Skip to content

Tria™ Firm M0061902220Recalls & reported problems

Boston Scientific Corporation · Ureteral Stent with Side Holes 6Fx24CM · Class 2 · Implant

All models of Tria Firm (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 08714729937845), received between Aug 2, 2019 and Apr 28, 2025 — classified as 2 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Calcified2
  • Detachment of Device or Device Component2
  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Remove1
  • Obstruction of Flow1
  • Unsealed Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • No Known Impact Or Consequence To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 28, 2025Break; Break

    IT WAS REPORTED THAT DURING THE PREPARATION, WHILE REMOVING THE SUTURE FROM THE STENT PRIOR TO PLACING IT IN THE PATIENT, IT CAUSED THE END OF THE STENT TO SNAP/BREAK OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21921956 ↗
  • MalfunctionApr 22, 2025Break; Break

    IT WAS REPORTED THAT A TRIA URETERAL STENT WAS USED IN A PROCEDURE. DURING THE PROCEDURE, IT WAS NOTICED THAT THE STENT WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21879054 ↗
  • MalfunctionMar 21, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING PREPARATION, AFTER REMOVING THE PACKAGING, THE STENT WAS BROKEN WHEN THE THREAD WAS REMOVED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THE PATIENT CONDITION WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21665693 ↗
  • MalfunctionJul 8, 2024Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED THAT, DURING UNPACKING, THE SEAL OF THE TRIA URETERAL STENT PACKAGING WAS TORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19697495 ↗
  • MalfunctionSep 14, 2023Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRIA FIRM URETERAL STENT WAS ATTEMPTED TO BE USED DURING A STENT REPLACEMENT PROCEDURE. THE EXACT EVENT DATE WAS NOT PROVIDED. DURING THE PROCEDURE, THE STENT WAS FOUND FRACTURED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER TRIA FIRM URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE AS NO PATIENT INJURY. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17749444 ↗
  • MalfunctionMay 5, 2023Break; Break

    IT WAS REPORTED A TRIA FIRM URETERAL STENT WAS USED DURING A STENT REPLACEMENT PROCEDURE IN THE URETER PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE STENT WAS SEPARATED WHEN REMOVING THE SUTURE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER TRIA FIRM URETERAL STENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16878085 ↗
  • MalfunctionNov 16, 2022Calcified; Difficult to Remove; Calcified; Difficult to Remove

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRIA URETERAL STENT WAS USED IN A STENT PLACEMENT PROCEDURE FOR MALIGNANT OBSTRUCTION IN THE LEFT URETER PERFORMED ON (B)(6) 2022, AS PART OF THE U0652 DOUBLE-J REGISTRY CLINICAL STUDY. THE SUBJECT WAS PRE-STENTED ON (B)(6) 2021 WHICH WAS REMOVED ON (B)(6) 2022, AND A NEW TRIA URETERAL STENT WAS PLACED RETROGRADE ON THE LEFT URETER UNDER CYSTOSCOPY. THE STENT WAS SUCCESSFULLY IMPLANTED UNDER THE DIRECTION FOR USE AND WAS PERFORMED WITHOUT ANY COMPLICATIONS. DISCHARGE MEDICATIONS INCLUDED ANTIBIOTIC AND PHENAZOPYRIDINE. ON (B)(6) 2022, THE STENT WAS REMOVED SUCCESSFULLY AS PER THE INITIAL PLAN. PAIN CONTROL WAS REQUIRED AND MONITORED ANESTHESIA CARE (MAC). THE STENT WAS DIFFICULT TO REMOVE DUE TO ENCRUSTATION OF STENT BUT WAS SUCCESSFULLY REMOVED WITH THE USE OF CYSTO/GRASPERS. THE STENT WAS STILL ABLE TO DRAIN AT THE TIME OF THE PROCEDURE. ACCORDING TO THE COMPLAINANT, IT WAS FOUND THAT THE TRIA URETERAL STENT WAS ENCRUSTED. A NEW STENT WAS IMPLANTED TO THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED ON DECEMBER 05, 2022, THE ONSET DATE OF THE EVENT DIFFI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15807241 ↗
  • InjuryOct 27, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED, TO BOSTON SCIENTIFIC CORPORATION. THAT A TRIA FIRM URETERAL STENT WAS USED, DURING AN UNSPECIFIED PROCEDURE ON (B)(6) 2022. DURING THE WITHDRAWAL/CLOSURE OF THE PROCEDURE, THE COIL OF THE STENT DETACHED. THE DETACHED COIL WAS REMOVED VIA GRASPING FORCEPS UNDER RIGID URETEROSCOPY. THE PROCEDURE WAS COMPLETED WITH ANOTHER TRIA FIRM URETERAL STENT. THERE WAS NO PATIENT INJURY OR COMPLICATION AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15681830 ↗
  • InjuryMar 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS TO REPORTED BOSTON SCIENTIFIC CORPORATION THAT A PREVIOUSLY IMPLANTED TRIA FIRM URETERAL STENT WAS REMOVED FROM THE LEFT URETER, PERFORMED ON (B)(6) 2022 AS PART OF THE (B)(6) CLINICAL STUDY. THE TRIA URETERAL STENT WAS IMPLANTED ON (B)(6) 2021 WITH NO PROBLEMS REPORTED DURING PLACEMENT. ON (B)(6) 2022, IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE ER WITH INFECTION. ON (B)(6) 2022, THE TRIA URETERAL STENTS WAS SUCCESSFULLY REMOVED FROM THE PATIENT IN THE OPERATING ROOM. NO ISSUES WERE REPORTED WITH THE DEVICE DURING REMOVAL. A TRIA FIRM URETERAL STENT WAS SUCCESSFULLY PLACED RETROGRADE ON THE LEFT SIDE UNDER FLUOROSCOPY. ON (B)(6) 2022 THE EVENT WAS CONSIDERED RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Infection

    MAUDE report 13830560 ↗
  • MalfunctionJul 2, 2021Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRIA URETERAL STENT WAS USED DURING A PROCEDURE IN THE URETER, PERFORMED ON AN UNKNOWN DATE. DURING PREPARATION, IT WAS NOTICED THAT THE STENT SHAFT WAS BROKEN. ANOTHER TRIA URETERAL STENT WAS OPENED AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12107426 ↗
  • MalfunctionAug 2, 2019Calcified; Obstruction of Flow; Calcified; Obstruction of Flow

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PREVIOUSLY IMPLANTED TRIA URETERAL STENT WAS USED REMOVED FROM A PATIENT DURING URETERAL STENT REMOVAL PROCEDURE PERFORMED IN THE URETER ON (B)(6) 2019. THE EXACT IMPLANT DATE WAS NOT PROVIDED. ACCORDING TO THE COMPLAINANT, THE STENT WAS IMPLANTED FOR 2 MONTHS AND WAS EXPLANTED ON (B)(6) 2019, WHEN THE PLANNED STENT REMOVAL PROCEDURE WAS PERFORMED, THE STENT WAS FOUND TO HAVE BLACK DISCOLORATION. IT WAS NOTICED THAT THE STENT WAS CALCIFIED AROUND THE BLADDER SIDE AND WAS OCCLUDED. REPORTEDLY, THE PHYSICIAN HAD TO CUT THE BLADDER SIDE OF THE STENT WITH A COOPER AND A GUIDEWIRE WAS INSERTED THROUGH THE STENT TO REMOVE THE STENT FROM THE PATIENT. ANOTHER TRIA URETERAL STENT WAS IMPLANTED AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WAS NO SERIOUS INJURY, NOR WERE THERE ANY ADVERSE PATIENT EFFECTS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8857203 ↗

Other Stent, ureteral devices

All Stent, ureteral devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page