IntellaTip MiFi™ XP M004PM47900Recalls & reported problems
Boston Scientific Corporation · Temperature Ablation Catheter · Class 3
All models of IntellaTip MiFi XP (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729848042), received between Jan 9, 2018 and Jun 21, 2021 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Bent” (1 report).
1 other recall cover devices cleared under the same FDA submission (P020025); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Bent1
- Communication or Transmission Problem1
- Device Contaminated During Manufacture or Shipping1
- Device Contamination with Chemical or Other Material1
- Insufficient Heating1
- Positioning Problem1
Patient problems
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 21, 2021Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT A FOREIGN OBJECT WAS FOUND INSIDE THE STERILIZATION BAG AFTER UNPACKING THE BOX CONTAINING THE INTELLATIP MIFI CATHETER. WHEN THE BAG WAS CHECKED CLOSELY, THERE WERE PLASTIC PIECES FOUND. SINCE THE COATING OF THE MAIN UNIT HANDLE PEELED OFF, IT IS POSSIBLE THAT IT WAS PART OF THE PLASTIC PIECES. NO PATIENT COMPLICATIONS WERE REPORTED. A DIFFERENT CATHETER WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12032055 ↗MalfunctionJun 4, 2019Communication or Transmission Problem; Communication or Transmission Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 20MAY2019. IT WAS REPORTED THE CATHETER COULD NOT BE RECOGNIZED. DURING AN ABLATION PROCEDURE WITH AN INTELLATIP MIFI XP CATHETER WHEN THE ABLATION CATHETER WAS PLACED IN THE HEART, THE CATHETER COULD NOT BE RECOGNIZED. RECOGNITION WAS IMPROVED BY CHECKING THE CONNECTION AND STARTING UP THE EQUIPMENT, BUT IMPEDANCE WAS HIGH, AND ENERGIZATION COULD NOT BE PERFORMED. CONNECTION CHECKING, EQUIPMENT STARTING UP, AND CONNECTION CABLE REPLACEMENT WERE PERFORMED, BUT THE ISSUE WAS NOT RESOLVED. IT WAS REQUESTED TO REPLACE TO ABLATION CATHETER, BUT SINCE IT ALREADY TOOK TIME, THE PHYSICIAN TOLD TO PERFORM THE PROCEDURE USING NON-BSC SYSTEM ABLATION CATHETER, AND THAT WAS FOLLOWED. NO PATIENT COMPLICATIONS WERE REPORTED. DEVICE ANALYSIS REVEALED RINGS WERE INSPECTED AND RING #1 SEAL WAS FOUND COMPROMISED SINCE BROKEN ADHESIVE WAS OBSERVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8666177 ↗MalfunctionMar 14, 2019Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT A BLACK FOREIGN OBJECT WAS FOUND INSIDE THE STERILE CATHETER PACKAGING. DURING PREPARATION FOR AN ATRIAL FLUTTER ABLATION PROCEDURE, A BLACK FOREIGN OBJECT WAS FOUND INSIDE THE STERILE CATHETER PACKAGING. THE DEVICE WAS NOT USED FOR THE PROCEDURE AND ANOTHER OF THE SAME CATHETER WAS USED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CURRENT CONDITION IS FINE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8420025 ↗MalfunctionFeb 21, 2018Insufficient Heating; Insufficient Heating
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2018. IT WAS REPORTED THAT THE DEVICE WAS UNABLE TO ABLATE. A INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER WAS SELECTED FOR A ATRIAL FLUTTER(AFL) ABLATION. HOWEVER ABLATION WAS PERFORMED WITH 60W AND 60 DEGREES SETTINGS, BUT WHILE THE W INCREASED, THE TEMPERATURE DID NOT INCREASE MORE THAN ABOUT 40 DEGREES, AND IT DID NOT CHANGE WITHOUT REDUCING THE POTENTIAL OF THE MINI ELECTRODE OF MIFI. ABLATION WAS PERFORMED FOR ABOUT 15 TIMES, THE PROCEDURE WAS COMPLETED WITH ANOTHER INTELLATIP MIFI. NO COMPLICATIONS WERE REPORTED. HOWEVER ANALYSIS REVELED BROKEN RING SEAL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7285931 ↗MalfunctionJan 9, 2018Bent; Positioning Problem; Bent; Positioning Problem
REPORTABLE BASED ON ANALYSIS COMPLETED (B)(6) 2017. IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE TO TREAT ATRIAL FLUTTER, THE INTELLATIP MIFI¿ XP ABLATION CATHETER WAS NOT ABLE TO COME IN CONTACT PROPERLY WITH THE TARGET AREA, AND ¿THE POTENTIAL DID NOT ATTENUATE EVEN THOUGH ENERGIZATION WAS PERFORMED.¿ THE CATHETER WAS REMOVED AND ¿THE ELECTRODE PART OF NO. 3-4¿ WAS BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED THE SEALS ON RING 1 AND RING 2 WERE COMPROMISED AND A PIECE OF THE RING 1 SEAL WAS MISSING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7175910 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020025. One submission often covers many models, so check the product and lot details in each record.
- Z-0202-2019 · Jun 20, 2018 · Open, ClassifiedBoston Scientific Maestro 4000 Cardiac Ablation System, Firmware Version 5.14, Model Numbers: M00440000, M0044000D0, M0044000H0,M0044000R0, M0044000Z0,
Other catheter, percutaneous, cardiac ablation, for treatment of atrial flutter devices
- QDOT MICRO D139505Biosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter27 deaths915 reports
- IntellaNav MiFi™ Open-Irrigated M004PMR96200Boston Scientific Corporation · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter4 deaths269 reports
- QDOT MICRO D139504Biosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter7 deaths138 reports
- Sphere-9™ Catheter AFR-00001Medtronic, Inc. · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter1 recall5 deaths77 reports
- QDOT MICRO D139401Biosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter2 deaths114 reports
- NaviStar ThermoCool NI75TCDHBiosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter111 reports
- NaviStar RMT ThermoCool NR7TCSIYBiosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter9 deaths103 reports
- IntellaNav MiFi™ Open-Irrigated M004PMR9620K20Boston Scientific Corporation · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter78 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page