NA DB-3128-95Recalls & reported problems

Boston Scientific Neuromodulation Corporation · Lead Extension Kit, 2x8 Contact, 95cm · Class 3 · Implant

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729966388), received between Jul 3, 2025 and Nov 20, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “High impedance” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (P150031); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High impedance2
  • Appropriate Device Problem Term/Code Not Available1
  • Fracture1

Patient problems

  • Movement Disorder1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 20, 2025High impedance; Appropriate Device Problem Term/Code Not Available; High impedance; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED A LOSS OF THERAPY AFTER AN ACCIDENT PLAYING FOOTBALL. THE PATIENT EXPERIENCED DYSTONIC MOVEMENTS IN THE ARM AND IN THE GAIT MOVEMENT ON THE RIGHT SIDE OF THE BODY. X-RAY IMAGING WAS PERFORMED AND CONFIRMED THE PRESENCE OF HIGH IMPEDANCES. THE PATIENT UNDERWENT A REVISION PROCEDURE WHERE THE LEAD EXTENSION WAS REPLACED, AND THE HIGH IMPEDANCES RESOLVED. THE PATIENT WAS DOING WELL POSTOPERATIVELY.

    Patient problems: Movement Disorder

    MAUDE report 23611409 ↗
  • MalfunctionJul 3, 2025Fracture; High impedance; Fracture; High impedance

    IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED HIGH IMPEDANCES CAUSED BY A FRACTURED LEAD EXTENSION. THE PATIENT UNDERWENT A REVISION PROCEDURE WHERE THE LEAD EXTENSION WAS REPLACED. AFTER REPLACING THE LEAD EXTENSION THERE WERE STILL HIGH IMPEDANCE MEASUREMENTS PRESENT. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS REPLACED. ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE CORRECT EVENT DATE OF THE HIGH IMPEDANCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22396230 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P150031. One submission often covers many models, so check the product and lot details in each record.

  • Z-2558-2025 · Jul 8, 2025 · Open, Classified
    Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
  • Z-2340-2025 · Jul 8, 2025 · Open, Classified
    Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
  • Z-1890-2024 · Apr 18, 2024 · Open, Classified
    Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
  • Z-2125-2020 · Apr 24, 2019 · Terminated
    Fifty-five (55) cm 8 Contact Extension Kit, part number M365NM3138550 - Product Usage: The Deep Brain Stimulation (DBS) System is indicated for use in bilateral stimulation of the