Skip to content

Wolverine Coronary Cutting Balloon H74939401102750Recalls & reported problems

Boston Scientific Corporation · Microsurgical Dilatation Device · Class 3

All models of Wolverine Coronary Cutting Balloon (22)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729888253), received between Jan 11, 2021 and Jul 29, 2025 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

6 other recalls cover devices cleared under the same FDA submission (P950020); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Adverse Event Without Identified Device or Use Problem1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information2
  • Cardiac Perforation1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 29, 2025Break; Break

    IT WAS REPORTED THAT A SHAFT BREAK OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED ARTERY. A 10MMX2.75MM WOLVERINE CORONARY CUTTING BALLOON WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE CATHETER SHAFT BROKE WHILE DELIVERING AND WAS ABLE TO REMOVE IN TWO PIECES WITHOUT SNARING. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22649972 ↗
  • InjuryDec 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SHOCKWAVE LITHOPLASTY COMPARED TO CUTTING BALLOON TREATMENT IN CALCIFIED CORONARY DISEASE: A RANDOMIZED CONTROLLED TRIAL (SHORT-CUT TRIAL). IT WAS REPORTED THAT CARDIAC PERFORATION OCCURRED. ON (B)(6) 2024, THE SUBJECT PRESENTED WITH STABLE ANGINA AND UNDERWENT PERCUTANEOUS CORONARY INTERVENTION (PCI) USING CUTTING BALLOON ANGIOPLASTY IN THE MID LEFT DESCENDING CORONARY ARTERY (LAD). PRE-PROCEDURE TIMI FLOW WAS I WITH 99% STENOSIS. A 10MMX2.75MM WOLVERINE CUTTING BALLOON WAS SELECTED FOR PROCEDURE. DURING PROCEDURE, DRUG ELUTING STENT WAS IMPLANTED IN THE MID LAD FOLLOWED BY POST-STENT DILATION WITH A NON-COMPLAINT BALLOON. A SMALL CARDIAC PERFORATION OF THE MID LAD WAS NOTED AND WAS TREATED WITH A COVERED STENT AND PROTAMINE. POST PROCEDURE TIMI FLOW WAS III WITH 0% STENOSIS. THE PATIENT WAS DISCHARGED THE SAME DAY.

    Patient problems: Cardiac Perforation

    MAUDE report 21036718 ↗
  • MalfunctionMar 2, 2022Break; Break

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 10FEB2022. IT WAS REPORTED THAT THE BALLOON BROKE. THE TARGET LESION WAS LOCATED IN THE CALCIFIED DIAGONAL ARTERY. A 10MMX2.75MM WOLVERINE CORONARY CUTTING BALLOON WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON BROKE NEAR THE HUB. THE DEVICE WAS PULLED OUT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT A HYPOTUBE BREAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13656548 ↗
  • MalfunctionJun 3, 2021Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT THE CATHETER SHAFT BROKE. THE TARGET LESION WAS LOCATED IN THE TORTUOUS AND MODERATELY CALCIFIED ARTERY. A 10MM X 2.75MM WOLVERINE CORONARY CUTTING BALLOON WAS SELECTED FOR TREATMENT WITHIN THE TARGET LESION. THE BALLOON WAS INFLATED AND DEFLATED WITHOUT ISSUE. AS THE WOLVERINE CORONARY CUTTING BALLOON WAS BEING WITHDRAWN, IT KINKED AND THEN BROKE INTO TWO PIECES. THE DEVICE WAS PULLED BACK INTO THE GUIDE CATHETER AND FULLY REMOVED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11931583 ↗
  • MalfunctionJan 27, 2021Break; Break

    IT WAS REPORTED VIA MAUDE REPORT 3600840000-2021-8005 THAT SHAFT BREAK OCCURRED. A 10MM X 2.75MM WOLVERINE CORONARY CUTTING BALLOON WAS SELECTED FOR USE. DURING THE PROCEDURE, UPON ADVANCING THE BALLOON THROUGH THE GUIDE OVER A WIRE, IT WAS NOTED THAT THE SHAFT PROXIMAL TO THE ENTRY POINT SNAPPED IN HALF. THE DEVICE WAS REMOVED COMPLETELY FROM THE GUIDE WITHOUT ANY INTERVENTION OR DEVICES NEEDED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WAS NO UNUSUAL HANDLING OF THE BALLOON PRIOR TO ENTRY INTO THE GUIDE AND THE HEMOSTASIS VALVE WAS OPEN AS WELL. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Insufficient Information; No Known Impact Or Consequence To Patient

    MAUDE report 11238011 ↗
  • MalfunctionJan 11, 2021Break

    ELDERLY FEMALE WITH HISTORY OF CORONARY ARTERY DISEASE. PROCEDURE IS ELECTIVE STAGED PERCUTANEOUS CORONARY INTERVENTION OF DIAGONAL ARTERY. AS THE WOLVERINE (2.75 X 10 BALLOON) ENTERED THE GUIDE, SHAFT BROKE INSIDE CATHETER UPON ADVANCEMENT. ALL PARTS WERE SUCCESSFULLY REMOVED WITHOUT KNOWN HARM TO PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 11146806 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P950020. One submission often covers many models, so check the product and lot details in each record.

  • Z-2138-2013 · Jul 31, 2013 · Terminated
    Boston Scientific, FlextomeTM Cutting BalloonTM, Sterilized using irradiation: 1) 4.00mm x 15mm FlextomeTM Cutting BalloonTM MonorailTM Microsurgical Dilatation Device, UPN #H74
  • Z-2137-2013 · Jul 31, 2013 · Terminated
    Boston Scientific, Small Peripheral Cutting BalloonTM, Sterilized using irradiation: The Peripheral Cutting Balloon Catheters are indicated for Percutaneous Transluminal Angiopla
  • Z-0621-2009 · Nov 21, 2008 · Terminated
    Flextome Cutting Balloon Device Over-the-Wire Delivery System, Coronary Atherotomy System, Over-the-Wire Delivery System, Catalog No. CB0340006, 4.00mm, Sterile. Made in Ireland:
  • Z-1512-06 · Jul 21, 2006 · Terminated
    Boston Scientific Cutting Balloon Ultra2 Monorail Device, Boston Scientific, One Boston Scientific Place, Natick, MA 01760.
  • Z-0410-06 · Dec 7, 2005 · Terminated
    The Peripheral Cutting Balloon Device small Monorail Delivery System features a balloon with 3 or 4 atherotomes (microsurgical blades) mounted longitudinally on its outer surface.
  • Z-0409-06 · Dec 7, 2005 · Terminated
    The Flextome Cutting Balloon Device consists of a balloon with 3 or 4 atherotomes (microsurgical blades) mounted longitudinally on its outer surface. When the Flextome Cutting Ball

Other Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring devices

All Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring devices →