NA DB-3128-95Recalls & reported problems
Boston Scientific Neuromodulation Corporation · Lead Extension Kit, 2x8 Contact, 95cm · Class 3 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729966388), received between Jul 3, 2025 and Nov 20, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “High impedance” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (P150031); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- High impedance2
- Appropriate Device Problem Term/Code Not Available1
- Fracture1
Patient problems
- Movement Disorder1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 20, 2025High impedance; Appropriate Device Problem Term/Code Not Available; High impedance; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED A LOSS OF THERAPY AFTER AN ACCIDENT PLAYING FOOTBALL. THE PATIENT EXPERIENCED DYSTONIC MOVEMENTS IN THE ARM AND IN THE GAIT MOVEMENT ON THE RIGHT SIDE OF THE BODY. X-RAY IMAGING WAS PERFORMED AND CONFIRMED THE PRESENCE OF HIGH IMPEDANCES. THE PATIENT UNDERWENT A REVISION PROCEDURE WHERE THE LEAD EXTENSION WAS REPLACED, AND THE HIGH IMPEDANCES RESOLVED. THE PATIENT WAS DOING WELL POSTOPERATIVELY.
Patient problems: Movement Disorder
MAUDE report 23611409 ↗MalfunctionJul 3, 2025Fracture; High impedance; Fracture; High impedance
IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED HIGH IMPEDANCES CAUSED BY A FRACTURED LEAD EXTENSION. THE PATIENT UNDERWENT A REVISION PROCEDURE WHERE THE LEAD EXTENSION WAS REPLACED. AFTER REPLACING THE LEAD EXTENSION THERE WERE STILL HIGH IMPEDANCE MEASUREMENTS PRESENT. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS REPLACED. ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE CORRECT EVENT DATE OF THE HIGH IMPEDANCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22396230 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P150031. One submission often covers many models, so check the product and lot details in each record.
- Z-2558-2025 · Jul 8, 2025 · Open, ClassifiedVercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
- Z-2340-2025 · Jul 8, 2025 · Open, ClassifiedBurr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
- Z-1890-2024 · Apr 18, 2024 · Open, ClassifiedVercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
- Z-2125-2020 · Apr 24, 2019 · TerminatedFifty-five (55) cm 8 Contact Extension Kit, part number M365NM3138550 - Product Usage: The Deep Brain Stimulation (DBS) System is indicated for use in bilateral stimulation of the
Other Stimulator, electrical, implanted, for parkinsonian symptoms devices
- ACTIVA RC 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS2,839 reports
- Activa® 37612Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS1,325 reports
- Percept™ RC B35300Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian symptoms1 recall3 deaths796 reports
- Vercise™ Cartesia™ DB-2202-45Boston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS18 deaths718 reports
- Activa® PC 37601Medtronic, Inc. · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS3 deaths521 reports
- Activa® RC 37612Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian symptoms1 death469 reports
- Activa® RC 37612Medtronic, Inc. · Stimulator, electrical, implanted, for parkinsonian symptoms410 reports
- NA NM-3138-55Boston Scientific Neuromodulation Corporation · STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS198 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.