ELUVIA™ H74939294701010Recalls & reported problems

Boston Scientific Corporation · Drug-Eluting Vascular Stent System · Class 3 · Implant

All models of ELUVIA (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729876724), received between Sep 27, 2019 and Oct 19, 2023 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem1
  • Deformation Due to Compressive Stress1
  • Difficult to Advance1
  • Entrapment of Device1
  • Fracture1
  • Material Integrity Problem1
  • Unsealed Device Packaging1

Patient problems

  • Discomfort1
  • Hematoma1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • Pain1
  • Swelling/ Edema1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 19, 2023Fracture; Deformation Due to Compressive Stress; Activation, Positioning or Separation Problem; Difficult to Advance; Material Integrity Problem; Fracture; Deformation Due to Compressive Stress; Activation, Positioning or Separation Problem; Difficult to Advance; Material Integrity Problem

    IT WAS REPORTED THAT STENT PARTIAL DEPLOYMENT AND FRACTURE OCCURRED REQUIRING ADDITIONAL INTERVENTION. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). A 7X100, 130 CM ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM WAS SELECTED FOR A LEFT LEG INTERVENTION. PEDAL ACCESS, AND ANGIOGRAM WAS PERFORMED. LASER ATHERECTOMY OF THE SFA/POPLITEAL (POP)/AT VESSELS. BALLOON ANGIOPLASTY WAS PERFORMED POST ATHERECTOMY. A PERFORATED SEGMENT IN THE DISTAL POPLITEAL ARTERY WITH ANGIOGRAPHIC EVIDENCE WAS NOTED. THE ELUVIA STENT WAS PLACED OVER THE 014 NON-BOSTON SCIENTIFIC GUIDEWIRE TO BE DEPLOYED IN THE PROXIMAL SFA. THE PHYSICIAN HAD TROUBLE ADVANCING THE DEVICE TO THE LESION LOCATION IN THE DISTAL SFA/POP LOCATION. THE ENTIRE SEGMENT FROM TPT-COMMON FEMORAL ARTERY (CFA) WAS STENTED PREVIOUSLY WITH IN-STENT RESTENOSIS (ISR) EFFECTING THE ENTIRE LENGTH. WHEN THE DEVICE COULD NOT BE ADVANCED TO THAT AREA FOR IMPLANTATION, THE PHYSICIAN DECIDED TO DEPLOY THE STENT IN THE AREA OF PERFORATION, POPLITEAL ARTERY. THE PHYSICIAN WAS ADVANCING THE STENT, TO THE PRIMARY SFA LESION LOCATION, AGGRESSIVELY AND THE STENT SHAFT WAS

    Patient problems: Discomfort; Hematoma; Pain; Swelling/ Edema

    MAUDE report 17966964 ↗
  • MalfunctionDec 21, 2022Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED THAT ONE CORNER OF THE STERILE PACKAGING WAS NOT COMPLETELY SEALED. DURING DEVICE PREPARATION FOR THE ANGIOGRAM PROCEDURE, AN ELUVIA US, 7X100, 130 CM WAS SELECTED FOR USE TO TREAT PERIPHERAL ARTERY DISEASE. AS THE DEVICE PACKAGING WAS OPENED, IT WAS OBSERVED THAT ONE CORNER OF THE STERILE PACKAGING WAS NOT PROPERLY SEALED, COMPROMISING THE STERILITY OF THE DEVICE. THE POTENTIALLY CONTAMINATED ELUVIA WAS NOT USED AND WAS DISPOSED OF BY THE STAFF. THE ELUVIA WAS REPLACED IN ORDER TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16030041 ↗
  • MalfunctionSep 27, 2019Entrapment of Device; Entrapment of Device

    IT WAS REPORTED THAT GUIDEWIRE AND CATHETER ENTRAPMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A .035 MAGIC TORQUE GUIDEWIRE WAS USED WITH A 7X100, 130CM ELUVIA SELF-EXPANDING STENT. DURING THE PROCEDURE THE GUIDEWIRE BECAME STUCK WITHIN THE STENT DELIVERY SYSTEM AND COULD NOT BE REMOVED. THE MAGIC TORQUE GUIDEWIRE AND THE ELUVIA STENT DELIVERY SYSTEM WERE REMOVED TOGETHER AS ONE UNIT. THE PROCEDURE WAS THEN COMPLETED USING A NEW GUIDEWIRE AND A 7X40MM ELUVIA STENT. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9131412 ↗

Other Stent, superficial femoral artery, drug-eluting devices

All Stent, superficial femoral artery, drug-eluting devices →