RHYTHMIA HDx™ M004RA6201US0Recalls & reported problems
Boston Scientific Corporation · Ablation Connection Box (Maestro™) · Class 3
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729966326), received between Dec 14, 2021 and Apr 19, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P020025, P150005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Displays Incorrect Message1
- Material Integrity Problem1
- Temperature Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 19, 2024Temperature Problem; Temperature Problem
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, AN UNKNOWN CATHETER WAS INSERTED INTO THE PATIENT AND THE CATHETER TEMPERATURE WAS "ABNORMAL." THE TEMPERATURE WAS AROUND 45 DEGREES CELSIUS. REPLACING THE "BIG END", REINSERTING AND UNPLUGGING THE TAIL WIRE, RESTARTING IT ALL DID NOT WORK. EVEN REPLACING IT WITH A NEW NEGATIVE ELECTRODE DID NOT WORK. ONLY WHEN THE SIGNAL STATION WAS TURNED OFF AND ONLY THE RF INSTRUMENT WAS TURNED ON, WOULD THE TEMPERATURE RETURN TO 35 DEGREES CELSIUS. AFTER REPLACING THE CONNECTION BOX - MAESTRO G2, THEY WERE ABLE TO CONTINUE THE PROCEDURE WITH THE SAME CATHETER. NO PATIENT COMPLICATIONS OCCURRED. IT IS UNKNOWN IF THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19143895 ↗MalfunctionDec 14, 2021Device Displays Incorrect Message; Material Integrity Problem; Device Displays Incorrect Message; Material Integrity Problem
IT WAS REPORTED THAT DURING A CARDIAC MAPPING AND ABLATION PROCEDURE, AN ERROR OF D07 EXCESSIVE TEMP WAS OBSERVED ON THE MAESTRO SYSTEM. THE METRIQ TUBING WAS SECURED AND NO LEAKS WERE OBSERVED. THE CATHETER AND CABLE WERE EXCHANGED, HOWEVER THE ERROR PERSISTED. THE RHYTHMIA HDX MAESTRO G2 CATHETER CONNECTION BOX WAS NOTED TO BE FRAYED. AND ALSO REPLACED, HOWEVER, THE ERROR WAS NOT RESOLVED. THE PROCEDURE WAS CANCELLED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13006812 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020025, P150005. One submission often covers many models, so check the product and lot details in each record.
- Z-0699-2019 · Nov 28, 2018 · Open, ClassifiedBlazer Open-Irrigated Ablation Batch Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, ar
- Z-0202-2019 · Jun 20, 2018 · Open, ClassifiedBoston Scientific Maestro 4000 Cardiac Ablation System, Firmware Version 5.14, Model Numbers: M00440000, M0044000D0, M0044000H0,M0044000R0, M0044000Z0,
Other catheter, percutaneous, cardiac ablation, for treatment of atrial flutter devices
- QDOT MICRO D139505Biosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter27 deaths915 reports
- IntellaNav MiFi™ Open-Irrigated M004PMR96200Boston Scientific Corporation · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter4 deaths269 reports
- QDOT MICRO D139504Biosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter7 deaths138 reports
- Sphere-9™ Catheter AFR-00001Medtronic, Inc. · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter1 recall5 deaths77 reports
- QDOT MICRO D139401Biosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter2 deaths114 reports
- NaviStar ThermoCool NI75TCDHBiosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter111 reports
- NaviStar RMT ThermoCool NR7TCSIYBiosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter9 deaths103 reports
- IntellaNav MiFi™ Open-Irrigated M004PMR9620K20Boston Scientific Corporation · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter78 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page