Promus PREMIER™ H7493952824300Recalls & reported problems
Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant
All models of Promus PREMIER (56)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729844853), received between Mar 16, 2017 and May 15, 2018 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Bent” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Bent3
- Break1
- Device Damaged Prior to Use1
- Device Damaged by Another Device1
- Device Dislodged or Dislocated1
- Material Perforation1
- Material Rupture1
Patient problems
- No Consequences Or Impact To Patient6
- No Patient Involvement1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 15, 2018Bent; Bent
IT WAS REPORTED THAT STENT DAMAGE OCCURRED. A 3.00X24MM PROMUS PREMIER¿ DRUG-ELUTING STENT WAS SELECTED FOR USE TO TREAT THE LESION. HOWEVER, IT WAS NOTICED THAT THE TIP OF THE STENT SEEMED TO BE DAMAGED AND THE STRUTS WERE NOT IN PLACED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7513847 ↗MalfunctionMay 9, 2018Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS IN THE LEFT CORONARY VESSEL. A 3.00X24MM PROMUS PREMIER¿ STENT WAS ADVANCED TO TREAT THE LESION. HOWEVER, THE DEVICE FAILED TO CROSS THE LESION AND IT WAS NOTED THAT THE SHAFT BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7500360 ↗MalfunctionDec 12, 2017Material Perforation; Device Damaged by Another Device; Material Perforation; Device Damaged by Another Device
IT WAS REPORTED THAT SHAFT PERFORATION OCCURRED. AFTER A GUIDE WIRE CROSSED THE LESION, A 3.00X24MM PROMUS PREMIER¿ DRUG-ELUTING STENT (DES) WAS ADVANCED TO TREAT THE LESION. THE GUIDE WIRE WAS INSERTED AT THE DISTAL END OF THE STENT. HOWEVER, IT WAS NOTED THAT THE GUIDE WIRE EXITED DISTAL TO THE PROXIMAL WIRE LUMEN PORT AND THAT THE PHYSICIAN PERFORATED THE WIRE LUMEN OF THE STENT WITH THE GUIDE WIRE BEING USED. THE STENT WAS REMOVED WITH NO PATIENT HARM. A NEW 3.00X24MM PROMUS PREMIER¿ DES WAS USED ON THE SAME WIRE AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS DISCHARGED THE FOLLOWING DAY WITH GREAT PROGNOSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7110234 ↗MalfunctionNov 8, 2017Bent; Bent
IT WAS REPORTED STENT DAMAGE OCCURRED. A 3.00X24 MM PROMUS PREMIER¿ STENT WAS SELECTED TO TREAT THE LESION. HOWEVER, A BENT STENT STRUT WAS NOTED UPON LOADING THE DEVICE ONTO THE WIRE. THE DEVICE DID NOT ENTER THE PATIENT'S BODY. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7009665 ↗MalfunctionAug 31, 2017Material Rupture; Material Rupture
SAME CASE AS MDR ID# 2134265-2017-08692. IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 3.00X24MM PROMUS PREMIER¿ DRUG-ELUTING STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING THE PROCEDURE, THE DISTAL TIP OF THE BALLOON RUPTURED AND THE STENT WAS NOT IMPLANTED. SUBSEQUENTLY, A 2.25X12MM PROMUS PREMIER¿ DRUG-ELUTING STENT WAS SELECTED TO TREAT A DIFFERENT VESSEL. HOWEVER, DURING UNPACKING, IT WAS NOTED THAT THE STENT APPEARED TO BE DAMAGED AND WAS NOT CRIMPED WELL ON THE BALLOON. THE DEVICE NEVER ENTERED THE PATIENT'S BODY AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6833939 ↗InjuryApr 4, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE PROXIMAL LEFT ANTERIOR DESCENDING (LAD) ARTERY. A 3.00X24MM PROMUS PREMIER¿ DRUG-ELUTING STENT WAS ADVANCED TO TREAT THE LESION. HOWEVER, DURING THE PROCEDURE, THE STENT CAME OFF THE BALLOON. A BALLOON CATHETER WAS THEN ADVANCED INTO THE STENT AND THE STENT WAS DEPLOYED WHERE IT WAS AND THE PROCEDURE WAS ENDED. LATER THAT EVENING, THE PATIENT WAS BROUGHT BACK TO THE CATH LAB AND ANOTHER STENT WAS IMPLANTED IN THE INITIALLY INTENDED LOCATION AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6457345 ↗MalfunctionMar 16, 2017Bent; Device Damaged Prior to Use; Bent; Device Damaged Prior to Use
IT WAS REPORTED THAT STENT DAMAGE OCCURRED. A 3.00X24MM PROMUS PREMIER¿ DRUG-ELUTING STENT WAS SELECTED FOR USE TO TREAT THE LESION. DURING UNPACKING, WHEN THE PHYSICIAN PULLED THE DEVICE ABOUT 2/3 OUT OF THE HOOP, IT WAS NOTICED THAT THE DEVICE HAD NO LESS THAN FIVE BENDS IN THE METAL TUBING. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Patient Involvement
MAUDE report 6409615 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page