Wolverine Coronary Cutting Balloon H74939401102750Recalls & reported problems
Boston Scientific Corporation · Microsurgical Dilatation Device · Class 3
All models of Wolverine Coronary Cutting Balloon (22)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729888253), received between Jan 11, 2021 and Jul 29, 2025 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
6 other recalls cover devices cleared under the same FDA submission (P950020); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Adverse Event Without Identified Device or Use Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information2
- Cardiac Perforation1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 29, 2025Break; Break
IT WAS REPORTED THAT A SHAFT BREAK OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED ARTERY. A 10MMX2.75MM WOLVERINE CORONARY CUTTING BALLOON WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE CATHETER SHAFT BROKE WHILE DELIVERING AND WAS ABLE TO REMOVE IN TWO PIECES WITHOUT SNARING. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE DEVICE. NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22649972 ↗InjuryDec 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SHOCKWAVE LITHOPLASTY COMPARED TO CUTTING BALLOON TREATMENT IN CALCIFIED CORONARY DISEASE: A RANDOMIZED CONTROLLED TRIAL (SHORT-CUT TRIAL). IT WAS REPORTED THAT CARDIAC PERFORATION OCCURRED. ON (B)(6) 2024, THE SUBJECT PRESENTED WITH STABLE ANGINA AND UNDERWENT PERCUTANEOUS CORONARY INTERVENTION (PCI) USING CUTTING BALLOON ANGIOPLASTY IN THE MID LEFT DESCENDING CORONARY ARTERY (LAD). PRE-PROCEDURE TIMI FLOW WAS I WITH 99% STENOSIS. A 10MMX2.75MM WOLVERINE CUTTING BALLOON WAS SELECTED FOR PROCEDURE. DURING PROCEDURE, DRUG ELUTING STENT WAS IMPLANTED IN THE MID LAD FOLLOWED BY POST-STENT DILATION WITH A NON-COMPLAINT BALLOON. A SMALL CARDIAC PERFORATION OF THE MID LAD WAS NOTED AND WAS TREATED WITH A COVERED STENT AND PROTAMINE. POST PROCEDURE TIMI FLOW WAS III WITH 0% STENOSIS. THE PATIENT WAS DISCHARGED THE SAME DAY.
Patient problems: Cardiac Perforation
MAUDE report 21036718 ↗MalfunctionMar 2, 2022Break; Break
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 10FEB2022. IT WAS REPORTED THAT THE BALLOON BROKE. THE TARGET LESION WAS LOCATED IN THE CALCIFIED DIAGONAL ARTERY. A 10MMX2.75MM WOLVERINE CORONARY CUTTING BALLOON WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON BROKE NEAR THE HUB. THE DEVICE WAS PULLED OUT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT A HYPOTUBE BREAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13656548 ↗MalfunctionJun 3, 2021Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT THE CATHETER SHAFT BROKE. THE TARGET LESION WAS LOCATED IN THE TORTUOUS AND MODERATELY CALCIFIED ARTERY. A 10MM X 2.75MM WOLVERINE CORONARY CUTTING BALLOON WAS SELECTED FOR TREATMENT WITHIN THE TARGET LESION. THE BALLOON WAS INFLATED AND DEFLATED WITHOUT ISSUE. AS THE WOLVERINE CORONARY CUTTING BALLOON WAS BEING WITHDRAWN, IT KINKED AND THEN BROKE INTO TWO PIECES. THE DEVICE WAS PULLED BACK INTO THE GUIDE CATHETER AND FULLY REMOVED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11931583 ↗MalfunctionJan 27, 2021Break; Break
IT WAS REPORTED VIA MAUDE REPORT 3600840000-2021-8005 THAT SHAFT BREAK OCCURRED. A 10MM X 2.75MM WOLVERINE CORONARY CUTTING BALLOON WAS SELECTED FOR USE. DURING THE PROCEDURE, UPON ADVANCING THE BALLOON THROUGH THE GUIDE OVER A WIRE, IT WAS NOTED THAT THE SHAFT PROXIMAL TO THE ENTRY POINT SNAPPED IN HALF. THE DEVICE WAS REMOVED COMPLETELY FROM THE GUIDE WITHOUT ANY INTERVENTION OR DEVICES NEEDED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WAS NO UNUSUAL HANDLING OF THE BALLOON PRIOR TO ENTRY INTO THE GUIDE AND THE HEMOSTASIS VALVE WAS OPEN AS WELL. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Insufficient Information; No Known Impact Or Consequence To Patient
MAUDE report 11238011 ↗MalfunctionJan 11, 2021Break
ELDERLY FEMALE WITH HISTORY OF CORONARY ARTERY DISEASE. PROCEDURE IS ELECTIVE STAGED PERCUTANEOUS CORONARY INTERVENTION OF DIAGONAL ARTERY. AS THE WOLVERINE (2.75 X 10 BALLOON) ENTERED THE GUIDE, SHAFT BROKE INSIDE CATHETER UPON ADVANCEMENT. ALL PARTS WERE SUCCESSFULLY REMOVED WITHOUT KNOWN HARM TO PATIENT.
Patient problems: Insufficient Information
MAUDE report 11146806 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P950020. One submission often covers many models, so check the product and lot details in each record.
- Z-2138-2013 · Jul 31, 2013 · TerminatedBoston Scientific, FlextomeTM Cutting BalloonTM, Sterilized using irradiation: 1) 4.00mm x 15mm FlextomeTM Cutting BalloonTM MonorailTM Microsurgical Dilatation Device, UPN #H74
- Z-2137-2013 · Jul 31, 2013 · TerminatedBoston Scientific, Small Peripheral Cutting BalloonTM, Sterilized using irradiation: The Peripheral Cutting Balloon Catheters are indicated for Percutaneous Transluminal Angiopla
- Z-0621-2009 · Nov 21, 2008 · TerminatedFlextome Cutting Balloon Device Over-the-Wire Delivery System, Coronary Atherotomy System, Over-the-Wire Delivery System, Catalog No. CB0340006, 4.00mm, Sterile. Made in Ireland:
- Z-1512-06 · Jul 21, 2006 · TerminatedBoston Scientific Cutting Balloon Ultra2 Monorail Device, Boston Scientific, One Boston Scientific Place, Natick, MA 01760.
- Z-0410-06 · Dec 7, 2005 · TerminatedThe Peripheral Cutting Balloon Device small Monorail Delivery System features a balloon with 3 or 4 atherotomes (microsurgical blades) mounted longitudinally on its outer surface.
- Z-0409-06 · Dec 7, 2005 · TerminatedThe Flextome Cutting Balloon Device consists of a balloon with 3 or 4 atherotomes (microsurgical blades) mounted longitudinally on its outer surface. When the Flextome Cutting Ball
Other Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring devices
- Wolverine Coronary Cutting Balloon H74939401103000Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring1 death36 reports
- Wolverine Coronary Cutting Balloon H74939401103500Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring21 reports
- Wolverine Coronary Cutting Balloon H74939401153000Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring19 reports
- Wolverine Coronary Cutting Balloon H74939401153500Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring19 reports
- Wolverine Coronary Cutting Balloon H74939401104000Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring19 reports
- Wolverine Coronary Cutting Balloon H74939401102500Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring16 reports
- Wolverine Coronary Cutting Balloon H74939401154000Boston Scientific Corporation · Catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring15 reports
- AngioSculpt PTCA RX Scoring Balloon Catheter 2200-3010Spectranetics Corporation · Catheter, Percutaneous Transluminal Coronary Angioplasty (Ptca), Cutting/Scoring2 deaths14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page