NC Emerge® H7493926708250Recalls & reported problems
Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2
All models of NC Emerge (61)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729846246), received between Sep 14, 2020 and Mar 3, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Material Integrity Problem2
- Detachment of Device or Device Component1
- Failure to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 3, 2025Break; Failure to Advance; Break; Failure to Advance
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 8MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, RESISTANCE WAS ENCOUNTERED AS THE DEVICE WAS ADVANCED THROUGH THE LESION. WHILE TRYING TO PUSH THE DEVICE, THE SHAFT BROKE IN HALF. THE DISTAL PORTION OF THE SHAFT THAT REMAINED ON THE GUIDEWIRE WAS BEING CAPTURED BY THE GUIDE CATHETER. EVERYTHING WAS REMOVED SUCCESSFULLY, AND NEW EQUIPMENT WAS THEN USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED, AND NO HARM WAS DONE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21504422 ↗MalfunctionFeb 1, 2024Detachment of Device or Device Component; Material Integrity Problem; Detachment of Device or Device Component; Material Integrity Problem
IT WAS REPORTED THAT DEVICE DETACHMENT OCCURRED. THE PATIENT PRESENTED WITH CORONARY ARTERY DISEASE AND UNDERWENT PERCUTANEOUS CORONARY INTERVENTION. A 2.50MM X 8MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING INSERTION OF THE BALLOON CATHETER THROUGH THE HEMOSTASIS VALVE, THE BALLOON HUB WAS NOTED TO BE BENT AND FRACTURED. THE TECHNICIAN TRIED THE BEND THE HUB BACK; HOWEVER, IT SNAPPED OFF THE CATHETER RIGHT AT THE HUB. THE DEVICE WAS REMOVED FROM THE HEMOSTASIS VALVE WITH NO OTHER ISSUES. A 2.5 X 8 EMERGE BALLOON CATHETER WAS USED, AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18627505 ↗MalfunctionAug 8, 2022Material Integrity Problem; Material Integrity Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2022. IT WAS REPORTED THAT TIP DAMAGE OCCURRED. THE PATIENT PRESENTED WITH CORONARY ARTERY DISEASE. A 2.50MM X 8MM NC EMERGE BALLOON CATHETER WAS SELECTED FOR USE IN A PERCUTANEOUS CORONARY INTERVENTION. HOWEVER, DURING PREPARATION, THE CATHETER TIP WAS BENT AND DAMAGED. ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER; RETURNED DEVICE ANALYSIS REVEALED SHAFT BREAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15186271 ↗MalfunctionSep 13, 2021Break; Break
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 01SEP2021. IT WAS REPORTED THAT SHAFT BREAK NEAR THE HUB OCCURRED. DURING THE INTRODUCTION OF A 2.50MM X 8MM NC EMERGE BALLOON CATHETER, THE SHAFT BROKE OFF NEAR THE HUB. THE DETACHED PORTION WAS PULLED OUT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A SHAFT BREAK 19CM FROM THE STRAIN RELIEF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12463348 ↗MalfunctionSep 14, 2020Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 2.50MM X 8MM NC EMERGE BALLOON CATHETER WAS SELECTED FOR USE. HOWEVER, BEFORE THE DEVICE WAS PUT INTO THE BODY, IT WAS NOTED THAT THE SHAFT WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10533028 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page