NA DB-3216-55Recalls & reported problems

Boston Scientific Neuromodulation Corporation · 16 Contact Lead Extension Kit, 55 cm · Class 3 · Implant

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08714729966395), received between Jul 4, 2025 and Aug 6, 2026 — classified as 8 injuries.

The most frequently reported specific device problem is “High impedance” (3 reports).

4 other recalls cover devices cleared under the same FDA submission (P150031); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • High impedance3
  • Fracture1
  • Migration1

Patient problems

  • Wound Dehiscence3
  • Bacterial Infection2
  • Fluid Discharge2
  • Impaired Healing2
  • Erosion1
  • Fall1
  • Localized Skin Lesion1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 6, 2026Fracture; High impedance; Migration

    It was reported that the patient experienced inadequate stimulation. The patient had a history of falls attributed to orthostatic hypotension associated with Parkinson's disease. The treating physician suspected that a fall may have damaged the Deep Brain Stimulation (DBS) system. As part of the investigation, a surgical exploration was performed at the implantable pulse generator (IPG) pocket. Upon examination, it was observed that the suture loops securing the IPG had broken, the IPG had rotated within the pocket, and the extensions were visibly fractured and twisted. The physician also explored the intermediate incision site and removed the damaged extensions. New extensions were tunneled and connected to the existing leads and IPG. Following reconnection, impedance measurements were within normal limits, and the system was confirmed to be fully operational with no damage identified in the leads or IPG. Postoperatively, the patient was returned to the pre-existing stimulation settings and discharged following the outpatient procedure. The explanted devices will not be returned for analysis, as they were disposed of.

    Patient problems: Shaking/Tremors

    MAUDE report 50168278 ↗
  • InjuryJul 31, 2026High impedance

    It was reported that the deep brain stimulation (DBS) patient underwent a surgical procedure to clean and restore the area above the ear where the wound incision reopened and the lead extension became exposed. Additionally, there were high impedances discovered on a few contacts. The physician believes the cause of the wound reopening was due to how the incision was closed by the physicians assistant, or that this model of lead extension is stiffer than an earlier model. The patient was doing well postoperatively.

    Patient problems: Impaired Healing; Wound Dehiscence

    MAUDE report 50101821 ↗
  • InjuryJul 14, 2026High impedance

    IT WAS REPORTED THAT DURING THE STAGE TWO IMPLANT PROCEDURE OF THE DEEP BRAIN STIMULATION (DBS) HIGH IMPEDANCES WERE OBSERVED ON FIVE CONTACTS. MULTIPLE ATTEMPTS WERE MADE TO RESOLVE THE IMPEDANCES, SUCH AS DISCONNECTING AND RECONNECTING THE LEAD EXTENSION FROM THE LEAD HOWEVER THE IMPEDANCES CONTINUED. THEREFORE, IT WAS DECIDED TO SWITCH OUT THE LEAD EXTENSION WITH A NEW LEAD EXTENSION. HOWEVER, THE HIGH IMPEDANCES CONTINUE ON ONE OF THE CONTACTS, AND IT WAS DETERMINED THAT HIGH IMPEDANCES WERE ON THE LEFT LEAD AND NOT THE LEAD EXTENSION. THE PHYSICIAN ASSESSED THAT THE IMPEDANCES MAY HAVE BEEN TISSUE RELATED, THAT THERE MAY BE AN AIR BUBBLE IN THE BRAIN, HOWEVER THIS WAS NOT CONFIRMED. THE PROCEDURE WAS PROLONGED AND ULTIMATELY COMPLETED, THE PATIENT WAS ADMITTED TO THE HOSPITAL OVERNIGHT BEYOND STANDARD OF CARE AND WAS DISCHARGED THE FOLLOWING DAY AND IS DOING WELL POST OPERATIVELY. IT IS UNCLEAR AS TO WHICH OF THE TWO IMPLANTED LEADS, WAS PLACED ON THE LEFT SIDE AND WAS EXHIBITING THE HIGH IMPEDANCES. IT IS ALSO UNCLEAR AS TO WHICH OF THE TWO LEAD EXTENSIONS WAS ULTIMATELY IMPLANTED AND WHICH WAS DISCARDED BY THE FACILITY DURING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Pneumocephalus

    MAUDE report 25808847 ↗
  • InjuryMay 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION BEHIND THE RIGHT EAR AT THE LEAD EXTENSION INCISION SITE OF THE DEEP BRAIN STIMULATION (DBS) SYSTEM. IT WAS NOTED THAT THERE WAS PUS AND THE WOUND SITE WAS OPEN. THE PATIENT ADMITTED TO PICKING AT THE INCISION SITE, WHICH IS THOUGHT TO BE THE CAUSE OF THE INFECTION. THE PATIENT UNDERWENT A PROCEDURE IN WHICH THE TWO LEAD EXTENSIONS WERE EXPLANTED, WAS PLACED ON ANTIBIOTICS, AND IS DOING WELL POST OPERATIVELY. THE DEVICES WILL NOT BE RETURNED AS THEY WERE DISCARDED BY THE HOSPITAL.

    Patient problems: Bacterial Infection; Fluid Discharge; Wound Dehiscence

    MAUDE report 25225951 ↗
  • InjuryApr 2, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED EROSION AT THE EXTENSION SITE WHERE IT CONNECTS TO THE LEAD, RESULTING IN EXPOSURE OF THE LEAD EXTENSION. THE PATIENT SUBSEQUENTLY UNDERWENT A REVISION PROCEDURE WHERE THE LEAD EXTENSIONS WERE REPLACED. THE PATIENT WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICES WERE DISPOSED OF AT THE FACILITY AND WERE NOT RETURNED.

    Patient problems: Erosion

    MAUDE report 24769514 ↗
  • InjuryDec 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT PRESENTED WITH DEHISCENCE AT THE LEAD EXTENSION INCISION SITE BEHIND THE EAR. IT WAS NOTED THAT THE INCISION WAS NOT HEALING PROPERLY. THE PATIENT UNDERWENT INTERVENTION WHEREIN THE INCISION SITE WAS CLEANED WITH IODINE AND TREATED WITH ANTIBIOTICS BEFORE COMPLETING THE PROCEDURE. THE LEAD EXTENSIONS REMAIN IMPLANTED. IT IS UNCLEAR WHETHER CULTURES WERE OBTAINED. THE PATIENT WAS PLACED ON A REGIMEN OF STEROIDS AND ANTIBIOTICS AND HAS SINCE DEVELOPED A SCAB. ONGOING CARE IS BEING PROVIDED BY THE TREATING PHYSICIAN. ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE PATIENT EXPERIENCED PAIN AND PURULENT DISCHARGE AT THE INCISION SITE. THE PATIENT IS DOING WELL AND WOUND HAS SINCE THEN CLOSED.

    Patient problems: Impaired Healing; Localized Skin Lesion; Pain; Purulent Discharge; Wound Dehiscence

    MAUDE report 23824067 ↗
  • InjuryNov 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT EXPERIENCED A FALL WHERE SHE BUMPED HER HEAD CAUSING A LACERATION NEAR THE LEAD EXTENSION. THE PATIENT UNDERWENT A REVISION PROCEDURE WHERE THE LEAD EXTENSIONS WERE REPLACED TO AVOID ANY POSSIBLE RISK OF INFECTION. THE PATIENT WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICES WERE RETAINED BY THE HOSPITAL AND WERE NOT RETURNED.

    Patient problems: Fall

    MAUDE report 23535384 ↗
  • InjuryJul 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE DEEP BRAIN STIMULATION (DBS) PATIENT UNDERWENT AN EXPLANT PROCEDURE INVOLVING COMPLETE REMOVAL OF THE DBS SYSTEM DUE TO INFECTION. THE PATIENT EXHIBITED FLUID DISCHARGE AT THE EXTENSION SITE. THE PATIENT REMAINS IN RECOVERY, AND ANTIBIOTIC THERAPY WAS INITIATED TO MANAGE THE INFECTION. THE HOSPITAL HAS RETAINED THE EXPLANTED DEVICES, AND THEY WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Bacterial Infection; Fluid Discharge

    MAUDE report 22405274 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P150031. One submission often covers many models, so check the product and lot details in each record.

  • Z-2558-2025 · Jul 8, 2025 · Open, Classified
    Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
  • Z-2340-2025 · Jul 8, 2025 · Open, Classified
    Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
  • Z-1890-2024 · Apr 18, 2024 · Open, Classified
    Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
  • Z-2125-2020 · Apr 24, 2019 · Terminated
    Fifty-five (55) cm 8 Contact Extension Kit, part number M365NM3138550 - Product Usage: The Deep Brain Stimulation (DBS) System is indicated for use in bilateral stimulation of the

Other Stimulator, electrical, implanted, for parkinsonian symptoms devices

All Stimulator, electrical, implanted, for parkinsonian symptoms devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.