Skip to content

NC Emerge® H7493926712220Recalls & reported problems

Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2

All models of NC Emerge (61)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729846369), received between Nov 12, 2020 and Dec 11, 2025 — classified as 3 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture4
  • Break3
  • Detachment of Device or Device Component1
  • Device-Device Incompatibility1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Discomfort1
  • Foreign Body In Patient1
  • Heart Block1
  • No Consequences Or Impact To Patient1
  • Reduced Blood Flow1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 11, 2025Break; Nonstandard Device; Device-Device Incompatibility; Break; Nonstandard Device; Device-Device Incompatibility

    IT WAS REPORTED VIA MEDSUN REPORT (B)(4) THAT SHAFT BREAK OCCURRED, RESULTING PATIENT COMPLICATIONS. THE PATIENT PRESENTED WITH IN-STENT RESTENOSIS. VASCULAR ACCESS WAS OBTAINED VIA THE RADIAL ARTERY WITH A NON-BOSTON SCIENTIFIC (BSC) GUIDE CATHETER. ONE TARGET LESION WAS LOCATED IN THE MID LEFT ANTERIOR DESCENDING ARTERY LAD) AND ANOTHER IN THE 1ST DIAGONAL BRANCH. A NON-BSC WIRE WAS PASSED INTO THE 1ST DIAGONAL AND ANOTHER NON-BSC WIRE WAS PLACED DOWN THE LAD. MULTIPLE BALLOONS WERE USED TO OPEN THE 1ST DIAGONAL AND THEN THE LAD. AFTER BALLOONING, INTRAVASCULAR ULTRASOUND WAS PERFORMED WHICH SHOWED UNDER EXPANSION IN THE REGION OF THE PREVIOUSLY PLACED STENT IN THE 1ST DIAGONAL. A 2.25MM X 12MM NC EMERGE BALLOON CATHETER WAS DELIVERED TO THE 1ST DIAGONAL STENT AREA AND INFLATED. HOWEVER, AFTER REMOVING THE DEVICE FROM THE PATIENT, IT WAS OBSERVED THAT THE DISTAL PORTION OF THE BALLOON BROKE OFF AND WAS NO LONGER ATTACHED TO THE END OF THE DELIVERY SYSTEM. THERE WAS NO EXCESSIVE FORCE APPLIED ON THE DEVICE AND THE END OF THE BALLON REMAINED LODGED IN THE PATIENT. A SNARE DEVICE WAS USED TO ATTEMPT TO RETRIEVE THE DETACHED PORTION BUT WAS UNSUCCESSFUL. THE ACCESS SITE WAS CHANGED T

    Patient problems: Discomfort; Foreign Body In Patient; Heart Block; Reduced Blood Flow; Thrombosis/Thrombus

    MAUDE report 23780816 ↗
  • InjuryMar 24, 2025Material Rupture; Detachment of Device or Device Component; Material Rupture; Detachment of Device or Device Component

    IT WAS REPORTED THAT BALLOON RUPTURE AND DETACHMENT OCCURRED. THE TARGET LESION WAS LOCATED IN A CALCIFIED CORONARY ARTERY. THE 2.25 X 12 MM NC EMERGE BALLOON WAS SELECTED FOR USE. AFTER SEVERAL INFLATIONS, THE BALLOON RUPTURED AND DURING REMOVAL, BROKE FROM THE SHAFT. A NON-BOSTON SCIENTIFIC HEART PUMP WAS PLACED FOR SUPPORT AND THE PATIENT WAS FLOWN TO ANOTHER FACILITY FOR SURGERY TO REMOVE THE BALLOON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21682356 ↗
  • MalfunctionJan 30, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON PINHOLE OCCURRED. A 2.25MM X 12MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING PROCEDURE, A LEAK WAS NOTED AND FOUND A PINHOLE IN THE DISTAL TIP. THE PROCEDURE WAS COMPLETED WITH NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21276130 ↗
  • MalfunctionMay 2, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.25MM X 12MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT 14 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE LESION WAS PRE-TREATED WITH A 2.00MM BALLOON. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19231864 ↗
  • MalfunctionAug 30, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT WAS PRESENTED WITH CORONARY ARTERY DISEASE AND WAS TO UNDERGO PERCUTANEOUS CORONARY INTERVENTION. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED MID RIGHT CORONARY ARTERY (RCA). A 1.2MM X 12MM THREADER BALLOON CATHETER WAS ADVANCED BUT FAILED TO PASS THE LESION. HOWEVER, DURING THE SECOND INFLATION AT 12 ATMOSPHERES (ATM), THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND ANOTHER 1.2MM X 12MM THREADER BALLOON CATHETER WAS ADVANCED, BUT IT ALSO FAILED TO PASS THE LESION AND DURING THIRD INFLATION AT 12 ATM, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND A 1.5X12 NON-BOSTON SCIENTIFIC BALLOON CATHETER WAS PLACED TO TREAT THE LESION. SUBSEQUENTLY, THE PHYSICIAN PROCEEDED TO THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED PROXIMAL RCA WITH 95% STENOSIS. A 2.25MM X 12MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATATION; HOWEVER, DURING SECOND INFLATION AT 16 ATM, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND DUE TO THE NATURE OF THE PROCEDURE, THE PATIENT WILL BE RETURNING FOR A ROTABLATOR PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17654783 ↗
  • MalfunctionJun 30, 2022Break; Break

    IT WAS REPORTED THAT A SHAFT BREAK OCCURRED. THE PATIENT PRESENTED WITH ACUTE CORONARY SYNDROME. A 2.25MM X 12MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, AS THE PHYSICIAN WAS LOADING THE BALLOON ON THE WIRE, A CHANGE IN ABILITY TO PUSH WAS NOTED WHILE MANEUVERING THE BALLOON IN THE GUIDE. THE DEVICE WAS PULLED BACK AND IT WAS NOTICED THE DEVICE HAD FRACTURED AT THE HYPOTUBE JUNCTION BEFORE ENTERING THE PATIENT'S ANATOMY. ALL OF THE DEVICE WAS PULLED AND REMOVED INTACT FROM THE PATIENT AND A NEW NC EMERGE BALLOON CATHETER WAS USED SUCCESSFULLY. THE PATIENT WAS RECOVERING AND PROCEDURE WAS COMPLETED WITHOUT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14881890 ↗
  • InjuryNov 12, 2020Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE TORTUOUS DISTAL RIGHT CORONARY ARTERY. AFTER A SYNERGY STENT WAS IMPLANTED, A 2.25MM X 12MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR POST-DILATATION. THE DEVICE DID NOT MOVE AS EXPECTED AND WHEN THE PHYSICIAN REMOVED THE DEVICE, THE DISTAL 30-40CM OF THE DEVICE REMAINED IN THE PATIENT. THE DETACHED SHAFT WAS TRAPPED IN THE GUIDE CATHETER USING ANOTHER BALLOON AND THE ENTIRE SYSTEM WAS SUCCESSFULLY REMOVED FROM THE BODY. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10831483 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page