Skip to content

NC Emerge® H7493926708350Recalls & reported problems

Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2

All models of NC Emerge (61)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729846284), received between Oct 13, 2021 and Feb 19, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Material Rupture” (3 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 24, 2019Class IITerminated

    Certain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process.

    Z-0579-2020 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Rupture3
  • Detachment of Device or Device Component2
  • Break1
  • Difficult to Remove1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 19, 2025Difficult to Remove; Material Rupture; Detachment of Device or Device Component; Difficult to Remove; Material Rupture; Detachment of Device or Device Component

    IT WAS REPORTED VIA USER MEDWATCH (B)(4) THAT BALLOON RUPTURE OCCURRED REQUIRING ADDITIONAL INTERVENTION. THE IN-STENT RESTENOSIS LESIONS WERE LOCATED IN THE RIGHT CORONARY ARTERY (RCA). A 3.50MM X 8MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATION. A PROLONGED HIGH-PRESSURE INFLATION WAS PERFORMED AND THE BALLOON RUPTURED. THERE WAS DIFFICULTY RETRIEVING THE BALLOON INTO THE NON-BOSTON SCIENTIFIC (BSC) GUIDE CATHETER. A GUIDE EXTENSION WAS ADVANCED OVER THE BALLOON TO REMOVE THE BALLOON, GUIDEWIRE, AND NON-BSC GUIDE CATHETER AS ONE UNIT. HOWEVER, THIS LED TO BALLOON AVULSION OFF OF THE SHAFT OF THE BALLOON DELIVERY SYSTEM. THE BALLOON FRAGMENT SEEN IN THE PROXIMAL RCA STENT SEGMENT. SNARING WAS ATTEMPTED USING A 2-WIRE TWIST TECHNIQUE WITH NEW GUIDEWIRES BUT WAS UNSUCCESSFUL. SUBSEQUENTLY, AN ATTEMPT WAS MADE TO SUCTION THE REMAINING FRAGMENT USING A NON-BSC DEVICE, BUT IT WAS ALSO UNSUCCESSFUL. THE PHYSICIAN THEN USED A 1.5 X 12 MM NON-BSC BALLOON CATHETER TO PULL IT BACK INTO THE GUIDE, BUT THIS ONLY MOVED A SEGMENT OF THE BALLOON FROM THE PROXIMAL SEGMENT TO THE MID-SEGMENT, AS COULD BE SEEN BY THE BALLOON MARKERS. SHORTLY AFTER, THE PHYSICIAN DECIDED THAT THE BEST OPTION

    Patient problems: Foreign Body In Patient

    MAUDE report 21417333 ↗
  • MalfunctionMar 6, 2024Nonstandard Device; Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 3.50MM X 8MM FG NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION, IT WAS NOTED THAT THE BALLOON BURST. THE BALLOON WAS REMOVED AND COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18849933 ↗
  • InjuryOct 13, 2021Break; Material Rupture; Detachment of Device or Device Component; Break; Material Rupture; Detachment of Device or Device Component

    IT WAS REPORTED THAT BALLOON RUPTURE, BALLOON DETACHMENT, AND SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 3.50MM X 8MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, THE BALLOON RUPTURED AFTER INFLATION AND CAME OFF THE RAIL SYSTEM. IT WAS ALSO NOTED THAT THE SHAFT FRACTURED. THE PHYSICIAN REMOVED THE BALLOON INTACT VIA A GOOSENECK SNARE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12625017 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page