- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729877622), received between Jun 1, 2021 and Oct 18, 2024 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove2
- Adverse Event Without Identified Device or Use Problem1
- Contamination1
- Defective Device1
- Device Contamination with Chemical or Other Material1
- Material Perforation1
- Packaging Problem1
- Tear, Rip or Hole in Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Vasoconstriction1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 18, 2024Contamination; Tear, Rip or Hole in Device Packaging; Contamination; Tear, Rip or Hole in Device Packaging
THE DYNAMIC XT ELECTRODE CATHETER WAS SELECTED FOR USE DURING THE PROCEDURE. IT WAS REPORTED THAT WHEN THE OUTER PACKAGE OF THE DEVICE WAS REMOVED, THEY FOUND THE PLASTIC PROTECTIVE SLEEVE OUTSIDE THE CATHETER WAS FRACTURED. THE DEVICE WAS CONTAMINATED AND COULD NOT BE USED DUE TO DAMAGE TO THE PROTECTIVE SLEEVE. CATHETER WAS REPLACED WITH A NEW ONE FROM THE SAME MODEL. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT PATIENT COMPLICATIONS. THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20488681 ↗MalfunctionAug 10, 2024Device Contamination with Chemical or Other Material; Adverse Event Without Identified Device or Use Problem; Device Contamination with Chemical or Other Material; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE PREPARATION OF AN ABLATION PROCEDURE TO TREAT A SUPRAVENTRICULAR TACHYCARDIA (SVT), A DYNAMIC XT STEERABLE DIAGNOSTIC CATHETER WAS SELECTED FOR USE. ON OPENING STERILE PACKAGING, SCOUT NURSE IDENTIFIED A SMALL FIBER THAT APPEARED TO REPRESENT A HAIR/STRAND OF GLUE ATTACHED NEAR A JOIN TOWARD THE END OF THE BLACK SECTION OF THE HANDLE. DEVICE WAS REPLACED AND THE ISSUE WAS RESOLVED. PROCEDURE WAS ABLE TO BE COMPLETED SUCCESSFULLY WITHOUT ANY PATIENT COMPLICATIONS. CATHETER HAS BEEN RECEIVED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19957764 ↗MalfunctionJun 18, 2024Packaging Problem; Packaging Problem
THE DYNAMIC XT ELECTRODE CATHETER WAS SELECTED FOR A ABLATION PROCEDURE. PRIOR TO USE, IT WAS NOTED THAT THE PLASTIC OUTER PACKAGING OF THE DEVICE WAS DAMAGED AND NO LONGER STERILE. THE DEVICE WAS REPLACED WITH A NEW ONE FROM THE SAME MODEL. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT PATIENT COMPLICATIONS. IT IS UNKNOWN IF DEVICE IS AVAILABLE FOR RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19559727 ↗MalfunctionJul 6, 2023Material Perforation; Material Perforation
IT WAS REPORTED THAT DURING PREPARATION FOR A PROCEDURE USING A DYNAMIC XT CATHETER THE SHAFT OF THE SHEATH WAS BROKEN. THEY EXCHANGED THE CATHETER, AND THE ISSUE WAS RESOLVED. THE PROCEDURE WAS THEN COMPLETED WITH NO PATIENT COMPLICATIONS. THE CATHETER HAS BEEN RECEIVED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17270174 ↗InjuryDec 8, 2022Difficult to Remove; Difficult to Remove
DURING A RADIOFREQUENCY ABLATION PROCEDURE, A DYNAMIC XT ELECTRODE CATHETER WAS SELECTED FOR USE. IT WAS REPORTED THAT WHEN THE CATHETER WAS DELIVERED INTO THE BRANCH OF THE VEIN, VENOUS CONTRACTURE OCCURRED, AND THE CATHETER COULD NOT BE WITHDRAWN. THE PHYSICIAN WAS EVENTUALLY ABLE TO REMOVE THE CATHETER, AND A DIFFERENT CATHETER WAS USED TO COMPLETE THE PROCEDURE. NO DAMAGE OR PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Vasoconstriction
MAUDE report 15941100 ↗MalfunctionJun 1, 2021Difficult to Remove; Defective Device; Difficult to Remove; Defective Device
IT WAS REPORTED THAT A DYNAMIC XT ELECTRODE CATHETER WAS USED IN A TACHYCARDIA ELECTROPHYSIOLOGICAL EXAMINATION AND RADIOFREQUENCY ABLATION PROCEDURE. IT WAS NOTED THAT THE PATIENT HAD A SEVERELY TORTUOUS LEFT FEMORAL VEIN. UPON ADVANCING TO THE JUNCTION OF RIGHT FEMORAL VEIN AND INFERIOR VENA CAVA THE BLOOD VESSEL HAD ANGULATION OF 90 DEGREES. AFTER PUSHING THOUGHT THE ANGULATION THE CATHETER BECAME STUCK AND WAS DAMAGED. THE CATHETER WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITHOUT PATIENT COMPLICATIONS. IT WAS FURTHER REPORTED THAT THE CATHETER'S FAILURE WAS THAT THE CATHETER WAS UNABLE TO BEND. THE CATHETER WAS REMOVED BY SIMPLY PULLING THE CATHETER OUT WITHOUT INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11913507 ↗
FDA clearance and classification
The FDA regulates product code DRF (catheter, electrode recording, or probe, electrode recording) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Dynamic XT™ M0042011020
Has the Dynamic XT™ M0042011020 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729877622), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Dynamic XT™ M0042011020?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jun 1, 2021 and Oct 18, 2024. In those reports, the most common specific device problems are “Difficult to Remove” (2), “Contamination” (1) and “Defective Device” (1); the most common patient problem is “Vasoconstriction” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Dynamic XT™ M0042011020?
None of the reports are classified as deaths, 1 as an injury and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Dynamic XT™ M0042011020, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Dynamic XT™, model M0042011020). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices
- IntellaMap Orion™ M004RC64S0Boston Scientific Corporation · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths356 reports
- Advisor™ HD Grid, Sensor Enabled™ D-AVHD-DF16St. Jude Medical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths356 reports
- ACHIEVE™ 990063-020Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING7 deaths129 reports
- Inquiry™ IBI-81104Irvine Biomedical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 recall2 deaths64 reports
- DECANAV R7F282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING4 deaths108 reports
- LASSO NAV D134301Biosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING2 deaths82 reports
- DECANAV R7D282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 death76 reports
- ACHIEVE ADVANCE™ 2ACH20Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING3 deaths69 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page