Vercise™ DB-9218-55Recalls & reported problems
Boston Scientific Neuromodulation Corporation · M8 Adapter, 55cm · Class 3 · Implant
All models of Vercise (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729905295), received between Jun 29, 2023 and Feb 16, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “High impedance” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (P150031); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- High impedance2
- Adverse Event Without Identified Device or Use Problem1
- Unexpected Therapeutic Results1
Patient problems
- Pain1
- Shaking/Tremors1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 16, 2024High impedance; Unexpected Therapeutic Results; Adverse Event Without Identified Device or Use Problem; High impedance; Unexpected Therapeutic Results; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED INADEQUATE STIMULATION RESULTING IN A LACK OF TREMOR CONTROL. HIGH IMPEDANCE READINGS WERE OBSERVED, AND THE PHYSICIAN ASSESSED THAT A FALL THAT HAD OCCURRED SEVERAL MONTHS PRIOR WAS A CONTRIBUTING FACTOR. REPROGRAMMING ATTEMPTS WERE UNSUCCESSFUL, AND THE PATIENT UNDERWENT A PROCEDURE FOR SURGICAL TROUBLESHOOTING. DURING THE PROCEDURE, WHEN DISCONNECTING AND RECONNECTING THE M8 ADAPTER FROM THE IMPLANTABLE PULSE GENERATOR (IPG) DID NOT RESOLVE THE IMPEDANCE ISSUES, THE M8 ADAPTER WAS REPLACED. WHEN THE IMPEDANCE READINGS WERE ABNORMAL FOR WHAT WOULD BE EXPECTED WHEN CONNECTING TO A NON-BOSTON SCIENTIFIC LEAD, THE IPG WAS REPLACED. THE PATIENT DID WELL POSTOPERATIVELY, AND IMPEDANCE READINGS WERE WITHIN NORMAL LIMITS.
Patient problems: Shaking/Tremors
MAUDE report 18728072 ↗InjuryJun 29, 2023High impedance; High impedance
IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED HIGH IMPEDANCES. REPROGRAMMING AROUND THE IMPEDANCES WAS ATTEMPTED BUT WAS UNSUCCESSFUL. THE PATIENT UNDERWENT A REVISION PROCEDURE WHERE THE DBS ADAPTOR WAS EXPLANTED. THE EXPLANTED DEVICE WAS NOT RETURNED TO BSC.
Patient problems: Pain
MAUDE report 17227345 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P150031. One submission often covers many models, so check the product and lot details in each record.
- Z-2340-2025 · Jul 8, 2025 · Open, ClassifiedBurr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
- Z-2558-2025 · Jul 8, 2025 · Open, ClassifiedVercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
- Z-1890-2024 · Apr 18, 2024 · Open, ClassifiedVercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
- Z-2125-2020 · Apr 24, 2019 · TerminatedFifty-five (55) cm 8 Contact Extension Kit, part number M365NM3138550 - Product Usage: The Deep Brain Stimulation (DBS) System is indicated for use in bilateral stimulation of the
Other STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS devices
- ACTIVA RC 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS2,839 reports
- Activa® 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS1,325 reports
- Percept™ RC B35300Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian symptoms1 recall3 deaths796 reports
- Vercise™ Cartesia™ DB-2202-45Boston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS18 deaths718 reports
- Activa® PC 37601Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS3 deaths521 reports
- Activa® RC 37612Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian symptoms1 death469 reports
- Activa® RC 37612Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian symptoms410 reports
- NA NM-3138-55Boston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS198 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.