Skip to content

Ultra ICE Plus M00499100Recalls & reported problems

Boston Scientific Corporation · IntraCardiac Echo Catheter · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729904373), received between Mar 15, 2018 and Mar 15, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 15, 2018Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT A CATHETER BREAK / DETACHMENT OCCURRED. AN ULTRA ICE PLUS ULTRASOUND IMAGING CATHETER AND ANOTHER MANUFACTURER¿S SNARE WERE SELECTED FOR USE DURING AN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) LEAD EXTRACTION PROCEDURE IN THE RIGHT ATRIUM. DURING A TWISTING MANEUVER WITH THE SNARE RETRIEVAL DEVICE, THE CATHETER WAS ENTRAPPED WHILE IT WAS ROTATING AND THEREFORE BECAME DAMAGED. DURING RETRIEVAL OF THE DEVICES, THE PIEZOELECTRIC CRYSTAL BROKE AND DETACHED FROM THE CATHETER BODY. IT WAS NOTED THE DETACHMENT ¿PROBABLY¿ OCCURRED INSIDE THE SHEATH. ALL COMPONENTS WERE REMOVED FROM THE PATIENT WITHOUT PROBLEMS AND THE PROCEDURE WAS COMPLETED USING A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7341241 ↗

Other Transducer, ultrasonic, diagnostic devices

All Transducer, ultrasonic, diagnostic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page