Skip to content

Peripheral RotaWire™ and wireClip® Torquer M0011373310Recalls & reported problems

Boston Scientific Corporation · Guidewire and Guidewire Manipulation Device · Class 2

All models of Peripheral RotaWire and wireClip Torquer (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729863632), received between Aug 11, 2025 and Aug 11, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE TIP BROKE OFF AND WAS LEFT INSIDE THE PATIENT. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY (LAD). A PERIPHERAL ROTAWIRE GUIDEWIRE AND A PERIPHERAL ROTALINK PLUS WERE SELECTED FOR USE. DURING THE PROCEDURE, THE SPEED WAS SET TO 160 RPM; HOWEVER, THE SPEED REACHED UP TO 163 RPM. UPON EXCHANGING ROTAWIRE TO ANOTHER GUIDEWIRE, IT WAS NOTED THAT THE 5MM TIP OF THE ROTAWIRE BROKE OFF AND REMAINED DISTAL LAD VERY CLOSE TO THE APEX. THE PHYSICIAN BELIEVE THAT THE TIP OF THE WIRE WAS PULLED TOO HARD WHILE BEING TAKEN OUT OF THE WIRE HOOP. THE PROCEDURE PROCEEDED AS PLANNED AND COMPLETED USING ORIGINAL METHOD. THE FRAGMENT WAS NOT REMOVED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 22778236 ↗

Other Catheter, peripheral, atherectomy devices