Skip to content

IntellaNav™ XP M004R4500THK20Recalls & reported problems

Boston Scientific Corporation · Temperature Ablation Catheter · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729885757), received between May 17, 2017 and May 17, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).

1 other recall cover devices cleared under the same FDA submission (P020025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Operates Differently Than Expected1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 17, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    REPORTABLE BASED ON ANALYSIS COMPLETED APRIL 21, 2017. IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, AN ARRHYTHMIA WAS INDUCED AND THE PHYSICIAN INTRODUCED THE INTELLANAV¿ XP ABLATION CATHETER TO MAP; HOWEVER, THE CATHETER COULD NOT BE SEEN ON THE RHYTHMIA MAPPING SYSTEM. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS. HOWEVER, RETURNED PRODUCT ANALYSIS REVEALED BODY FLUID UNDER RING 2 AND IN THE INTERIOR OF THE SHEATH, AND THAT IT IS LIKELY THE RING SEAL WAS COMPROMISED DURING THE PROCEDURE DUE TO A BEND IN THE DISTAL TIP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6571823 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020025. One submission often covers many models, so check the product and lot details in each record.

  • Z-0202-2019 · Jun 20, 2018 · Open, Classified
    Boston Scientific Maestro 4000 Cardiac Ablation System, Firmware Version 5.14, Model Numbers: M00440000, M0044000D0, M0044000H0,M0044000R0, M0044000Z0,

Other catheter, percutaneous, cardiac ablation, for treatment of atrial flutter devices

All catheter, percutaneous, cardiac ablation, for treatment of atrial flutter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page