Promus PREMIER™ H7493952832250Recalls & reported problems
Boston Scientific Corporation · Everolimus-Eluting Platinum Chromium Coronary Stent System · Class 3 · Implant
All models of Promus PREMIER (56)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729844952), received between Nov 21, 2016 and Dec 16, 2019 — classified as 1 death, 3 malfunctions.
The most frequently reported specific device problem is “Entrapment of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Entrapment of Device2
- Bent1
- Break1
- Defective Device1
- Fracture1
Patient problems
- No Consequences Or Impact To Patient2
- Death1
- No Patient Involvement1
- Thrombosis1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
DeathDec 16, 2019Defective Device; Defective Device
SAME CASE AS MDR: (B)(4) AND (B)(4). IT WAS REPORTED THAT STENT PROFILE PROBLEM OCCURRED AND PATIENT EXPERIENCED ACUTE THROMBOSIS AND DIED. THE HIGHLY CALCIFIED TARGET LESION WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY (LAD). AFTER CROSSING THE LESION WITH AN UNSPECIFIED GUIDEWIRE, TWO UNSPECIFIED BALLOON CATHETERS WERE ADVANCED WITH DIFFICULTY DUE TO THE CALCIFICATION. A 2.5X38MM PROMUS PREMIER DRUG-ELUTING STENT (DES) WAS IMPLANTED IN THE MID LAD. FOLLOWING IMPLANT, A DISTAL DISSECTION WAS NOTED ALONG WITH 18MM OF DISEASE THAT ALSO NEEDED TO BE TREATED. A 3.5X32MM PROMUS PREMIER DES WAS PROXIMALLY OVERLAPPED THE INITIALLY PLACED STENT. ANOTHER 2.5X32MM PROMUS PREMIER WAS ADVANCED DISTALLY THROUGH BOTH STENTS BUT WOULD NOT DELIVER DESPITE PUSHING HARD. THE DEVICE WAS REMOVED AND THE LESION WAS DILATED WITH AN UNSPECIFIED BALLOON CATHETER. THE 2.5X32MM STENT WAS READVANCED BUT WAS STILL UNSUCCESSFUL. THE DEVICE WAS REMOVED HOWEVER, THE STENT STARTED STRIPPING OFF THE BALLOON AND ACCORDIONED INSIDE THE PREVIOUS DEPLOYED STENTS. THE PHYSICIAN ATTEMPTED TO REMOVED THE DISLODGED STENT THAT BECAME STUCK IN THE PROXIMAL LAD. THE ARTERY STARTED TO CLOT AFTER 30 MINUTES AND A VENTRI
Patient problems: Death; Thrombosis
MAUDE report 9474802 ↗MalfunctionNov 1, 2017Entrapment of Device; Fracture; Entrapment of Device; Fracture
SAME CASE AS MDR ID: 2134265-2017-10552. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION TO TREAT A CHRONIC TOTAL OCCLUSION, THE STENT DELIVERY SYSTEM STUCK TO THE GUIDE WIRE AND THE CATHETER SHAFT BROKE. THE CALCIFIED TARGET LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY. THE PHYSICIAN WAS ABLE TO CROSS THE LESION AND DILATE SEVERAL TIMES WITH A BALLOON. A 300CM STRAIGHT TIP SAMURAI GUIDE WIRE WAS PLACED IN THE DISTAL VESSEL. A 2.50 X 32MM PROMUS PREMIER¿ DRUG-ELUTING STENT WAS ADVANCED, BUT FAILED TO CROSS THE LESION. UPON REMOVAL FROM THE BODY, THE STENT DELIVERY SYSTEM BECAME STUCK ON THE PROXIMAL PORTION OF THE GUIDE WIRE. WHILE TRYING TO REMOVE THE STENT DELIVERY SYSTEM FROM THE WIRE, THE CATHETER SHAFT SEPARATED AT THE WIRE EXIT PORT. ALL COMPONENTS WERE OUTSIDE OF THE BODY; NO DEVICE FRAGMENTS SEPARATED WITHIN THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS AND THE LESION WAS SUCCESSFULLY STENTED WITH A NEW 2.5 X 32MM STENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6993212 ↗MalfunctionApr 18, 2017Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN A CORONARY GRAFT TO THE LEFT ANTERIOR DESCENDING (LAD) ARTERY. A 190 CM FILTERWIRE EZ¿ WAS ADVANCED TO TREAT THE LESION. HOWEVER, DURING PROCEDURE, THE FILTERWIRE FAILED TO CROSS THE LESION. THE DEVICE WAS THEN REMOVED AND IT WAS NOTED THAT THE TIP WAS DAMAGED. SUBSEQUENTLY, ANOTHER FILTERWIRE WAS ADVANCED AND WAS ABLE TO TREAT THE LESION. A 2.50X32MM PROMUS PREMIER¿ STENT WAS ALSO ADVANCED TO TREAT THE LAD PORTION. MOREOVER, THE SHAFT BROKE IN HALF. THE DEVICE WAS REMOVED FROM THE PATIENT'S BODY AND THE PROCEDURE WAS COMPLETED WITH ANOTHER PROMUS PREMIER¿ STENT. NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS WAS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6500930 ↗MalfunctionNov 21, 2016Bent; Entrapment of Device; Bent; Entrapment of Device
IT WAS REPORTED THAT STENT DAMAGE AND CATHETER ENTRAPMENT OCCURRED. THE TARGET LESION WAS LOCATED BELOW THE KNEE VESSEL. A 2.50X32MM PROMUS PREMIER¿ DRUG-ELUTING STENT WAS SELECTED FOR USE TO TREAT THE LESION. HOWEVER, AS THE STENT WAS LOADED ONTO THE WIRE, OUTSIDE OF THE BODY, IT WOULD NO LONGER ADVANCE ON THE WIRE. THE PROXIMAL END OF THE STENT CRUMPLED ON THE BALLOON. THE PHYSICIAN ATTEMPTED TO REMOVE THE ENTIRE DELIVERY SYSTEM FROM THE WIRE AS IT HAD NOT ENTERED THE PATIENT¿S BODY BUT WAS UNABLE TO. THE ENTIRE STENT DELIVERY SYSTEM WAS REMOVED FROM THE WIRE AFTER A BLADE WAS USED TO STRIP THE STENT OFF THE BALLOON. THE PROCEDURE WAS COMPLETED WITH A NEW PROMUS PREMIER STENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Patient Involvement
MAUDE report 6118460 ↗
Other Coronary drug-eluting stent devices
- SYNERGY™ H7493926038300Boston Scientific Corporation · Coronary drug-eluting stent8 deaths124 reports
- SYNERGY™ H7493926016300Boston Scientific Corporation · Coronary drug-eluting stent7 deaths106 reports
- SYNERGY™ H7493926020300Boston Scientific Corporation · Coronary drug-eluting stent9 deaths90 reports
- SYNERGY™ H7493926016250Boston Scientific Corporation · Coronary drug-eluting stent6 deaths82 reports
- XIENCE Sierra™ 1550350-33Abbott Vascular Inc. · Coronary drug-eluting stent1 recall32 reports
- Resolute Onyx™ RONYX30022UXMedtronic, Inc. · Coronary drug-eluting stent6 deaths76 reports
- SYNERGY™ H7493926024300Boston Scientific Corporation · Coronary drug-eluting stent3 deaths75 reports
- Resolute Onyx™ RONYX30015UXMedtronic, Inc. · Coronary drug-eluting stent3 deaths73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page