Skip to content

NC Emerge® H7493926708320Recalls & reported problems

Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2

All models of NC Emerge (61)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729846277), received between Feb 8, 2019 and Jan 5, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break2
  • Material Deformation1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 5, 2023Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 3.25MM X 8MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR USE. HOWEVER, DURING INSERTION, THE SHAFT WAS KINKED AND WAS BROKE AT THE MIDDLE PART OUTSIDE THE BODY. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16107109 ↗
  • MalfunctionJun 27, 2022Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 3.25MM X 8MM NC EMERGE BALLOON CATHETER WAS SELECTED FOR USE. HOWEVER, DURING PREPARATION, THE BALLOON SHAFT BROKE INTO TWO PIECES. THE DEVICE WAS NOT INSERTED INTO THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH A NEW NC EMERGE BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14830361 ↗
  • MalfunctionFeb 8, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURED. A 3.25 X 8 NC EMERGE BALLOON WAS ADVANCED FOR TREATMENT. HOWEVER, AT 22 ATMOSPHERES THE BALLOON RUPTURED. NO PATIENT COMPLICATION INDICATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8321817 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page