ACUITY® Pro 8105 (UDI 08714729932628)Recalls & reported problems

Boston Scientific Corporation · Lead Delivery System · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729932628), received between Jan 29, 2019 and Mar 26, 2024 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem3
  • Defective Device1
  • Difficult to Insert1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1
  • Perforation1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 26, 2024Defective Device; Defective Device

    IT WAS REPORTED THAT THIS INNER GUIDING CATHETER WAS DAMAGED WHILE THE PHYSICIAN WAS CUTTING THE SHEATH. THIS GUIDING CATHETER WAS NEVER IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18983966 ↗
  • MalfunctionMar 18, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THIS INNER GUIDING CATHETER WAS FRACTURED. THIS GUIDING CATHETER WAS NEVER IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18926558 ↗
  • MalfunctionMar 1, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT DURING PROCEDURE, AFTER THE PHYSICIAN CUT THIS GUIDING CATHETER AWAY, A REMNANT WAS OBSERVED ON THE LEAD BODY OUTSIDE OF THE PATIENT. THE PHYSICIAN TRIED REMOVING THE REMNANT WITH A CUTTER BUT HAD TO REMOVE IT FROM THE LEAD BODY USING A SMALL PAIR OF SCISSORS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18822740 ↗
  • InjuryApr 17, 2019Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT DURING A LEAD IMPLANT PROCEDURE, IT WAS EVIDENT THROUGH CONTRAST THAT A VERY SMALL PERFORATION HAD OCCURRED IN THE CORONARY SINUS WHEN THE CATHETER WAS INSERTED. THE CONTRAST MEDIUM DISSIPATED BY THE END OF THE PROCEDURE AND THE PHYSICIAN STATED THAT NO FURTHER ACTION WAS NEEDED. THE PATIENT WAS STABLE THROUGHOUT AND FOLLOWING THE PROCEDURE WITH NO ADDITIONAL CONSEQUENCES.

    Patient problems: Perforation

    MAUDE report 8523066 ↗
  • MalfunctionJan 29, 2019Material Integrity Problem; Material Integrity Problem

    BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING EXPLANT, THIS CATHETER HAD A PROBLEM WITH USING THE SPLITTER TOOL TO SPLIT THE HUB OF THE CATHETER. ONCE IT WENT DOWN THE CATHETER, THE PARTS OF THE SHAFT HAD FRAGMENTED AND FRAYED. NOTHING WAS LEFT IN THE PATIENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8286043 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →