NC Emerge® H7493926712500Recalls & reported problems
Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2
All models of NC Emerge (61)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729846451), received between Apr 7, 2019 and May 2, 2025 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture3
- Break2
- Detachment of Device or Device Component2
- Difficult to Remove1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Embolism/Embolus1
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMay 2, 2025Material Rupture; Detachment of Device or Device Component; Material Rupture; Detachment of Device or Device Component
IT WAS REPORTED THAT A DEVICE ISSUE OCCURRED RESULTING IN EMBOLISM. THE TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED RENAL ARTERY. A 5.00MM X 12MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED, AND PART OF THE BALLOON WAS MISSING FROM THE END OF THE CATHETER UPON REMOVAL. ADDITIONALLY, A FRAGMENT OF THE BALLOON SHEARED OFF AND EMBOLIZED INTO THE PATIENT'S RENAL ARTERIES. DETAILS ON FURTHER ACTION FOR THE DETACHED PORTION WERE NOT PROVIDED.
Patient problems: Embolism/Embolus; Foreign Body In Patient
MAUDE report 21951438 ↗MalfunctionNov 21, 2024Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE CORONARY ARTERY. A 5.00MM X 12MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE PROCEDURE, THE SHAFT OF THE BALLOON CATHETER BROKE WITHIN THE GUIDE CATHETER, AND THE BALLOON BECAME TRAPPED IN THE VESSEL AFTER THE STENT WAS IMPLANTED. ADDITIONALLY, A KINK WAS OBSERVED IN THE MID-SHAFT OF THE CATHETER. THE DEVICE WAS REMOVED BY PULLING IT OUT. THE PROCEDURE WAS COMPLETED USING AN ALTERNATE DEVICE WITHOUT ANY COMPLICATIONS, AND NO PATIENT-RELATED ISSUES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20746600 ↗InjuryAug 25, 2022Difficult to Remove; Material Rupture; Detachment of Device or Device Component; Difficult to Remove; Material Rupture; Detachment of Device or Device Component
IT WAS REPORTED THAT BALLOON RUPTURED, TIP DETACHMENT AND DIFFICULTY IN REMOVAL OCCURRED. A 6F AL1 GUIDE WAS ENGAGED IN THE SAPHENOUS VEIN GRAFT (SVG) TO OBTUSE MARGINAL (OM). A NON BOSTON SCIENTIFIC WIRE WAS ADVANCED TO THE DISTAL PORTION ON THE DRAFT INTO THE OM BRANCH. THE 95% IN-STENT RESTENOTIC LESION WAS DILATED WITH A 2.25 X 15MM BALLOON WIT 60% RESIDUAL. USING A OPTICROSS 60MHZ CATHETER, PERFORMED INTRAVASCULAR ULTRASOUND TO ASSESS THE VESSEL SIZE AND LESION MORPHOLOGY. THEN DILATED AND STENTED A DOWNSTREAM LESION IN THE OM BRANCH, ULTIMATELY PLACING A NON BOSTON SCIENTIFIC STENT IN THE OM, POST-DILATED TO 3.0MM WITH EXCELLENT ANGIOGRAPHIC RESULT. RETURN TREATMENT OF THE PROXIMAL GRAFT IN-STENT RESTENOTIC LESION, USING A 3.5 X 15MM NC BALLOON TO DILATE THE LESION AT HIGH PRESSURE WITH 30% RESIDUAL. A NON BOSTON SCIENTIFIC STENT WAS PLACED TO COVER THE PROXIMAL GRAFT STENOSIS AND POST DILATED WITH A 5.00MM X 12MM NC EMERGE BALLOON CATHETER , HOWEVER, DURING PROCEDURE AT 20 ATMOSPHERES THE BALLOON RUPTURED. TRIED TO REMOVE THE BALLOON, BUT IT WAS STUCK IN THE GUIDE TIP. VISUALIZATION OF THE BALLOON MARKERS INSIDE THE GUIDE, BUT THERE WAS A FAINT FILLING DEFECT AT THE GUIDE TI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15297852 ↗MalfunctionMay 18, 2021Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 5.00MM X 12MM NC EMRGE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE PROCEDURE, THE SHAFT FRACTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11845820 ↗MalfunctionApr 7, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURED. A 5.00MM X 12MM NC EMERGE BALLOON CATHETER WAS FOR USE. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. NO KNOWN PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8489925 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page