture-placing device used in pelvic organ prolapse repair. Many of its 287 reports describe the device misfiring or failing to capture the suture, and some describe parts detaching. The reports include 48 injuries and no deaths. There are 3 recalls: 1 open, covering suture capture in some lots, and the rest closed, covering sutures that may break or detach.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
283
Deaths reported
0
Injuries reported
43
Recalls naming it
3

Summary

The FDA's MAUDE database contains 283 adverse event reports that list this exact device (DI 08714729842804), received between Nov 20, 2016 and Sep 16, 2026 — classified as 43 injuries, 240 malfunctions.

The most frequently reported specific device problem is “Retraction Problem” (86 reports).

3 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 2 reportsFeb 2024: 4 reportsMar 2024: 7 reportsApr 2024: 6 reportsMay 2024: 1 reportsJun 2024: 7 reportsJul 2024: 1 reportsAug 2024: 4 reportsSep 2024: 3 reportsOct 2024: 2 reportsNov 2024: 6 reportsDec 2024: 5 reportsJan 2025: 7 reportsFeb 2025: 6 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 6 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 3 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 4 reportsJan 2026: 6 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 3 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 27, 2024Class IIOpen, Classified

    Some lots of Capio SLIM do not consistently catch the suture after activation possibly resulting in a prolonged procedure or additional medical intervention.

    Z-2451-2024 · Root cause: Process change control · Recall details · FDA record ↗

  2. Feb 12, 2018Class IITerminated

    Potential for Capio sutures to break and /or detach.

    Z-1616-2018 · Root cause: Device Design · Recall details · FDA record ↗

  3. Feb 12, 2018Class IITerminated

    Potential for Capio sutures to break and /or detach.

    Z-1617-2018 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Retraction Problem86
  • Detachment of Device or Device Component70
  • Use of Device Problem53
  • Misfire48
  • Detachment Of Device Component40
  • Mechanical Problem15
  • Adverse Event Without Identified Device or Use Problem14
  • Break11

Patient problems

  • No Clinical Signs, Symptoms or Conditions144
  • No Consequences Or Impact To Patient55
  • Device Embedded In Tissue or Plaque32
  • No Known Impact Or Consequence To Patient23
  • Foreign Body In Patient18
  • Hemorrhage/Blood Loss/Bleeding5
  • No Patient Involvement5
  • Unspecified Tissue Injury4

Recent adverse event reports

The 25 most recent of 283 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Retraction Problem

    It was reported to Boston Scientific Corporation that, during a procedure using a Capio SLIM, the device appeared to become caught, causing the plunger to get stuck and not spring back. This occurred after the device had been fired twice. The procedure was completed using another of the same device, and no patient complications were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50788140 ↗
  • MalfunctionSep 8, 2026Detachment of Device or Device Component

    It was reported to Boston Scientific Corporation that, during the use of a Capio device, the dart detached from the suture during deployment. The suture was tensioned along the Capio handle. The break occurred at the dart. No attempt was made to recover the dart; however, an X-ray was performed in the operating room (OR). The procedure was completed using another of the same device. No patient complications were reported.

    Patient problems: Foreign Body In Patient

    MAUDE report 50667785 ↗
  • MalfunctionJul 8, 2026Misfire

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT, DURING USE OF A CAPIO SLIM DEVICE IN A PROLAPSE REPAIR PROCEDURE, THE DEVICE MISFIRED AND COULD NOT COMPLETE THE SUTURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25751106 ↗
  • MalfunctionJun 25, 2026Misfire

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT CAPIO SLIM WAS USED ON A SACROSPINOUS LIGAMENT REPAIR PROCEDURE. DURING THE PROCEDURE, TWO DEVICES MISFIRED. THE PROCEDURE WAS COMPLETED WITH ANOTHER SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED, AND THE PATIENT FULLY RECOVERED. NOTE: THIS MANUFACTURER REPORT PERTAINS TO THE FIRST OF TWO DEVICES USED DURING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25630307 ↗
  • MalfunctionJun 12, 2026

    TWO CAPIO SLIM DEVICES FOR SACROSPINAL LIGAMENT FIXATION WERE TOO STIFF AND NOT RETRIEVING BULLET/ FIRING PROPERLY. THIRD DEVICE WAS OPENED AND WORKED FINE. NO HARM TO THE PATIENT. MALFUNCTIONED DEVICES WERE TAKEN OFF THE FIELD. MANUFACTURER RESPONSE FOR INSTRUMENTATION, SURGICAL MESH, UROGYNECOLOGIC, TRANSVAGINAL REPAIR OF PELVIC OR, CAPIO SLIM (PER SITE REPORTER).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25484965 ↗
  • MalfunctionJun 3, 2026

    CAPIO DEVICE NOTED BY SURGEON TO BE MISFIRING. NO FURTHER DESCRIPTION FOUND IN OPERATIVE NOTE. MANUFACTURER RESPONSE FOR INSTRUMENTATION, SURGICAL MESH, UROGYNECOLOGIC, TRANSVAGINAL REPAIR OF PELVIC OR, CAPIO SLIM (PER SITE REPORTER). NOT KNOWN AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25374125 ↗
  • MalfunctionApr 9, 2026Use of Device Problem; Detachment of Device or Device Component; Difficult to Open or Close

    IT WAS REPORTED THAT A CAPIO SLIM DEVICE WAS USED DURING A VAULT SUSPENSION PROCEDURE ON (B)(6) 2026. DURING THE PROCEDURE, A NON-BOSTON SCIENTIFIC SUTURE DID NOT PENETRATE THE INTENDED LIGAMENT ON THE PATIENT'S RIGHT SIDE. THE SURGEON REPLACED THE SUTURE AND ATTEMPTED TO RETHROW IT. AT THAT TIME, THE SURGEON OBSERVED THAT THE DART WAS NO LONGER ATTACHED TO THE END OF THE SUTURE. THE DEVICE CAGE WAS OPENED BY BENDING THE METAL HOUSING, AT WHICH POINT THE DETACHED DART WAS LOCATED WITHIN THE DEVICE WITH A VERY SMALL PORTION OF SUTURE REMAINING ATTACHED (LESS THAN 1 CM). THE PROCEDURE WAS COMPLETED USING ANOTHER CAPIO SLIM WITH A DIFFERENT SUTURE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24834991 ↗
  • MalfunctionMar 20, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A SACROSPINOUS LIGAMENT FIXATION PROCEDURE USING A CAPIO SLIM DEVICE, THE BULLET DID NOT ENGAGE WITH THE RECEIVING MECHANISM AND DETACHED FROM THE SUTURE, RESULTING IN THE BULLET BEING RETAINED IN THE PATIENT. THE SUTURE BROKE ON THE RIGHT SIDE OF THE PATIENT RIGHT AT THE DART. ATTEMPTS WERE MADE TO RETRIEVE THE BULLET; HOWEVER, THESE ATTEMPTS WERE UNSUCCESSFUL. THE PROCEDURE WAS COMPLETED USING ANOTHER SUTURE WITH THE SAME CAPIO DEVICE, AND NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 24651549 ↗
  • MalfunctionMar 12, 2026Misfire

    IT WAS REPORTED THAT DURING THE SACROSPINOUS LIGAMENT PROCEDURE, TWO CAPIO SLIM DEVICES MISFIRED. THE PROCEDURE WAS COMPLETED USING ANOTHER CAPIO DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. NOTE: THIS REPORT PERTAINS TO THE SECOND OF TWO DEVICES USED DURING THE SAME PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24579433 ↗
  • MalfunctionMar 12, 2026Misfire

    IT WAS REPORTED THAT DURING THE SACROSPINOUS LIGAMENT PROCEDURE, TWO CAPIO SLIM DEVICES MISFIRED. THE PROCEDURE WAS COMPLETED USING ANOTHER CAPIO DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. NOTE: THIS REPORT PERTAINS TO THE FIRST OF TWO DEVICES USED DURING THE SAME PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24579407 ↗
  • MalfunctionJan 29, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING PROCEDURE, THE BULLET GOT STUCK IN THE DEVICE, REQUIRING THE USE OF INSTRUMENTS TO PRY OPEN THE DEVICE TO RETRIEVE THE SUTURE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24207331 ↗
  • MalfunctionJan 28, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING PROCEDURE, THE BULLET GOT STUCK IN THE DEVICE, REQUIRING THE USE OF INSTRUMENTS TO PRY OPEN THE DEVICE TO RETRIEVE THE SUTURE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24205646 ↗
  • MalfunctionJan 28, 2026Retraction Problem; Use of Device Problem

    IT WAS REPORTED THAT DURING A PROCEDURE USING A CAPIO SLIM DEVICE, THE HOOK DID NOT FIT INTO THE HOLE, AND THE HANDLE GOT STUCK AND FAILED TO MOVE SMOOTHLY. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24200247 ↗
  • MalfunctionJan 9, 2026Retraction Problem

    IT WAS REPORTED THAT, DURING A SACROSPINOUS LIGAMENT FIXATION PROCEDURE USING A CAPIO SLIM DEVICE, THE CARRIER FAILED TO RETRACT. AFTER MULTIPLE UNSUCCESSFUL ATTEMPTS TO THROW THE SUTURE USING A CAPIO DEVICE, THE PHYSICIAN DECIDED NOT TO USE THE CAPIO DEVICE TO PLACE THE FINAL SUTURE AND INSTEAD MANUALLY LOCATED THE SACROSPINOUS LIGAMENT AND PERFORMED TRADITIONAL (NON-DEVICE) SUTURING TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24030666 ↗
  • MalfunctionJan 9, 2026Use of Device Problem

    IT WAS REPORTED THAT DURING A SACROSPINOUS LIGAMENT FIXATION PROCEDURE FOR VAGINAL PROLAPSE USING A CAPIO SLIM, THE DEVICE WOULD NOT CAPTURE THE NEEDLE. A SECOND CAPIO SLIM WAS USED BUT IT ALSO FAILED TO CAPTURE THE NEEDLE. THE PROCEDURE WAS THEN CANCELLED. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION STATUS UNKNOWN. NOTE: THIS MANUFACTURER REPORT PERTAINS TO THE FIRST OF TWO DEVICES USED DURING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24028043 ↗
  • MalfunctionJan 9, 2026Use of Device Problem

    IT WAS REPORTED THAT DURING A SACROSPINOUS LIGAMENT FIXATION PROCEDURE FOR VAGINAL PROLAPSE USING A CAPIO SLIM, THE DEVICE WOULD NOT CAPTURE THE NEEDLE. A SECOND CAPIO SLIM WAS USED BUT IT ALSO FAILED TO CAPTURE THE NEEDLE. THE PROCEDURE WAS THEN CANCELLED. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER ANESTHESIA OR SEDATION STATUS UNKNOWN. NOTE: THIS MANUFACTURER REPORT PERTAINS TO THE SECOND OF TWO DEVICES USED DURING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24028029 ↗
  • MalfunctionDec 22, 2025Misfire; Misfire

    IT WAS REPORTED THAT DURING A SACROSPINOUS LIGAMENT FIXATION PROCEDURE USING A CAPIO SLIM DEVICE, THE DEVICE MISFIRED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23865926 ↗
  • MalfunctionDec 19, 2025Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED THAT UPON OPENING THE PACKAGING DURING PREPARATION, THERE WAS A HOLE OBSERVED IN THE LAST LAYER OF THE PACKAGING, COMPROMISING THE STERILITY OF THE PRODUCT. THE PROCEDURE WAS COMPLETED USING A SECOND DEVICE AND THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23856765 ↗
  • MalfunctionDec 9, 2025Misfire; Misfire

    IT WAS REPORTED THAT DURING THE SACROSPINOUS LIGAMENT PROCEDURE, TWO CAPIO SLIM DEVICES MISFIRED. THE PROCEDURE WAS COMPLETED USING ANOTHER CAPIO DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT NUMBER 2124215-2025-89001 FOR THE FIRST CAPIO SLIM DEVICE, AND 2124215-2025-89003 FOR THE SECOND CAPIO SLIM DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23759647 ↗
  • MalfunctionDec 9, 2025Misfire; Misfire

    IT WAS REPORTED THAT DURING THE SACROSPINOUS LIGAMENT PROCEDURE, TWO CAPIO SLIM DEVICES MISFIRED. THE PROCEDURE WAS COMPLETED USING ANOTHER CAPIO DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT NUMBER 2124215-2025-89001 FOR THE FIRST CAPIO SLIM DEVICE, AND 2124215-2025-89003 FOR THE SECOND CAPIO SLIM DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23759645 ↗
  • MalfunctionOct 31, 2025Misfire; Misfire

    IT WAS REPORTED THAT DURING THE PROCEDURE, TWO CAPIO DEVICES MISFIRED. THE PROCEDURE WAS COMPLETED USING A THIRD CAPIO DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT NUMBER 2124215-2025-77408 FOR THE FIRST CAPIO SLIM DEVICE, AND FOR THE SECOND CAPIO SLIM DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23441703 ↗
  • MalfunctionOct 31, 2025Misfire; Misfire

    IT WAS REPORTED THAT DURING THE PROCEDURE, TWO CAPIO DEVICES MISFIRED. THE PROCEDURE WAS COMPLETED USING A THIRD CAPIO DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT NUMBER 2124215-2025-77408 FOR THE FIRST CAPIO SLIM DEVICE, AND 2124215-2025-77421 FOR THE SECOND CAPIO SLIM DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23441696 ↗
  • MalfunctionSep 16, 2025Misfire; Misfire

    IT WAS REPORTED THAT, DURING A SACROSPINOUS LIGAMENT FIXATION PROCEDURE USING A CAPIO SLIM, THE DEVICE MISFIRED. NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23062701 ↗
  • MalfunctionSep 2, 2025Break; Break

    IT WAS REPORTED THAT DURING A SACROSPINOUS FIXATION PROCEDURE, THE DEVICE WAS NOT WORKING PROPERLY. THE REPORTED FAILURE MODE SUGGESTS BREAK/FRACTURE, THOUGH IT IS UNCERTAIN WHETHER THERE WAS A COMPLETE DETACHMENT. THE REPORTED EVENT MOST LIKELY INDICATES THAT THE CARRIER BROKE. NO PATIENT COMPLICATIONS WERE REPORTED. NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING A PROCEDURE. REFER TO MANUFACTURER REPORT NUMBER 2124215-2025-59160 FOR THE SECOND CAPIO SLIM DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22937896 ↗
  • MalfunctionSep 2, 2025Break; Break

    IT WAS REPORTED THAT DURING A SACROSPINOUS FIXATION PROCEDURE, THE DEVICE WAS NOT WORKING PROPERLY. THE REPORTED FAILURE MODE SUGGESTS BREAK/FRACTURE, THOUGH IT IS UNCERTAIN WHETHER THERE WAS A COMPLETE DETACHMENT. THE REPORTED EVENT MOST LIKELY INDICATES THAT THE CARRIER BROKE. NO PATIENT COMPLICATIONS WERE REPORTED. NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE A PROCEDURE. REFER TO MANUFACTURER REPORT NUMBER 2124215-2025-59042 FOR THE FIRST CAPIO SLIM DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22937227 ↗

FDA clearance and classification

The FDA regulates product code PWI (instrumentation, surgical mesh, urogynecologic, transvaginal repair of pelvic organ prolapse) under 21 CFR 884.4910 as a Class II device. 21 CFR 884.4910 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: Used to aid in insertion, placement, fixation, or anchoring of surgical mesh for transvaginal pelvic organ prolapse repair

Common questions about the Capio SLIM M0068318260

Has the Capio SLIM M0068318260 been recalled?

Yes. 3 FDA recall records name this device's identifier: 1 still open and 2 closed (terminated) by the FDA. The most recent, Z-2451-2024 (Class II), began on Jun 27, 2024: Some lots of Capio SLIM do not consistently catch the suture after activation possibly resulting in a prolonged procedure or additional medical intervention. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Capio SLIM M0068318260?

The FDA's MAUDE database lists 283 adverse event reports for this device, received between Nov 20, 2016 and Sep 16, 2026. In those reports, the most common specific device problems are “Retraction Problem” (86), “Detachment of Device or Device Component” (70) and “Use of Device Problem” (53); the most common patient problems are “Device Embedded In Tissue or Plaque” (32), “Foreign Body In Patient” (18) and “Hemorrhage/Blood Loss/Bleeding” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Capio SLIM M0068318260?

None of the reports are classified as deaths, 43 as injuries and 240 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Capio SLIM M0068318260, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Capio SLIM, model M0068318260). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Instrumentation, Surgical Mesh, Urogynecologic, Transvaginal Repair Of Pelvic Organ Prolapse devices

All Instrumentation, Surgical Mesh, Urogynecologic, Transvaginal Repair Of Pelvic Organ Prolapse devices →