Skip to content

NC Emerge® H7493926715550Recalls & reported problems

Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2

All models of NC Emerge (61)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729846598), received between Sep 30, 2024 and Oct 30, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 30, 2024Break; Difficult to Remove; Break; Difficult to Remove

    IT WAS REPORTED THAT CATHETER REMOVAL DIFFICULTIES AND A SHAFT BREAK WERE ENCOUNTERED. THE PATIENT EXPERIENCING CARDIOVASCULAR INSTABILITY AND PERSISTENT ISCHEMIC SYMPTOMS, INCLUDING CHEST PAINS. AN ELECTROCARDIOGRAM (EKG) INDICATED A CONCERN FOR AN INFERIOR ST ELEVATION MYOCARDIAL INFARCTION (STEMI). A 5.50MM X 15MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INTERVENTION, THERE WAS DIFFICULTY WITHDRAWING THE BALLOON, WHICH ULTIMATELY RESULTED IN A BALLOON SHAFT FRACTURE. THE BALLOON TIP AND SHAFT REMAINED JUST PROXIMAL TO THE GUIDELINER. WHEN PULLED BACK INTO THE GUIDELINER, ABOUT 20CM OF THE BALLOON REMAINED IN THE PATIENT OUT THE GUIDE. THE BALLOON SHAFT AND TIP WERE SUCCESSFULLY RETRIEVED. ALL EQUIPMENT WAS REMOVED INTACT, WITH NO APPARENT HARM TO THE PATIENT. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS MODERATELY TORTUOUS WITH 70% STENOSIS AND SNARING WAS PERFORMED TO COMPLETELY REMOVED THE FRACTURED DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20568310 ↗
  • InjurySep 30, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT BALLOON TIP DETACHMENT OCCURRED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 5.50 MM X 15 MM NC EMERGE BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE PROCEDURE, THE BALLOON TIP DETACHED FROM THE BALLOON SHAFT. THE DEVICE WAS REMOVED WITH A SNARE, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER 5.5 MM NC EMERGE BALLOON CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20334468 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page