Skip to content

REBEL™ H7493927424350Recalls & reported problems

Boston Scientific Corporation · Platinum Chromium Coronary Stent System · Class 3 · Implant

All models of REBEL (24)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729843931), received between Nov 29, 2018 and Nov 29, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

1 other recall cover devices cleared under the same FDA submission (P130030); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Material Deformation1

Patient problems

  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 29, 2018Material Deformation; Adverse Event Without Identified Device or Use Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT STENT DAMAGE OCCURRED. DURING PREPARATION OF A 24X3.50MM REBEL STENT, IT WAS NOTED THAT THE STENT WAS DAMAGED. NO PATIENT COMPLICATIONS WERE REPORTED AS THE DEVICE DID NOT ENTER THE PATIENT'S BODY. DEVICE EVALUATED BY MFR.: THE OUTER SHAFT, INNER SHAFT, BALLOON, STENT AND TIP WERE MICROSCOPICALLY EXAMINED. THERE ARE NUMEROUS SHAFT KINKS. THERE IS SHAFT DAMAGE AT THE EXIT NOTCH WHICH IS CONSISTENT WITH DAMAGE SEEN WITH THE USE OF A GUIDEWIRE. THE STENT IS DAMAGED 10MM FROM THE PROXIMAL MARKERBAND. INSPECTION OF THE REMAINDER OF THE DEVICE PRESENTED NO OTHER DAMAGE OR IRREGULARITIES. THERE WAS NO EVIDENCE OF ANY MATERIAL OR MANUFACTURING DEFICIENCIES CONTRIBUTING TO THE DAMAGE AND THE REPORTED EVENT.

    Patient problems: No Consequences Or Impact To Patient; No Patient Involvement

    MAUDE report 8115970 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P130030. One submission often covers many models, so check the product and lot details in each record.

  • Z-0580-2020 · Oct 24, 2019 · Terminated
    Rebel Monorail PtCr Coronary Stent System, REF / Catalog No. H7493927324300, Rebel MR OUS 3.0mm x 24mm. The Rebel Monorail (MR) Stent System is a balloon expandable Platinum Chro

Other STENT, CORONARY devices

All STENT, CORONARY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page