WallFlex™ Colonic M00555780Recalls & reported problems

Boston Scientific Corporation · Stent System with Anchor Lock Delivery System · Class 2 · Implant

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729902607), received between Jun 1, 2022 and Dec 27, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Remove1
  • Material Deformation1
  • Material Integrity Problem1
  • Positioning Failure1
  • Positioning Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 27, 2024Positioning Failure; Material Deformation; Adverse Event Without Identified Device or Use Problem; Positioning Failure; Material Deformation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A WALLFLEX ENTERAL COLONIC STENT WAS TO BE USED TO TREAT A STENOSIS IN THE LOWER COLON DURING A COLONIC STENTING FOR COLON CANCER WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE STENT WAS IMPOSSIBLE TO DEPLOY. ADDITIONALLY. IT WAS NOTED THAT THE INNER YELLOW SHEATH WAS KINKED, WHICH PREVENTED THE STENT FROM BEING RELEASED. AS A RESULT, THE STENT WAS REMOVED, AND THE PROCEDURE WAS NOT COMPLETED. NO PATIENT COMPLICATION WAS REPORTED DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21026420 ↗
  • MalfunctionAug 23, 2023Difficult to Remove; Difficult to Remove

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2023, THAT A WALLFLEX ENTERAL COLONIC UNCOVERED STENT WAS TO BE IMPLANTED IN THE HEPATIC FLEXURE BILIARY TO TREAT A STRICTURE DURING A COLONIC STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2023. THE PATIENT'S ANATOMY WAS TIGHT. DURING THE PROCEDURE, THE STENT WAS SUCCESSFULLY DEPLOYED; HOWEVER, THE DELIVERY SYSTEM WAS DIFFICULT TO REMOVE AS IT SEEMED TO GET STUCK. THE SURGEON CUT THE SHEATH FROM THE HANDLE AND PULLED IT OUT. THE STENT REMAINS IMPLANTED, AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17602287 ↗
  • InjuryJun 1, 2022Material Integrity Problem; Positioning Problem; Material Integrity Problem; Positioning Problem

    NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT # 3005099803-2022-02898 FOR THE ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2022 THAT TWO WALLFLEX COLONIC SOFT STENT SYSTEM WITH ANCHOR LOCK DELIVERY SYSTEMS WERE IMPLANTED TO TREAT A MALIGNANT COLONIC STRICTURE IN THE SIGMOID COLON AND AS A BRIDGE TO SURGERY DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2022. REPORTEDLY, THE PATIENT'S ANATOMY WAS TORTUOUS. DURING THE PROCEDURE, THE FIRST WALLFLEX COLONIC SOFT STENT (THE SUBJECT OF MFR. REPORT# 3005099803-2022-02898) WAS ABLE TO BE DEPLOYED BUT AFTER STENT DEPLOYMENT, THE STENT WAS NOT VISIBLE UNDER ENDOSCOPIC VIEW. THE PHYSICIAN SUSPECTED A STENT MALFUNCTION AND DECIDED TO REMOVE THE DELIVERY SYSTEM. A SECOND WALLFLEX COLONIC SOFT STENT (THE SUBJECT OF THIS REPORT) WAS DEPLOYED; BUT THE SAME PROBLEM OCCURRED, THE SECOND STENT WAS NOT VISIBLE UNDER ENDOSCOPIC VIEW. HOWEVER, WHEN THE PHYSICIAN CHECKED UNDER FLUOROSCOPY, IT WAS NOTED THAT TWO WALLFLEX COLONIC SOFT STENTS WERE IMPLANTED IN AN INCORRECT POSITION. THE STENTS REMAIN IMPLANTED AND THE PROCEDURE WAS STOPPED AS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14568737 ↗

Other Stent, colonic, metallic, expandable devices

All Stent, colonic, metallic, expandable devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.