Innova™ Vascular H74939293051030Recalls & reported problems

Boston Scientific Corporation · Self-Expanding Stent System · Class 3 · Implant

All models of Innova Vascular (42)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729873938), received between Aug 14, 2018 and Dec 21, 2023 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Premature Activation” (3 reports).

2 other recalls cover devices cleared under the same FDA submission (P140028); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Premature Activation3
  • Activation, Positioning or Separation Problem1
  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 21, 2023Premature Activation; Premature Activation

    IT WAS REPORTED THAT STENT INADVERTENT DEPLOYMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE FEMORAL ARTERY. A 5 X 100 X 130 INNOVA VASCULAR SELF-EXPANDING STENT WAS SELECTED FOR TREATMENT. HOWEVER, DURING UNPACKING, IT WAS NOTED THAT THE STENT WAS PARTIALLY DEPLOYED FROM THE CATHETER. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18373750 ↗
  • MalfunctionFeb 11, 2021Premature Activation; Premature Activation

    IT WAS REPORTED THAT THE STENT PREMATURELY DEPLOYED. A 5MM X 100MM X 130CM INNOVA WAS SELECTED FOR USE IN A STENTING PROCEDURE IN THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA). DURING THE PROCEDURE, THE STENT BEGAN TO DEPLOY BEFORE IT WAS IN POSITION. THE DEPLOYMENT SAFETY TOOL WAS STILL ON. THE DEVICE WAS REMOVED WITHOUT ISSUE. THE PROCEDURE WAS COMPLETED WITH ANOTHER 5MM X 100MM INNOVA. NO PATIENT COMPLICATIONS WERE REPORTED, AND IT WAS A GOOD OUTCOME. IT WAS FURTHER REPORTED THAT: A CONTRALATERAL APPROACH WAS TAKEN TO ACCESS THE LESION AND THE SAFETY LOCK WAS STILL IN PLACE PRIOR TO POSITIONING IN THE LESION. EXPOSING THE STENT PRIOR TO DEPLOYMENT WAS PERFORMED AS DIRECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11317213 ↗
  • MalfunctionDec 11, 2018Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    IT WAS REPORTED THAT THE STENT PARTIALLY DEPLOYED. A 5X100X130MM INNOVA VASCULAR STENT WAS SELECTED FOR USE FOR A SUBCLAVIAN GRAFT BYPASS PROCEDURE IN THE LEG. THE TARGET LESION DID NOT HAVE MUCH TORTUOSITY, HAD NO CALCIFICATION, AND NO STENOSIS WAS OBSERVED, HOWEVER, THE TARGET LESION WAS THROMBOSED. FOLLOWING PRE-DILATION, THE INNOVA STENT WAS ADVANCED OVER A .018 GUIDEWIRE AND DEPLOYMENT INITIATED. THE STENT ONLY BEGAN TO FLOWER FROM THE DEPLOYMENT SYSTEM WHEN THE THUMBWHEEL BEGAN TO SPIN FREELY WITHOUT THE STENT DEPLOYING ANY FURTHER. THE PULL GRIP WAS THEN ATTEMPTED AND AGAIN THE STENT DID NOT DEPLOY. THE PHYSICIAN ALSO ATTEMPTED TO BREAK OPEN THE HANDLE TO SEE IF THE STENT COULD BE DEPLOYED, BUT THIS WAS UNSUCCESSFUL. THE PHYSICIAN ALSO ATTEMPTED TO SWITCH OUT THE GUIDEWIRE FOR A .035 SIZE AND THAT WAS ALSO UNSUCCESSFUL. THE PHYSICIAN THEN INSERTED A LARGER AND LONGER SHEATH AND WAS ABLE TO WITHDRAW THE STENT INTO THE SHEATH AND REMOVE THE STENT AND DELIVERY SYSTEM TOGETHER. THERE WERE NO PATIENT COMPLICATIONS AND THE PATIENT'S STATUS POST PROCEDURE WAS FINE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8152384 ↗
  • MalfunctionAug 14, 2018Entrapment of Device; Premature Activation; Entrapment of Device; Premature Activation

    IT WAS REPORTED THE DELIVERY SYSTEM BECAME STUCK ON THE GUIDEWIRE. VASCULAR ACCESS WAS OBTAINED VIA RAIL ACCESS IN THE GROIN AND POPLITEAL. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY (SFA). A .035 ZIPWIRE WAS PUT INTO POSITION AND THEN A 5 X 120 X 130 INNOVA AND A 5 X 100 X 130 INNOVA WERE LOADED OVER BOTH ENDS OF THE WIRE AND ADVANCED TO THE TREATMENT SITE. THE STENT DELIVERY SYSTEMS BECAME STUCK ON THE ZIPWIRE DURING ADVANCEMENT TO THE TARGET LESION. WHILE TRYING TO REMOVE ONE OF THE DELIVERY SYSTEMS, THE STENT BEGAN TO DEPLOY. THE PHYSICIAN PUSHED THE ZIPWIRE OUT OF THE PATIENT ENOUGH THAT ONE OF THE STENTS WAS OUTSIDE THE PATIENT AND THE PHYSICIAN WAS ABLE TO CUT THE WIRE. THE REST OF THE WIRE AND THE SECOND INNOVA STENT WAS ABLE TO BE REMOVED FROM THE PATIENT WITH NO FURTHER COMPLICATIONS. THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICES. THE PATIENT WAS FINE POST PROCEDURE. RETURNED PRODUCT CONSISTED OF AN INNOVA SELF-EXPANDING STENT DELIVERY SYSTEM (SDS). VISUAL EXAMINATION SHOWED A KINK AT THE NOSECONE. NO DAMAGE WAS NOTICED ON THE MID-SHAFT. THE YELLOW THUMBWHEEL LOCK WAS INTACT ON THE DEVICE AND THE PULL HANDLE WAS IN THE START POSITION. THE STENT W

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7783202 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140028. One submission often covers many models, so check the product and lot details in each record.

  • Z-0686-2018 · Nov 6, 2017 · Open, Classified
    Innova Self-Expanding Stent System, 180mm. Labeled with Material Number and description as follows: Product Usage: The Innova Vascular Self-Expanding Stent System is indicated
  • Z-0685-2018 · Nov 6, 2017 · Open, Classified
    Innova Self-Expanding Stent System, 200 mm. Labeled with Material Number and description as follows: Product Usage: The Innova Vascular Self-Expanding Stent System is indicated t

Other STENT, SUPERFICIAL FEMORAL ARTERY devices

All STENT, SUPERFICIAL FEMORAL ARTERY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page