NA DB-3128-55Recalls & reported problems

Boston Scientific Neuromodulation Corporation · Lead Extension Kit, 2x8 Contact, 55cm · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729966371), received between Jan 10, 2025 and Jun 16, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “High impedance” (4 reports).

4 other recalls cover devices cleared under the same FDA submission (P150031); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High impedance4
  • Fracture1
  • Migration1
  • Unexpected Therapeutic Results1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Fall1
  • Fatigue1
  • Head Injury1
  • Movement Disorder1
  • Pain1
  • Undesired Nerve Stimulation1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 16, 2025High impedance; Unexpected Therapeutic Results; Migration; High impedance; Unexpected Therapeutic Results; Migration

    IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED BODY FATIGUE, PAIN, INADEQUATE THERAPY ON THEIR LEFT TORSO AND FELT THE IMPLANTABLE PULSE GENERATOR (IPG) ROTATE IN POCKET SITE. FUNCTIONAL TESTING REVEALED A HIGH IMPEDANCE CONTACT THAT WAS REPROGRAMMED AROUND RESULTING WITH ADEQUATE THERAPY AND PATIENT SYMPTOMS RESIDING. THE PHYSICIAN ASSESSED THAT THE PATIENTS SYMPTOMS WERE EXACERBATED WHILE THE DBS THERAPY WAS ON HOWEVER RESIDED ONCE REPROGRAMMED. THE PATIENT UNDERWENT A PROCEDURE WHEREIN THE IPG WAS REPOSITIONED AND SECURED IN THE POCKET SITE AND ADDITIONAL FUNCTIONAL TESTING CONFIRMED HIGH IMPEDANCE ON THE LEAD EXTENSION AS A RESULT WAS REPLACED WITH ANOTHER OF THE SAME MODEL BEFORE COMPLETING THE PROCEDURE. THE PATIENT DID WELL POST-OPERATIVELY.

    Patient problems: Fatigue; Movement Disorder; Pain; Undesired Nerve Stimulation

    MAUDE report 22214311 ↗
  • InjuryMar 31, 2025High impedance; High impedance

    IT WAS REPORTED THAT DURING A SCHEDULED DEEP BRAIN STIMULATION (DBS) ELECTIVE PROCEDURE A HIGH IMPEDANCE WAS DISCOVERED DURING FUNCTIONAL TESTING ON CONTACTS E1 AND E2. ATTEMPTS TO CLEAR THE HIGH IMPEDANCES WAS PERFORMED, HOWEVER WERE UNSUCCESSFUL. THE PHYSICIAN REPLACED THE LEAD EXTENSION WITH ANOTHER OF THE SAME MODEL AS A RESULT ADDITIONAL FUNCTIONAL TESTING REVEALED THE HIGH IMPEDANCES HAD RESOLVED AND PROCEEDED TO COMPLETE THE PROCEDURE. NO ADVERSE EFFECTS WERE NOTED FROM THE HIGH IMPEDANCE ISSUE. THE PATIENT DID WELL POST-OPERATIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21734348 ↗
  • MalfunctionMar 26, 2025Fracture; High impedance; Use of Device Problem; Fracture; High impedance; Use of Device Problem

    IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT UNDERWENT A REVISION PROCEDURE TO REMOVE A FRAGMENTED PORTION OF A LEAD EXTENSION THAT WAS CONNECTED TO THE IMPLANTABLE PULSE GENERATOR (IPG). THE LEAD EXTENSION HAD SNAPPED AND FRACTURED WHILE ATTEMPTING TO REMOVE IT DUE TO HIGH IMPEDANCES DURING A PREVIOUSLY REPORTED REVISION (SEE MFR REPORT TW (B)(4) ) AND REMAINED IMPLANTED IN THE PATIENT. THE LEAD EXTENSION WAS SUCCESSFULLY EXPLANTED. THE PATIENT WAS DOING WELL POSTOPERATIVELY. THE FRAGMENTED LEAD EXTENSION WAS DISPOSED BY THE HOSPITAL AND WAS NOT RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21704307 ↗
  • InjuryJan 10, 2025High impedance; High impedance

    IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED A NUMBER OF FALLS AND SUSTAINED A HEAD INJURY AFTER BEING IMPLANTED. IT WAS NOTED THAT THE HEAD INJURY WAS LIKELY THE CAUSE OF THE HIGH IMPEDANCES PER THE PHYSICIAN'S ASSESSMENT, HOWEVER NO STIMULATION ISSUES WERE NOTED. THE PATIENT UNDERWENT A PROCEDURE WHEREIN AN OPERATING ROOM (OR) CABLE WAS JUST TO DETERMINE THE HIGH IMPEDANCE MEASUREMENT WAS ON THE LEAD AND/OR LEAD EXTENSION. THE LEAD EXTENSIONS WERE REPLACED WITH OTHERS OF THE SAME MODEL WHICH SHOWED THE IMPEDANCE WITHIN RANGE AND THE LEADS REMAINED IMPLANTED. THE PATIENT DID WELL POST-OPERATIVELY. ADDITIONAL INFORMATION HAS NOT BEEN PROVIDED DESPITE GOOD FAITH EFFORTS.

    Patient problems: Fall; Head Injury

    MAUDE report 21123674 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P150031. One submission often covers many models, so check the product and lot details in each record.

  • Z-2558-2025 · Jul 8, 2025 · Open, Classified
    Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
  • Z-2340-2025 · Jul 8, 2025 · Open, Classified
    Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
  • Z-1890-2024 · Apr 18, 2024 · Open, Classified
    Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
  • Z-2125-2020 · Apr 24, 2019 · Terminated
    Fifty-five (55) cm 8 Contact Extension Kit, part number M365NM3138550 - Product Usage: The Deep Brain Stimulation (DBS) System is indicated for use in bilateral stimulation of the

Other Stimulator, electrical, implanted, for parkinsonian symptoms devices