OptiCross™ HD H74939352030Recalls & reported problems
Boston Scientific Corporation · Coronary Imaging Catheter · Class 2
All models of OptiCross HD (3)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific OptiCross HD imaging catheter has 421 reports, including 18 injuries and 1 death. Common problems described include poor or blank ultrasound images, catheters that kink, twist or get stuck inside the vessel, and procedures being finished with a different device. A few reports mention a foreign body in the patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 436
- Deaths reported
- 1
- Injuries reported
- 18
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 436 adverse event reports that list this exact device (DI 08714729960768), received between Jan 8, 2019 and Sep 30, 2026 — classified as 1 death, 18 injuries, 417 malfunctions.
The most frequently reported specific device problem is “Poor Quality Image” (159 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem212
- Poor Quality Image159
- Material Deformation60
- Material Integrity Problem58
- Detachment of Device or Device Component50
- Failure to Advance48
- Unable to Obtain Readings46
- Nonstandard Device39
Patient problems
- No Clinical Signs, Symptoms or Conditions421
- No Consequences Or Impact To Patient8
- Foreign Body In Patient3
- Chest Pain1
- Hematoma1
- Hemorrhage/Blood Loss/Bleeding1
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 436 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026No Display/Image; Poor Quality Image; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948184 ↗MalfunctionSep 30, 2026Entrapment of Device; Material Deformation
It was reported that a catheter issue occurred.The 86% stenosed target lesion was located in the moderately calcified left anterior descending artery. An OptiCross HD imaging catheter was advanced for intravascular ultrasound (IVUS) examination of the target lesion. During the procedure, the catheter fractured and became entangled. The procedure was completed using another device of the same model. There were no patient complications, and the patient remained stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945817 ↗MalfunctionSep 29, 2026Unable to Obtain Readings; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933345 ↗MalfunctionSep 28, 2026Unable to Obtain Readings; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923695 ↗MalfunctionSep 24, 2026Entrapment of Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50892208 ↗MalfunctionSep 24, 2026Device Contamination with Chemical or Other Material
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50890739 ↗MalfunctionSep 24, 2026Device Contamination with Chemical or Other Material
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50890387 ↗MalfunctionSep 24, 2026Unable to Obtain Readings; Material Integrity Problem; Air/Gas in Device
It was reported that catheter issue occurred.An OptiCross HD imaging catheter was selected for intravascular ultrasound (IVUS) examination of the target lesion. During preparation, the extension catheter was bulged, it was also noted that the air could not be discharged and the catheter was black during test outside the patient. The procedure was completed with another of same device. The patient was stable following the procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885291 ↗MalfunctionSep 24, 2026Material Deformation
It was reported that catheter issue occurred. An OptiCross HD imaging catheter was selected for ultrasound examination of the left anterior descending (LAD) artery. During the procedure, the device got fractured. Another of the same model was used to complete the procedure. No complications were reported, and the patient was stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50882181 ↗MalfunctionSep 23, 2026Unable to Obtain Readings; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50874656 ↗MalfunctionSep 23, 2026No Display/Image; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50874046 ↗MalfunctionSep 23, 2026No Display/Image; Poor Quality Image; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873775 ↗MalfunctionSep 23, 2026Unable to Obtain Readings; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873773 ↗MalfunctionSep 23, 2026Unable to Obtain Readings; Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870922 ↗MalfunctionSep 3, 2026Material Deformation; Adverse Event Without Identified Device or Use Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50600020 ↗MalfunctionAug 3, 2026Entrapment of Device; Failure to Advance
It was reported that the device became stuck and required intervention. The OptiCross HD imaging catheter was selected for ultrasound examination of the target lesion. During procedure, the physician used an intravascular ultrasound diagnostic catheter, which became stuck inside the vessel and could not advance. Everything was pulled and removed from patient. The procedure experienced a delay of 5 to 10 minutes and completed using a different device. The patient condition following the procedure was stable.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50116341 ↗MalfunctionApr 6, 2026No Display/Image; Nonstandard Device
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2026. IT WAS REPORTED THAT VISUALIZATION ISSUES OCCURRED. THE OPTICROSS HD IMAGING CATHETER WAS ADVANCED FOR ULTRASOUND EXAMINATION OF THE TARGET LESION; HOWEVER, IMAGING COULD NOT BE INTERPRETED DUE TO A BLACKOUT IMAGE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE CATHETER WAS TWISTED, AND THE IMAGING WINDOW WAS RIPPLED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24790989 ↗MalfunctionMar 30, 2026Nonstandard Device; Deformation Due to Compressive Stress
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 05 MAR 2026. IT WAS REPORTED THAT A CATHETER ISSUE OCCURRED. THE MODERATELY STENOSED TARGET LESION, WITH A LENGTH OF 17 MM, WAS LOCATED IN THE LEFT CIRCUMFLEX ARTERY. THE OPTICROSS 6 HD IMAGING CATHETER WAS ADVANCED FOR INTRAVASCULAR ULTRASOUND IVUS EXAMINATION OF THE TARGET LESION. DURING THE PROCEDURE, THE IVUS CATHETER KINKED AND COULD NOT BE INSERTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THE PATIENT REMAINED STABLE, AND NO COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE CATHETER WAS TWISTED AND THAT THE IMAGING WINDOW WAS RIPPLED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24716456 ↗MalfunctionMar 19, 2026No Display/Image; Nonstandard Device
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 26FEB2026. IT WAS REPORTED THAT VISUALIZATION ISSUES OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING (LAD) ARTERY. AN OPTICROSS HD IMAGING CATHETER WAS SELECTED FOR ULTRASOUND EXAMINATION OF THE TARGET LESION BUT COULD NOT SHOW THE IMAGE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE CATHETER WAS TWISTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24641360 ↗MalfunctionFeb 25, 2026Nonstandard Device; Failure to Advance
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 10FEB2026. IT WAS REPORTED THAT CROSSING DIFFICULTIES WERE ENCOUNTERED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS LEFT CIRCUMFLEX ARTERY. THE OPTICROSS HD IMAGING CATHETER WAS ADVANCED FOR ULTRASOUND EXAMINATION OF THE TARGET LESION BUT WAS NOT ABLE TO CROSS THE LESION DUE TO CALCIFICATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS. HOWEVER, DEVICE ANALYSIS REVEALED THE CATHETER WAS TWISTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24442435 ↗MalfunctionFeb 24, 2026No Display/Image; Nonstandard Device
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 02 FEBRUARY 2026. IT WAS REPORTED THAT VISUALIZATION ISSUES OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND SEVERELY CALCIFIED RIGHT CORONARY ARTERY. THE OPTICROSS HD IMAGING CATHETER WAS ADVANCED FOR ULTRASOUND EXAMINATION OF THE TARGET LESION; BUT THE IMAGE COULD NOT BE DISPLAYED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE CATHETER WAS TWISTED AND ENTANGLED WITH THE GUIDEWIRE, AND THE IMAGING WINDOW WAS RIPPLED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24430557 ↗MalfunctionFeb 24, 2026Difficult to Flush
IT WAS REPORTED THAT A CATHETER ISSUE OCCURRED, RESULTING IN AN ABORTED PROCEDURE. THE TARGET LESION WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY. THE OPTICROSS HD IMAGING CATHETER WAS ADVANCED FOR INTRAVASCULAR ULTRASOUND (IVUS) EXAMINATION OF THE TARGET LESION. DURING THE PROCEDURE, THE IVUS CATHETER COULD NOT BE FLUSHED. AS A RESULT, THE PROCEDURE COULD NOT BE COMPLETED. THE PATIENT REMAINED STABLE, AND NO COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24430546 ↗MalfunctionFeb 19, 2026Poor Quality Image; Air/Gas in Device
IT WAS REPORTED THAT A CATHETER ISSUE OCCURRED. THE 95% STENOSED, 32 X 3.0MM TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND CALCIFIED LEFT CIRCUMFLEX ARTERY (LCX). AN OPTICROSS HD CATHETER WAS SELECTED FOR INTRAVASCULAR ULTRASOUND EXAMINATION (IVUS) OF THE TARGET LESION. DURING INITIAL TESTING OUTSIDE THE PATIENT, THE CATHETER HAD BUBBLES WHICH LED TO POOR IMAGE QUALITY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE, AND THE PATIENT WAS STABLE FOLLOWING THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24395381 ↗MalfunctionFeb 19, 2026Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT A CATHETER ISSUE OCCURRED. THE TARGET LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). AN OPTICROSS HD IMAGING CATHETER WAS SELECTED FOR ULTRASOUND EXAMINATION OF THE TARGET LESION. AFTER PRIMING THE CATHETER, IMAGING WAS INITIATED AND SALINE WAS PUSHED IN. A FOREIGN BODY WAS FOUND AT THE TIP OF THE CATHETER, SO THE ORIGINAL CATHETER WAS NOT USED. A NEW IVUS TUBE WAS USED FOR THE PROCEDURE, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24391890 ↗MalfunctionFeb 18, 2026Difficult to Flush
IT WAS REPORTED THAT A CATHETER ISSUE OCCURRED, RESULTING IN AN ABORTED PROCEDURE. THE TARGET LESION WAS LOCATED IN THE ANTERIOR DESCENDING BRANCH ARTERY. THE OPTICROSS HD IMAGING CATHETER WAS ADVANCED FOR INTRAVASCULAR ULTRASOUND (IVUS) EXAMINATION OF THE TARGET LESION. DURING INTRODUCTION, THE DEVICE COULD NOT BE FLUSHED. AS A RESULT, THE PROCEDURE WAS NOT COMPLETED. THE PATIENT REMAINED STABLE, AND NO COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24381391 ↗
Other Catheter, ultrasound, intravascular devices
- OptiCross H749518100Boston Scientific Corporation · Catheter, ultrasound, intravascular2 deaths457 reports
- ViewFlex™ D087031Irvine Biomedical, Inc. · Catheter, ultrasound, intravascular3 deaths341 reports
- OptiCross 18 H7493932800180Boston Scientific Corporation · Catheter, ultrasound, intravascular186 reports
- OptiCross™ 6 HD H74939354080Boston Scientific Corporation · Catheter, ultrasound, intravascular4 deaths177 reports
- Visions PV .014P RX Digital IVUS Catheter 014RPhilips Image Guided Therapy Corporation · Catheter, Ultrasound, Intravascular1 recall1 death84 reports
- Visions PV .018 Digital IVUS Catheter 86700Philips Image Guided Therapy Corporation · Catheter, Ultrasound, Intravascular1 recall67 reports
- OptiCross™ HD H74939352040Boston Scientific Corporation · Catheter, ultrasound, intravascular3 deaths116 reports
- SOUNDSTAR 10439236Biosense Webster Inc · Catheter, ultrasound, intravascular4 deaths107 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page