Skip to content

Viking™ M0044000160Recalls & reported problems

Boston Scientific Corporation · Fixed Curve Diagnostic Catheter · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729878162), received between Aug 6, 2019 and Aug 8, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Mechanics Altered” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Mechanics Altered2

Patient problems

  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 8, 2019Mechanics Altered; Mechanics Altered

    REPORTABLE BASED ON ANALYSIS COMPLETED ON 12 JUL 2019. IT WAS REPORTED THAT A VIKING CATHETER HAD ITS CURVATURE REVERSED. THE PROCEDURE WAS COMPLETED VIA A ALTERNATE METHOD WITH NO PATIENT COMPLICATIONS. HOWEVER, ANALYSIS OF THE RETURNED DEVICE REVEALED EXPOSED SILVER EPOXY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8875522 ↗
  • MalfunctionAug 6, 2019Mechanics Altered; Mechanics Altered

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 12JUL2019. IT WAS REPORTED THAT THE CATHETER CURVE WAS BAD. DURING AN ABLATION PROCEDURE WITH A VIKING ELECTRODE CATHETER, THE CATHETER CURVATURE WAS REVERSED MAKING THE DEVICE UNUSABLE. NO PATIENT COMPLICATIONS OCCURRED. DEVICE ANALYSIS REVEALED THAT ELECTRODE 5 HAD BEEN MOVED IN THE DISTAL DIRECTION EXPOSING THE SILVER EPOXY AND ELECTRODE 5 SIGNAL WIRE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8866387 ↗

Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices

All CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page