NC Emerge® H7493926715450Recalls & reported problems

Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2

All models of NC Emerge (61)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729846574), received between Dec 18, 2020 and Mar 30, 2026 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture6
  • Nonstandard Device2
  • Device-Device Incompatibility1
  • Difficult to Remove1
  • Entrapment of Device1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 30, 2026Nonstandard Device; Material Rupture; Material Separation

    IT WAS REPORTED THAT BALLOON RUPTURE AND TIP DETACHMENT OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A CALCIFIED VESSEL. A 4.50MM X 15MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATATION. THE BALLOON WAS INFLATED TWICE; HOWEVER, THE PHYSICIAN BELIEVES THE BALLOON LIKELY RUPTURED DURING THE FIRST INFLATION. BECAUSE THERE WAS NO BLOOD RETURN IN THE INFLATION DEVICE, THE BALLOON WAS REINFLATED. UPON REMOVAL OF THE DEVICE, THE TIP OF THE CATHETER WAS FOUND TO BE DETACHED. THE PHYSICIAN SEARCHED FOR THE DETACHED COMPONENT UNDER X-RAY BUT WERE UNABLE TO LOCATE IT. THE PROCEDURE WAS DELAYED WHILE THE PHYSICIAN ATTEMPTED TO IDENTIFY THE MISSING TIP. THE PROCEDURE WAS COMPLETED WITH A NON-BOSTON SCIENTIFIC BALLOON. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS EXPECTED TO MAKE A FULL RECOVERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 24717415 ↗
  • MalfunctionOct 15, 2025Nonstandard Device; Material Rupture; Device-Device Incompatibility; Nonstandard Device; Material Rupture; Device-Device Incompatibility

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A NON-TORTUOUS AND NON-CALCIFIED CORONARY ARTERY. A 4.50 MM X 15 MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR POST-DILATATION. DURING THE INITIAL INFLATION AT 6 ATMOSPHERES FOR 3 SECONDS, THE BALLOON CAME INTO CONTACT WITH THE PROXIMAL END OF A PREVIOUSLY IMPLANTED STENT, RESULTING IN BALLOON RUPTURE. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23303563 ↗
  • MalfunctionJul 18, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS, AND SEVERELY CALCIFIED CORONARY ARTERY. A 4.50MM X 15MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING SECOND INFLATION AT 12 ATMOSPHERES FOR 15 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED SUCCESSFULLY, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19779783 ↗
  • MalfunctionOct 5, 2021Entrapment of Device; Difficult to Remove; Entrapment of Device; Difficult to Remove

    REPORTABLE BASED ON ADDITIONAL INFORMATION ON 15-SEPT-2021. IT WAS REPORTED THAT CATHETER ENTRAPMENT OCCURRED. THE 85% IN-STENT RESTENOSIS (ISR) TARGET LESION WAS LOCATED IN THE NON-CALCIFIED AND NON-TORTUOUS PROXIMAL SAPHENOUS VEIN GRAFT TO OBTUSE MARGINAL. A 4.50MM X 15MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATATION AND WAS INFLATED TWICE AT 26 ATMOSPHERES FOR 20 SECONDS. THE DEVICE WAS FULLY DEFLATED, HOWEVER, IT WOULD NOT COME OUT OF THE GUIDE CATHETER. THE PHYSICIAN THEN PULLED EVERYTHING OUT TOGETHER AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED THAT THE BALLOON WAS ALSO STUCK ON THE WIRE. THE PHYSICIAN HAD TO REMOVE THE GUIDE, BALLOON, AND WIRE TOGETHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12576716 ↗
  • MalfunctionFeb 23, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS AN AREA OF IN-STENT RESTENOSIS LESION IN THE RIGHT CORONARY ARTERY. A LASER WAS INITIALLY USED TO MODIFY THE LESION. A 4.50MM X 15MM NC EMERGE BALLOON CATHETER WAS ADVANCED AND INITIALLY INFLATED AT 16 ATMOSPHERES. HOWEVER, DURING THE SECOND INFLATION AT 10 ATMOSPHERES (ATM) IT RUPTURED. IT WAS REPLACED WITH ANOTHER 4.50MM X 15MM NC EMERGE BUT IT ALSO RUPTURED DURING INITIAL INFLATION AT 14 ATM. SUBSEQUENTLY, A 4.50MM X 20MM NC EMERGE WAS ADVANCED BUT ALSO RUPTURED DURING THE SECOND INFLATION AT 16 ATM. THE DEVICES WERE COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A WOLVERINE CUTTING BALLOON FOLLOWED BY A NON-BSC BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11369568 ↗
  • MalfunctionFeb 23, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS AN AREA OF IN-STENT RESTENOSIS LESION IN THE RIGHT CORONARY ARTERY. A LASER WAS INITIALLY USED TO MODIFY THE LESION. A 4.50MM X 15MM NC EMERGE BALLOON CATHETER WAS ADVANCED AND INITIALLY INFLATED AT 16 ATMOSPHERES. HOWEVER, DURING THE SECOND INFLATION AT 10 ATMOSPHERES (ATM) IT RUPTURED. IT WAS REPLACED WITH ANOTHER 4.50MM X 15MM NC EMERGE BUT IT ALSO RUPTURED DURING INITIAL INFLATION AT 14 ATM. SUBSEQUENTLY, A 4.50MM X 20MM NC EMERGE WAS ADVANCED BUT ALSO RUPTURED DURING THE SECOND INFLATION AT 16 ATM. THE DEVICES WERE COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A WOLVERINE CUTTING BALLOON FOLLOWED BY A NON-BSC BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11369564 ↗
  • InjuryDec 18, 2020Material Rupture; Material Rupture

    IT WAS FURTHER REPORTED THAT THIS IS THE SAME CASE AS USER FACILITY MAUDE REPORT #MW5098430 IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 70% STENOSED, TARGET LESION WAS LOCATED IN A CALCIFIED VESSEL. A 4.50MM X 15MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR POST DILATATION OF A SYNERGY DRUG-ELUTING STENT. HOWEVER, DURING INFLATION AT 22 ATMOSPHERES, THE BALLOON RUPTURED. WHEN THE DEVICE WAS REMOVED FROM THE PATIENT, IT LOOKED LIKE A PIECE OF THE BALLOON WAS MISSING. NO BALLOON FRAGMENTS WERE FOUND VIA FLUOROSCOPY BUT THE PHYSICIAN THOUGHT A PIECE REMAINED IN THE PATIENT. ALL DEVICES WERE REMOVED FROM THE PATIENT AND THE PROCEDURE WAS NOT COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 11039902 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →