Skip to content

Expel™ Nephroureteral Stent System with Twist-Loc™ Hub H7493935108260Recalls & reported problems

Boston Scientific Corporation · Nephroureteral Stent System · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729860945), received between Nov 5, 2020 and Nov 5, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Entrapment of Device1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2020Break; Entrapment of Device; Break; Entrapment of Device

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 27OCT2020. IT WAS REPORTED THAT THE FLEXIBLE CANNULA BECAME STUCK AND BROKE INSIDE THE CATHETER. AN 8.3/26 EXPEL NEPHROURETERAL STENT SYSTEM WITH TWIST-LOC HUB WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE PLASTIC STYLET WOULD GET STUCK IN THE DRAINAGE CATHETER WITH THE WIRE IN PLACE. THEN, IT WAS NOTED THAT THE SOFT PLASTIC STIFFENER BROKE OFF IN THE DRAINAGE CATHETER. THE PHYSICIAN HAD TO REMOVE EVERYTHING FROM THE PATIENT AND START OVER. THE PROCEDURE WAS COMPLETED WITH A NON-BOSTON SCIENTIFIC PRODUCT. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE SHAFT MIDDLE SECTION WAS DETACHED/SEPARATED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10791796 ↗

Other STENT, URETERAL devices

All STENT, URETERAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page